NA 672625Recalls & reported problems
Medtronic, Inc. · ACCESSORY 672625 KIT ADP/EXT DF-1 FNL PK · Class 3 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00721902066801), received between Apr 6, 2017 and Apr 1, 2024 — classified as 2 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Difficult to Open or Remove Packaging Material” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Open or Remove Packaging Material2
- Fracture2
- Mechanical Problem2
- Adverse Event Without Identified Device or Use Problem1
- High impedance1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Known Impact Or Consequence To Patient3
- Endocarditis1
- Unspecified Infection1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryApr 1, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT¿S DEVICE WAS MOVED TO THE PATIENT¿S ABDOMEN IN (B)(6) 2023 DUE TO RADIATION TREATMENTS AND AN EXTENDER WAS USED FOR THE RIGHT VENTRICULAR (RV) LEAD. NOW THE LEAD IS NOW EXHIBITING HIGH TIP-TO-COIL PACING IMPEDANCES. THE PHYSICIAN IS SUSPECTING THAT THE ADAPTOR IS THE ROOT CAUSE OF THE IMPEDANCE ISSUES. THE DEVICE IS NOW BEING RETURNED TO THE PECTORAL REGION AND THE ADAPTOR WAS REMOVED. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19013343 ↗MalfunctionDec 19, 2022Difficult to Open or Remove Packaging Material; Difficult to Open or Remove Packaging Material
IT WAS REPORTED THAT WHEN THE ADAPTOR KIT WAS ATTEMPTED TO BE OPENED IN ORDER TO ACCESS A WRENCH TOOL FOR USE AT AN IMPLANT IT WAS NOTED THAT THE WRENCH WAS GLUED TO THE TOP OF THE STERILE PACKAGING, CAUSING THE ADAPTOR TO "FLING OUT" OF THE PACKAGING WHEN THE KIT WAS OPENED, CONTAMINATING THE STERILE FIELD. THIS HAPPENED TWO TIMES WITH TWO ADAPTOR KITS. THE TOOLS WERE NOT USED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16006839 ↗MalfunctionDec 19, 2022Difficult to Open or Remove Packaging Material; Difficult to Open or Remove Packaging Material
IT WAS REPORTED THAT WHEN THE ADAPTOR KIT WAS ATTEMPTED TO BE OPENED IN ORDER TO ACCESS A WRENCH TOOL FOR USE AT AN IMPLANT IT WAS NOTED THAT THE WRENCH WAS GLUED TO THE TOP OF THE STERILE PACKAGING, CAUSING THE ADAPTOR TO "FLING OUT" OF THE PACKAGING WHEN THE KIT WAS OPENED, CONTAMINATING THE STERILE FIELD. THIS HAPPENED TWO TIMES WITH TWO ADAPTOR KITS. THE TOOLS WERE NOT USED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16006778 ↗MalfunctionMay 10, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE HIGH VOLTAGE LEAD ADAPTER WAS ATTEMPTED TO BE IMPLANTED BUT NOT USED. THERE WAS A PRODUCT PERFORMANCE ISSUE WITH THE LEAD ADAPTER. THE LEAD ADAPTER WAS REMOVED FROM IMPLANT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14350497 ↗InjuryAug 22, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO INFECTION OF ENDOCARDITIS. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Endocarditis; Unspecified Infection
