Skip to content

AMSCO 7052HP (REF FH18082)Recalls & reported problems

Corporation Steris Canada · The AMSCO 7052HP Washer-Disinfector with Steam Heat delivers a validated and cleared process for cleaning reusable da Vinci EndoWrist X/Xi, S/Si, and Single-Site instruments and staplers. · Class 2

All models of AMSCO (25)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00724995176563), received between Nov 2, 2023 and Mar 6, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Installation-Related Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Installation-Related Problem1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 6, 2024Use of Device Problem; Use of Device Problem

    THE USER FACILITY REPORTED VIA MEDWATCH (B)(4), "OUR AMSCO 7052 WASHER HAS THE ROBOT ACCESSORY WASHER OPTION ON IT, WHICH USES A SPECIAL WASHING RACK IN IT. THE 7052 HOLDS THE RACKS DOWN TO THE WATER NOZZLE USING TWIN ROLLERS IN THE MIDDLE OF THE MACHINE, WHICH HOLD ONTO A FLANGE ON THE WASHING RACK. THE ROBOT RACK THAT WE HAVE HAS THE FLANGE ONLY ON TWO SIDES. THE PROBLEM THAT WE HAVE FOUND IS THAT THE RACK CAN BE INSERTED INTO THE WASHER WITH FLANGES IN THE FORWARD TO BACK CONFIGURATION INSTEAD OF THE SIDE TO SIDE, WHICH DOESN'T ALLOW THE ROLLERS TO GRAB THE WASH RACK. WITH THE ROLLERS DISENGAGED, THE WASH RACK IS FREE TO FLOAT AROUND IN THE WASHER WHEN WATER FLOWS OUT OF THE CENTRAL POST. THIS RAISES THE POTENTIAL FOR DAMAGE IN THE MACHINE. THE PREVIOUS WASHER RACK DESIGN WE HAVE HAS THE FLANGE ALL THE WAY AROUND THE WATER ATTACHMENT POST AND HOLDS THE CART DOWN ON THE SIDE RAILS, SO THE ORIENTATION OF THE RACK ISN'T IMPORTANT." NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18852324 ↗
  • MalfunctionNov 2, 2023Installation-Related Problem; Installation-Related Problem

    THE USER FACILITY REPORTED VIA MEDWATCH # (B)(4) THAT AFTER STERIS'S INSTALLATION OF THE AMSCO 7052HP WASHER/DISINFECTOR, THE UNIT WAS EXPERIENCING A LOT OF "PROBLEMS". NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18062243 ↗

Other Disinfector, medical devices devices

All Disinfector, medical devices devices →