CAPSUREFIX® NOVUS 5076110Recalls & reported problems

Medtronic, Inc. · LEAD 5076110 SM BI SIL SCR-IN STER · Class 3 · Implant

All models of CAPSUREFIX NOVUS (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00721902309656), received between Dec 19, 2019 and Oct 16, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Over-Sensing” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Over-Sensing2
  • Break1
  • High Capture Threshold1
  • Under-Sensing1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 16, 2025Break; Over-Sensing; Break; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED AN INSULATION BREACH. THE RV LEAD WAS REVISED AND WAS ATTEMPTED TO BE FIXED USING SEVERAL LEAD ANCHORING SLEEVES. THE RV LEAD HAS SINCE EXHIBITED OVERSENSING. FAR FIELD R-WAVE (FFRW) OVERSENSING WAS OBSERVED WHILE THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) HAS A PIN PLUG IN THE ATRIAL PORT. THE RV LEAD AND CRT-P REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23308360 ↗
  • MalfunctionJun 12, 2023Over-Sensing; Under-Sensing; Over-Sensing; Under-Sensing

    IT WAS REPORTED THAT T-WAVE OVERSENSING (TWOS) AND VENTRICULAR OVERSENSING WERE OBSERVED ON ELECTROGRAMS, WHICH HAD OCCURRED WHILE THE PATIENT WAS WORKING OUT. IN ADDITION, HIGH THRESHOLDS WERE SEEN ON THE RIGHT VENTRICULAR (RV) LEAD, AND ONE MODE SWITCH HAD OCCURRED. LATER THE RV LEAD HAD TWOS POST VENTRICULAR SENSE (VS) OR UNDERSENSING. ADDITIONALLY, THE RIGHT ATRIAL (RA) LEAD HAD UNDERSENSING AND FAR FIELD R-WAVE (FFRW) OVERSENSING. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17112276 ↗
  • InjuryDec 19, 2019High Capture Threshold; High Capture Threshold

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) LONGEVITY WAS SHORTER THAN EXPECTED AND THE BATTERY VOLTAGE WAS RAPIDLY DEPLETING. IT WAS FURTHER REPORTED THAT THE ATRIAL LEAD HAD EXHIBITED HIGH THRESHOLDS FOR QUITE A WHILE AND THE ATRIAL OUTPUTS WERE REPROGRAMMED. THE ICD AND ATRIAL LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9496722 ↗

Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices

All Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices →