Medtronic, Inc. · ACCESSORY 672625 KIT ADP/EXT DF-1 FNL PK · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00721902066801), received between Apr 6, 2017 and Apr 1, 2024 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Remove Packaging Material” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Open or Remove Packaging Material2
  • Fracture2
  • Mechanical Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • High impedance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient3
  • Endocarditis1
  • Unspecified Infection1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 1, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE PATIENT¿S DEVICE WAS MOVED TO THE PATIENT¿S ABDOMEN IN (B)(6) 2023 DUE TO RADIATION TREATMENTS AND AN EXTENDER WAS USED FOR THE RIGHT VENTRICULAR (RV) LEAD. NOW THE LEAD IS NOW EXHIBITING HIGH TIP-TO-COIL PACING IMPEDANCES. THE PHYSICIAN IS SUSPECTING THAT THE ADAPTOR IS THE ROOT CAUSE OF THE IMPEDANCE ISSUES.  THE DEVICE IS NOW BEING RETURNED TO THE PECTORAL REGION AND THE ADAPTOR WAS REMOVED. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19013343 ↗
  • MalfunctionDec 19, 2022Difficult to Open or Remove Packaging Material; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT WHEN THE ADAPTOR KIT WAS ATTEMPTED TO BE OPENED IN ORDER TO ACCESS A WRENCH TOOL FOR USE AT AN IMPLANT IT WAS NOTED THAT THE WRENCH WAS GLUED TO THE TOP OF THE STERILE PACKAGING, CAUSING THE ADAPTOR TO "FLING OUT" OF THE PACKAGING WHEN THE KIT WAS OPENED, CONTAMINATING THE STERILE FIELD. THIS HAPPENED TWO TIMES WITH TWO ADAPTOR KITS. THE TOOLS WERE NOT USED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16006839 ↗
  • MalfunctionDec 19, 2022Difficult to Open or Remove Packaging Material; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT WHEN THE ADAPTOR KIT WAS ATTEMPTED TO BE OPENED IN ORDER TO ACCESS A WRENCH TOOL FOR USE AT AN IMPLANT IT WAS NOTED THAT THE WRENCH WAS GLUED TO THE TOP OF THE STERILE PACKAGING, CAUSING THE ADAPTOR TO "FLING OUT" OF THE PACKAGING WHEN THE KIT WAS OPENED, CONTAMINATING THE STERILE FIELD. THIS HAPPENED TWO TIMES WITH TWO ADAPTOR KITS. THE TOOLS WERE NOT USED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16006778 ↗
  • MalfunctionMay 10, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE HIGH VOLTAGE LEAD ADAPTER WAS ATTEMPTED TO BE IMPLANTED BUT NOT USED. THERE WAS A PRODUCT PERFORMANCE ISSUE WITH THE LEAD ADAPTER. THE LEAD ADAPTER WAS REMOVED FROM IMPLANT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14350497 ↗
  • InjuryAug 22, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO INFECTION OF ENDOCARDITIS. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Endocarditis; Unspecified Infection

    MAUDE report 8923575 ↗
  • MalfunctionNov 13, 2018Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT'S RIGHT VENTRICULAR (RV) LEAD HAD TRIGGERED AN ALERT FOR HIGH/UNDEFINED SUPERIOR VENA CAVA (SVC) DE FIBRILLATION COIL IMPEDANCE. IT WAS FURTHER DISCOVERED THAT THE RV DEFIBRILLATION COIL IMPEDANCE DISPLAYED HIGH/UNDEFINED IMPEDANCE AND THERE WAS A SMALL AMOUNT OF NOISY ELECTROGRAMS (EGM) INVOLVING THE RV COIL. ADDITIONALLY, THE RV TIP-TO-RV COIL LEAD IMPEDANCE HAD TRIGGERED A WARNING FOR HIGH/UNDEFINED IMPEDANCE, BUT THE LEAD IS CURRENTLY PROGRAMMED TO A BIPOLAR CONFIGURATION. A LEAD FRACTURE IS SUSPECTED. IT WAS NOTED THAT THE PATIENT HAS LEAD EXTENDERS CONNECTED TO BOTH THE SVC COIL AND RV COIL CONNECTIONS AND IT IS UNCLEAR IF THE SUSPECTED FRACTURE IS ON THE LEAD OR THE EXTENDERS. FOLLOW UP WAS CONDUCTED AND IT HAS BEEN VERIFIED THAT NO REPROGRAMMING NOR INTERVENTION HAS BEEN COMPLETED AT THIS TIME. THE LEAD AND EXTENDERS ALL REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8066816 ↗
  • MalfunctionNov 13, 2018Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT'S RIGHT VENTRICULAR (RV) LEAD HAD TRIGGERED AN ALERT FOR HIGH/UNDEFINED SUPERIOR VENA CAVA (SVC) DE FIBRILLATION COIL IMPEDANCE. IT WAS FURTHER DISCOVERED THAT THE RV DEFIBRILLATION COIL IMPEDANCE DISPLAYED HIGH/UNDEFINED IMPEDANCE AND THERE WAS A SMALL AMOUNT OF NOISY ELECTROGRAMS (EGM) INVOLVING THE RV COIL. ADDITIONALLY, THE RV TIP-TO-RV COIL LEAD IMPEDANCE HAD TRIGGERED A WARNING FOR HIGH/UNDEFINED IMPEDANCE, BUT THE LEAD IS CURRENTLY PROGRAMMED TO A BIPOLAR CONFIGURATION. A LEAD FRACTURE IS SUSPECTED. IT WAS NOTED THAT THE PATIENT HAS LEAD EXTENDERS CONNECTED TO BOTH THE SVC COIL AND RV COIL CONNECTIONS AND IT IS UNCLEAR IF THE SUSPECTED FRACTURE IS ON THE LEAD OR THE EXTENDERS. FOLLOW UP WAS CONDUCTED AND IT HAS BEEN VERIFIED THAT NO REPROGRAMMING NOR INTERVENTION HAS BEEN COMPLETED AT THIS TIME. THE LEAD AND EXTENDERS ALL REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8066808 ↗
  • MalfunctionApr 6, 2017High impedance; High impedance

    IT WAS REPORTED THAT AN ADAPTOR WAS ATTEMPTED BUT NOT USED DURING IMPLANT. THE PHYSICIAN NOTED THAT THE PATIENT HAD A LEFT SIDE SYSTEM BEING MOVED TO THEIR ABDOMEN. THE PATIENT'S RIGHT VENTRICULAR (RV) LEAD COIL IMPEDANCE WAS HIGH THROUGH THE ADAPTOR. THE LEAD WAS TESTED UPON REPLACEMENT OF THE ADAPTOR AND THE IMPEDANCE LEVELS WERE NORMAL. THE ADAPTOR WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6468626 ↗

Other permanent pacemaker Electrode devices