METRx® System 9560757Recalls & reported problems
Medtronic Sofamor Danek, Inc. · INSTRUMENT 9560757 RDINCEXILLUMIN SYS · Class 2
All models of METRx System (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00721902663444), received between Jul 13, 2020 and Jul 13, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K993021); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Material Separation1
Patient problems
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 13, 2020Material Separation; Material Separation
INFORMATION WAS RECEIVED FROM HEALTH CARE PROFESSIONAL VIA MANUFACTURING REPRESENTATIVE REGARDING PATIENT WITH STENOSIS SPONDYLOLISTHESIS FOR SPINAL THERAPY. PATIENT DEMOGRAPHICS: GENDER: MALE; AGE: (B)(6). OUTCOME: ALIVE NO INJURY. EVENT WAS INTRA-OP. PROCEDURE USED WAS MAST TLIF. LEVELS IMPLANTED: L4-S1 IT WAS REPORTED THAT THE LIGHT SOURCE IS INSERTED INTO THE METRX TUBE PER ITS IFU. HOWEVER, AFTER ABOUT AN HOUR, THE LIGHT SOURCE COMES APART AT THE END. THE TWO PIECES OF METAL BASICALLY SPLIT IN TWO, WITH ONE PIECE FALLING OFF, EXPOSING THE LIGHT. OF COURSE THE LIGHT IS HOT SO THIS POSES A BURN RISK TO THE PATIENT AS WELL AS THE SURGICAL STAFF. IT ALSO POSES A FIRE HAZARD TO THE PATIENT AND THE OR. LASTLY, THERE IS THE POTENTIAL FOR THE ONE PIECE OF METAL THAT FALLS OFF TO BECOME LODGED INTO THE PATIENT OR FALL INTO THE WOULD BE UNKNOWN TO THE SURGEON AND STAFF. FORTUNATELY IN THIS CASE, NONE OF THESE MORE SERIOUS ISSUES OCCURRED. THE LIGHTS AND THE PIECE OF DISASSOCIATED METAL WERE REMOVED FROM THE PATIENT AND DISCARDED. NO PATIENT SYMPTOMS WERE REPORTED. NO FURTHER COMPLICATIONS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10262744 ↗MalfunctionJul 13, 2020Break; Break
INFORMATION WAS RECEIVED FROM HEALTH CARE PROFESSIONAL VIA MANUFACTURING REPRESENTATIVE REGARDING PATIENT WITH STENOSIS SPONDYLOLISTHESIS FOR SPINAL THERAPY. PATIENT DEMOGRAPHICS: GENDER: MALE; AGE: (B)(6). OUTCOME: ALIVE NO INJURY. EVENT WAS INTRA-OP. PROCEDURE USED WAS MAST TLIF. LEVELS IMPLANTED: L4-S1 IT WAS REPORTED THAT THE LIGHT SOURCE IS INSERTED INTO THE METRX TUBE PER ITS IFU. HOWEVER, AFTER ABOUT AN HOUR, THE LIGHT SOURCE COMES APART AT THE END. THE TWO PIECES OF METAL BASICALLY SPLIT IN TWO, WITH ONE PIECE FALLING OFF, EXPOSING THE LIGHT. OF COURSE THE LIGHT IS HOT SO THIS POSES A BURN RISK TO THE PATIENT AS WELL AS THE SURGICAL STAFF. IT ALSO POSES A FIRE HAZARD TO THE PATIENT AND THE OR. LASTLY, THERE IS THE POTENTIAL FOR THE ONE PIECE OF METAL THAT FALLS OFF TO BECOME LODGED INTO THE PATIENT OR FALL INTO THE WOULD BE UNKNOWN TO THE SURGEON AND STAFF. FORTUNATELY IN THIS CASE, NONE OF THESE MORE SERIOUS ISSUES OCCURRED. THE LIGHTS AND THE PIECE OF DISASSOCIATED METAL WERE REMOVED FROM THE PATIENT AND DISCARDED. NO PATIENT SYMPTOMS WERE REPORTED. NO FURTHER COMPLICATIONS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10262733 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K993021. One submission often covers many models, so check the product and lot details in each record.
- Z-0992-05 · Jun 20, 2005 · TerminatedMedtronic Sofamor Danek brand METRx System 18 mm Med Procedure Kit; Ref 9560018.
- Z-0993-05 · Jun 20, 2005 · TerminatedMedtronic Sofamor Danek brand METRx System MD Procedure Kit; Ref 9560030.
- Z-0991-05 · Jun 20, 2005 · TerminatedMedtronic Sofamor Danek brand METRx System 16 mm Med Procedure Kit; Ref 9560016.
Other ARTHROSCOPE devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.