CAPSUREFIX® NOVUS 5076110Recalls & reported problems
Medtronic, Inc. · LEAD 5076110 SM BI SIL SCR-IN STER · Class 3 · Implant
All models of CAPSUREFIX NOVUS (17)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00721902309656), received between Dec 19, 2019 and Oct 16, 2025 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Over-Sensing” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Over-Sensing2
- Break1
- High Capture Threshold1
- Under-Sensing1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryOct 16, 2025Break; Over-Sensing; Break; Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED AN INSULATION BREACH. THE RV LEAD WAS REVISED AND WAS ATTEMPTED TO BE FIXED USING SEVERAL LEAD ANCHORING SLEEVES. THE RV LEAD HAS SINCE EXHIBITED OVERSENSING. FAR FIELD R-WAVE (FFRW) OVERSENSING WAS OBSERVED WHILE THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) HAS A PIN PLUG IN THE ATRIAL PORT. THE RV LEAD AND CRT-P REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23308360 ↗MalfunctionJun 12, 2023Over-Sensing; Under-Sensing; Over-Sensing; Under-Sensing
IT WAS REPORTED THAT T-WAVE OVERSENSING (TWOS) AND VENTRICULAR OVERSENSING WERE OBSERVED ON ELECTROGRAMS, WHICH HAD OCCURRED WHILE THE PATIENT WAS WORKING OUT. IN ADDITION, HIGH THRESHOLDS WERE SEEN ON THE RIGHT VENTRICULAR (RV) LEAD, AND ONE MODE SWITCH HAD OCCURRED. LATER THE RV LEAD HAD TWOS POST VENTRICULAR SENSE (VS) OR UNDERSENSING. ADDITIONALLY, THE RIGHT ATRIAL (RA) LEAD HAD UNDERSENSING AND FAR FIELD R-WAVE (FFRW) OVERSENSING. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17112276 ↗InjuryDec 19, 2019High Capture Threshold; High Capture Threshold
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) LONGEVITY WAS SHORTER THAN EXPECTED AND THE BATTERY VOLTAGE WAS RAPIDLY DEPLETING. IT WAS FURTHER REPORTED THAT THE ATRIAL LEAD HAD EXHIBITED HIGH THRESHOLDS FOR QUITE A WHILE AND THE ATRIAL OUTPUTS WERE REPROGRAMMED. THE ICD AND ATRIAL LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9496722 ↗
Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes438 deaths19,632 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes200 deaths12,708 reports
- INGEVITY™ MRI 7741Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes36 deaths7,874 reports
- INGEVITY™+ 7841Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes15 deaths7,419 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes190 deaths6,776 reports
- INGEVITY™+ 7842Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes21 deaths5,995 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes50 deaths5,027 reports
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes40 deaths5,000 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page