MAUDE report 8923575 ↗MalfunctionNov 13, 2018Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT'S RIGHT VENTRICULAR (RV) LEAD HAD TRIGGERED AN ALERT FOR HIGH/UNDEFINED SUPERIOR VENA CAVA (SVC) DE FIBRILLATION COIL IMPEDANCE. IT WAS FURTHER DISCOVERED THAT THE RV DEFIBRILLATION COIL IMPEDANCE DISPLAYED HIGH/UNDEFINED IMPEDANCE AND THERE WAS A SMALL AMOUNT OF NOISY ELECTROGRAMS (EGM) INVOLVING THE RV COIL. ADDITIONALLY, THE RV TIP-TO-RV COIL LEAD IMPEDANCE HAD TRIGGERED A WARNING FOR HIGH/UNDEFINED IMPEDANCE, BUT THE LEAD IS CURRENTLY PROGRAMMED TO A BIPOLAR CONFIGURATION. A LEAD FRACTURE IS SUSPECTED. IT WAS NOTED THAT THE PATIENT HAS LEAD EXTENDERS CONNECTED TO BOTH THE SVC COIL AND RV COIL CONNECTIONS AND IT IS UNCLEAR IF THE SUSPECTED FRACTURE IS ON THE LEAD OR THE EXTENDERS. FOLLOW UP WAS CONDUCTED AND IT HAS BEEN VERIFIED THAT NO REPROGRAMMING NOR INTERVENTION HAS BEEN COMPLETED AT THIS TIME. THE LEAD AND EXTENDERS ALL REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8066816 ↗MalfunctionNov 13, 2018Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT'S RIGHT VENTRICULAR (RV) LEAD HAD TRIGGERED AN ALERT FOR HIGH/UNDEFINED SUPERIOR VENA CAVA (SVC) DE FIBRILLATION COIL IMPEDANCE. IT WAS FURTHER DISCOVERED THAT THE RV DEFIBRILLATION COIL IMPEDANCE DISPLAYED HIGH/UNDEFINED IMPEDANCE AND THERE WAS A SMALL AMOUNT OF NOISY ELECTROGRAMS (EGM) INVOLVING THE RV COIL. ADDITIONALLY, THE RV TIP-TO-RV COIL LEAD IMPEDANCE HAD TRIGGERED A WARNING FOR HIGH/UNDEFINED IMPEDANCE, BUT THE LEAD IS CURRENTLY PROGRAMMED TO A BIPOLAR CONFIGURATION. A LEAD FRACTURE IS SUSPECTED. IT WAS NOTED THAT THE PATIENT HAS LEAD EXTENDERS CONNECTED TO BOTH THE SVC COIL AND RV COIL CONNECTIONS AND IT IS UNCLEAR IF THE SUSPECTED FRACTURE IS ON THE LEAD OR THE EXTENDERS. FOLLOW UP WAS CONDUCTED AND IT HAS BEEN VERIFIED THAT NO REPROGRAMMING NOR INTERVENTION HAS BEEN COMPLETED AT THIS TIME. THE LEAD AND EXTENDERS ALL REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8066808 ↗MalfunctionApr 6, 2017High impedance; High impedance
IT WAS REPORTED THAT AN ADAPTOR WAS ATTEMPTED BUT NOT USED DURING IMPLANT. THE PHYSICIAN NOTED THAT THE PATIENT HAD A LEFT SIDE SYSTEM BEING MOVED TO THEIR ABDOMEN. THE PATIENT'S RIGHT VENTRICULAR (RV) LEAD COIL IMPEDANCE WAS HIGH THROUGH THE ADAPTOR. THE LEAD WAS TESTED UPON REPLACEMENT OF THE ADAPTOR AND THE IMPEDANCE LEVELS WERE NORMAL. THE ADAPTOR WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6468626 ↗
Other permanent pacemaker Electrode devices
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · permanent pacemaker Electrode136 deaths16,004 reports
- CapSureFix Novus MRI™ SureScan® 5076-58Medtronic, Inc. · permanent pacemaker Electrode124 deaths8,405 reports
- Capsurefix® Novus 5076-45Medtronic, Inc. · permanent pacemaker Electrode85 deaths7,050 reports
- CapSureFix Novus MRI™ SureScan™ 407652Medtronic, Inc. · permanent pacemaker Electrode69 deaths6,097 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · permanent pacemaker Electrode56 deaths4,850 reports
- Sprint Quattro Secure® 694765Medtronic, Inc. · permanent pacemaker Electrode73 deaths3,852 reports
- Attain Performa™ MRI SureScan™ 429888Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,377 reports
- CapSureFix Novus MRI™ SureScan™ 407645Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,370 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.