NA R2290Recalls & reported problems
Medtronic, Inc. · ANALYZER R2290 PATIENT GLOBAL MODEL 2290 · Class 3
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Patient Global Model 2290 analyzer is a benchtop unit used with a programmer to check pacemakers and other implanted heart devices before and after surgery. The 58 reports mostly describe bench equipment trouble: loss of communication with the programmer, cracked displays, broken connector pins, units falling from a table and damaging their sockets, and measurements drifting outside expected ranges. No deaths and no injuries were reported, and no reports involved a patient. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 58
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 58 adverse event reports that list this exact device (DI 00721902710322), received between Aug 9, 2017 and Mar 9, 2026 — classified as 58 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (24 reports).
10 other recalls cover devices cleared under the same FDA submission (P890003); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical /Electronic Property Problem24
- Use of Device Problem11
- Inadequate User Interface6
- Connection Problem5
- Device Sensing Problem4
- Computer Software Problem3
- Impedance Problem3
- Mechanical Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions30
- No Patient Involvement16
- No Known Impact Or Consequence To Patient12
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 58 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 9, 2026Electrical /Electronic Property Problem
THE ANALYZER WAS RETURNED FOR EVALUATION AS AN ASSOCIATE PRODUCT AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24542360 ↗MalfunctionDec 19, 2025Use of Device Problem; Inadequate User Interface; Use of Device Problem; Inadequate User Interface
IT WAS REPORTED THAT THE PROGRAMMER WAS UNABLE TO RECOGNIZE ITS ANALYZER, AS IT GENERATED AN ERROR MESSAGE INDICATING ¿WARNING: LOSS OF COMMUNICATION,¿ AND THAT THE ANALYZER WAS UNABLE TO FUNCTION PROPERLY. IT WAS ALSO NOTED THAT THE POWER SUPPLY CONNECTOR DID NOT MATCH APPROPRIATELY WITH THE POWER SUPPLY CABLE, AND THE DEVICE WAS SOMETIMES UNABLE TO TURN ON AND, AT OTHER TIMES, GENERATED UNUSUAL NOISE AND WAS UNABLE TO POWER ON. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23847533 ↗MalfunctionMar 24, 2025Device Sensing Problem; Impedance Problem; Device Sensing Problem; Impedance Problem
IT WAS REPORTED THAT THE ANALYZER'S SENSING MEASUREMENTS TRENDED LOWER THAN EXPECTED AND THE LEAD IMPEDANCE MEASUREMENTS TRENDED HIGHER THAN EXPECTED DURING LEAD ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21677878 ↗MalfunctionSep 9, 2024Break; Break
IT WAS REPORTED THAT THE PROGRAMMER ANALYZER CONNECTOR PINS BECAME BROKEN WHILE ATTEMPTING TO CONNECT TO THE ANALYZER SOCKET. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20167237 ↗MalfunctionFeb 28, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE PROGRAMMERS DISPLAY WAS CRACKED. IT WAS NOTED THE RADIOFREQUENCY (RF) HEAD AND ANALYZER WERE NOT WORKING. THE PROGRAMMER, RF HEAD AND ANALYZER WERE RETURNED FOR EVALUATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18798871 ↗MalfunctionJan 11, 2024Signal Artifact/Noise; Use of Device Problem; Impedance Problem; Signal Artifact/Noise; Use of Device Problem; Impedance Problem
IT WAS REPORTED THAT THE ANALYZER WAS UNABLE TO WORK CORRECTLY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18491900 ↗MalfunctionNov 21, 2023Connection Problem; Connection Problem
IT WAS FURTHER REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18180728 ↗MalfunctionAug 17, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE ANALYZER, WHICH WAS RETURNED FOR SERVICE AS IT HAD FALLEN FROM TABLE TO FLOOR WHEN CABLE WAS CONNECTED BREAKING THE ANALYZER SOCKET PUSHING IT INSIDE THE INSTRUMENT, SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17562496 ↗MalfunctionJun 7, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE ANALYZER WAS RETURNED FOR PREVENTIVE MAINTENANCE AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFA CTURER¿S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17080073 ↗MalfunctionMay 22, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE ANALYZER WAS RETURNED TO SERVICE AS AN ASSOCIATE PRODUCT AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16972569 ↗MalfunctionMay 17, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE ANALYZER WHICH WAS RETURNED FOR EVALUATION AS AN ASSOCIATE PRODUCT SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16944147 ↗MalfunctionFeb 17, 2023Electrical /Electronic Property Problem; Inadequate User Interface; Electrical /Electronic Property Problem; Inadequate User Interface
IT WAS REPORTED THAT THE ANALYZER LOST CONTACT WITH THE PROGRAMMER DURING AN ANALYZER SESSION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16395840 ↗MalfunctionJan 23, 2023Computer Software Problem; Operating System Becomes Nonfunctional; Computer Software Problem; Operating System Becomes Nonfunctional
IT WAS REPORTED THAT WHEN PREPARING FOR A PROCEDURE THE ANALYZER DISPLAYED AND ERROR MESSAGE INDICATING THAT CONNECTION HAD BEEN LOST. IN TROUBLESHOOTING THE ANALYZER WAS POWER CYCLED MULTIPLE TIMES FAILING TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16222729 ↗MalfunctionNov 10, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE ANALYZER WAS RETURNED TO SERVICE AS AN ASSOCIATE PRODUCT AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15769233 ↗MalfunctionSep 17, 2022Output Problem; Output Problem
