N/A 7480100Recalls & reported problems
Medtronic Sofamor Danek, Inc. · INSTRUMENT 7480100 DUAL ENDED FEELER · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00721902760303), received between Apr 13, 2020 and Jun 29, 2021 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break3
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 29, 2021Break; Break
INFORMATION RECEIVED FROM HEALTHCARE PROVIDER VIA MANUFACTURER REPRESENTATIVE REGARDING AN EVENT HAPPENED DURING INTRA-OP FOR A PATIENT UNDERGOING A PROCEDURE OF T10-PELVIS INSTRUMENTED FUSION. THE PRE-OPERATIVE DIAGNOSIS WAS MENTIONED AS ADULT SCOLIOSIS OF THE LUMBAR SPINE. IT WAS REPORTED THAT DUAL ENDED PROBE, THE SURGEON REMOVED THE INSTRUMENT AFTER FEELING THE CANNULA CREATED PRIOR TO PLACING A PEDICLE SCREW AND NOTICED THE BALL END WAS MISSING. UPON X-RAYS TAKEN IN THE ROOM, NOTICED THE METALLIC METAL PIECE LEFT IN THE PATIENT. THEY WERE ABLE TO SUCTION THAT ITEM OUT AND REPLACED THE INSTRUMENT WITH ANOTHER PROBE FOR THE REMAINDER OF THE PROCEDURE. THE TIP WAS BROKEN ABOUT 10MM OFF AND THE BROKEN PART WAS RETRIEVED SUCCESSFULLY. THERE WAS A DELAY OF ABOUT 30 MINUTES REPORTED AND MENTIONED DELAY WAS DUE TO THE BROKEN INSTRUMENT. ALL THE REPORTED PRODUCTS CAME IN CONTACT WITH PATIENT. THERE WERE NO SYMPTOMS TO PATIENT OR PHYSICIAN WERE REPORTED. NO ADDITIONAL TREATMENT OR SURGERY PERFORMED. NO FURTHER COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12082418 ↗InjuryJan 19, 2021Break; Break
INFORMATION WAS RECEIVED FROM THE HEALTHCARE PROFESSIONAL (HCP) VIA MANUFACTURER REPRESENTATIVE FOR AN EVENT WHICH OCCURRED DURING POSTERIOR LUMBAR TLIF SURGERY ON A PATIENT DIAGNOSED WITH LUMBAR DISC HERNIATION. IT WAS REPORTED THAT THE PROBE BROKE IN THE PATIENT DURING THE NORMAL USE OF THE PROBE IN THE SURGERY. IN ORDER TO REMOVE THE PROBE, MORE PEDICLES NEEDED TO BE DEALT WITH IN ORDER TO REMOVE THE BROKEN PROBE IN THE PATIENT. THE OPERATION TIME WAS EXTENDED BY ONE AND A HALF HOURS WHILE REMOVING MORE BONE, AND THE DOCTOR¿S EXPOSURE TIME TO X-RAYS INCREASED BY ONE AND A HALF HOURS. PRODUCT WILL BE RETURNED, AND IT WAS REPLACED WITH MEDTRONIC PRODUCT. NO OTHER PATIENT INJURY / COMPLICATION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11192296 ↗InjuryApr 13, 2020Break; Break
IT WAS REPORTED THAT INTRA-OP, WHILE THE SOUNDING FEELER PROBE WAS BEING USED IN THE VERTEBRAL BODY, THE TIP OF THE INSTRUMENT BROKE AND WAS LEFT INSIDE THE VERTEBRAL BODY. X-RAY WAS DONE TO CONFIRM THE EXACT LOCATION OF THE BROKEN TIP OF THE INSTRUMENT. THE RESULTS OF THIS X-RAY ARE UNKNOWN CURRENTLY.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9953725 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page