Medtronic, Inc. · LEAD 6996SQ41 SUBCUTANEOUS LEAD US · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00721902183065), received between Feb 8, 2018 and Apr 3, 2026 — classified as 8 injuries, 5 malfunctions.

The most frequently reported specific device problem is “High impedance” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High impedance7
  • Fracture5
  • Impedance Problem5
  • Adverse Event Without Identified Device or Use Problem2
  • Low impedance1
  • Positioning Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Known Impact Or Consequence To Patient4
  • Unspecified Infection2
  • No Code Available1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 3, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED AND REPLACED DUE TO INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 24781784 ↗
  • InjuryAug 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO AN INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 22783486 ↗
  • InjuryFeb 5, 2025Fracture; Impedance Problem; Fracture; Impedance Problem

    IT WAS REPORTED THAT THERE WAS RISING RIGHT VENTRICULAR (RV) DEFIBRILLATOR COIL IMPEDANCE AND A SUSPECTED FRACTURE WITH THE TRANSVENOUS RV LEAD. IN ADDITION, THERE WAS RISING SUPERIOR VENA CAVA (SVC) COIL IMPEDANCE AND A SUSPECTED FRACTURE WITH THE RV SUBCUTANEOUS LEAD. THE LEADS WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21314639 ↗
  • MalfunctionNov 7, 2024High impedance; Impedance Problem; High impedance; Impedance Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) COIL LEAD AND SUPERIOR VENA CAVA (SVC) COIL LEAD TRIGGERED ALERTS FOR HIGH OUT OF RANGE (OOR) IMPEDANCES AND A GRADUAL RISE IN BOTH IMPEDANCES WERE NOTED. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20631800 ↗
  • InjuryJul 31, 2023Fracture; High impedance; Fracture; High impedance

    IT WAS FURTHER REPORTED THAT THE SVC WAS PROGRAMMED OFF. THE LEAD REMAINS IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17425077 ↗
  • InjuryDec 12, 2022Positioning Problem; Positioning Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD PULLED UP INTO THE POCKET. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15960237 ↗
  • MalfunctionDec 22, 2021High impedance; High impedance

    IT WAS REPORTED THAT THERE WAS AN ALERT SUPERIOR VENA CAVA (SVC) LEAD FOR HIGH/UNDEFINED PACING IMPEDANCE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13059765 ↗
  • InjuryJul 13, 2021Fracture; High impedance; Fracture; High impedance

    IT WAS REPORTED THAT THERE WAS HIGH/UNDEFINED IMPEDANCES ON THE SUBCUTANEOUS HIGH VOLTAGE LEAD COIL. THE LEAD WAS SUSPECT OF A FRACTURE. THE SUBCUTANEOUS LEAD WAS CAPPED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12160051 ↗
  • MalfunctionJan 27, 2021Low impedance; Impedance Problem; Low impedance; Impedance Problem

    IT WAS REPORTED THAT THE EPICARDIAL PACING LEADS HAS LOW IMPEDANCE THAT HAS BEEN TRENDING DOWN AND TRIGGERED AN IMPEDANCE WARNING. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11237997 ↗
  • MalfunctionSep 13, 2018Fracture; High impedance; Fracture; High impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) HIGH VOLTAGE SUBCUTANEOUS COIL MEASURED HIGH OUT OF RANGE IMPEDANCE. AN X-RAY REVEAL ED THAT THE RV COIL HAD FRACTURED. THE RV COIL REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7875348 ↗
  • InjuryJul 18, 2018Fracture; High impedance; Impedance Problem; Fracture; High impedance

    IT WAS FURTHER REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED THE ALERT FOR HIGH/UNDEFINED SVC IMPEDANCE.

    Patient problems: No Code Available; No Known Impact Or Consequence To Patient

    MAUDE report 7698534 ↗
  • MalfunctionApr 10, 2018Impedance Problem; Impedance Problem

    IT WAS REPORTED THAT THE PATIENT'S RIGHT VENTRICULAR (RV) HIGH VOLTAGE LEADS DISPLAY ELEVATED IMPEDANCE MEASUREMENTS ON THE HIGH VOLTAGE COILS. FOLLOW-UP WAS COMPLETED AND NO REPROGRAMMING HAS BEEN COMPLETED AT THIS TIME. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7417914 ↗
  • InjuryFeb 8, 2018High impedance; High impedance

    IT WAS REPORTED THAT THE PATIENT RECEIVED MULTIPLE INAPPROPRIATE SHOCKS FROM THE RIGHT VENTRICULAR (RV) LEAD DUE TO NOISE. THE LEAD ALSO EXHIBITED SHORT VENTRICULAR INTERVALS. HIGH AND UNDEFINED IMPEDANCE WERE ALSO OBSERVED ON THE PACING AND RIGHT VENTRICULAR DEFIBRILLATION PORTIONS OF THE LEAD. HIGH AND UNDEFINED IMPEDANCE WERE ALSO NOTED ON THE SUBCUTANEOUS DEFIBRILLATION COIL LOCATED IN THE SUPERIOR VENA CAVA (SVC). DEFIBRILLATION THERAPIES WERE TURNED OFF AND THE LEADS REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7257730 ↗