IT WAS REPORTED THAT THE ANALYZER GAVE INCORRECT MEASUREMENTS DURING TESTING/MAINTENANCE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15443071 ↗MalfunctionSep 6, 2022No Display/Image; Failure to Power Up; Use of Device Problem; No Display/Image; Failure to Power Up; Use of Device Problem
IT WAS REPORTED THAT THE ANALYZER SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ASSESSMENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15359738 ↗MalfunctionAug 25, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE ANALYZER WHICH WAS RETURNED FOR SERVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ASSE SSMENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15292555 ↗MalfunctionAug 18, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE ANALYZER WHICH WAS RETURNED WITH THE PROGRAMMER AS AN ASSOCIATED DEVICE SUBSEQUENTLY TESTED OUT-OF-SPECIFICATION DURING MANUFACTURER'S ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15248346 ↗MalfunctionJun 21, 2022Intermittent Communication Failure; Intermittent Communication Failure
IT WAS REPORTED THAT THE PROGRAMMER ANALYZER WHICH WAS RETURNED AS PART OF A PULLBACK PROJECT SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ASSESSMENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14751916 ↗MalfunctionJun 9, 2022Use of Device Problem; Use of Device Problem
IT WAS FURTHER REPORTED THAT THE ANALYZER RETURNED TO SERVICE AND THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14641370 ↗MalfunctionJan 17, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE PROGRAMMER WAS UNABLE TO MAINTAIN TELEMETRY WITH A DEVICE. THE RADIOFREQUENCY HEAD WAS REPLACED BUT THE ISSUE REMAINED. THE PROGRAMMER WAS RETURNED FOR SERVICE. IT WAS FURTHER REPORTED THAT THE ANALYZER SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13265945 ↗MalfunctionNov 3, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE PROGRAMMER ANALYZER WHICH WAS RETURNED FOR PREVENTIVE MAINTENANCE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12743183 ↗MalfunctionNov 2, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT WHEN ATTEMPTING TO CONNECT TO THE ANALYZER, THE PROGRAMMER DISPLAYED AN ERROR MESSAGE. THE ANALYZER WAS RETURNED TO SERVICE AS AN ASSOCIATE PRODUCT AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12737213 ↗MalfunctionAug 18, 2021Electrical /Electronic Property Problem; Failure to Power Up; Use of Device Problem; Electrical /Electronic Property Problem; Failure to Power Up; Use of Device Problem
IT WAS REPORTED THAT THE ANALYZER HAD STOPPED WORKING AND WOULD NO LONGER CONNECT TO THE PROGRAMMER. A SECOND ANALYZER WAS TESTED AND WAS ABLE TO WORK WITH THE PROGRAMMER. THE ANALYZER WAS RETURNED TO SERVICE AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12336670 ↗MalfunctionAug 4, 2021Use of Device Problem; Connection Problem; Use of Device Problem; Connection Problem
IT WAS REPORTED THAT THE PROGRAMMER ANALYZER DOES NOT WORK DUE TO A PROBLEM WITH THE ANALYZER SOCKET. THE ANALYZER WAS RETURNED FOR SERVICE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12276317 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P890003. One submission often covers many models, so check the product and lot details in each record.
- Z-2169-2025 · Jun 18, 2025 · Open, ClassifiedMedtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria A
- Z-2168-2025 · Jun 18, 2025 · Open, ClassifiedMedtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria
- Z-2128-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
- Z-2127-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
- Z-2505-2020 · Jun 4, 2020 · Open, ClassifiedCareLink SmartSync Device Manager, Model Number 24970A
- Z-2506-2020 · Jun 4, 2020 · Open, ClassifiedPatient Connector, Model Number 24967
- Z-0845-2019 · Jan 17, 2019 · TerminatedMedtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer
- Z-0400-2018 · Jul 14, 2017 · TerminatedMyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic device that communicates
- Z-0399-2018 · Jul 14, 2017 · TerminatedMyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The monitor is an external electronic device that interfaces with
- Z-2209-2010 · Jun 11, 2010 · TerminatedMedtronic Carelink Monitor Model 2490C. The Medtronic CareLink Monitor Model 2490C is designed to automatically gather information from the implanted heart device. This automatic w
Other Pacemaker/icd/crt non-implanted components devices
- Gallant™ CDDRA500QSt. Jude Medical, Inc. · Pacemaker/icd/crt non-implanted components2 recalls33 deaths3,028 reports
- NA 229047Medtronic, Inc. · Pacemaker/icd/crt non-implanted components1 recall662 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls559 reports
- CareLink Encore™ 29901AMedtronic, Inc. · Pacemaker/icd/crt non-implanted components2 recalls474 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls422 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls151 reports
- Medtronic CareLink® Programmer 2090AA4Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls82 reports
- Entrant™ CDDRA300QSt. Jude Medical, Inc. · Pacemaker/icd/crt non-implanted components1 recall2 deaths158 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.