GETINGE CM320-SERIES (REF CM320WUWD)Recalls & reported problems
Getinge Disinfection AB · CM320-4 CHAMBER WUWD. Washer disinfector with four chambers; Washer chamber followed by Ultra-sonic chamber, washer chamber and dryer chamber. · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07340153700192), received between May 26, 2025 and Apr 20, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fire” (2 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 13, 2022Open, Classified
Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury or slip and fall injury. May lead to a patient procedural delay and/or infection
Z-1249-2022 · Root cause: Process control · FDA record ↗
- Dec 21, 2020Terminated
The potential that the plastic inlet hose connections may deform or break as a result of material fatigue and/or over tightening of the clamps. A broken hose connection may lead to a hazardous situation of water leakage or water spray. The water leakage or water spray may result in a burn to the user and/or an inconvenience due to a procedural delay if the Washer Disinfector is taken out of service.
Z-1124-2021 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Fire2
- Detachment of Device or Device Component1
- Overheating of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Abrasion1
- Angioedema1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 20, 2026Detachment of Device or Device Component
A COMPLAINT WAS RECEIVED BY GETINGE REGARDING A CM320 WASHER-DISINFECTOR. DURING OPERATION ON THE UNIT, AN INCIDENT OCCURRED IN WHICH THE PANEL THAT SITS ABOVE THE SCREEN/ENTRY CHAMBER 1 DOOR (SOILED SIDE) DETACHED FROM THE UNIT AND FELL ONTO A STAFF MEMBER¿S ARM. INITIAL INSPECTION REVEALED ONE OF THE FOUR HOOKS THAT SECURES THE PANEL TO THE UNIT SNAPPED OFF AND APPEARS BENT. THE STAFF MEMBER WAS ABLE TO COMPLETE THEIR SHIFT BUT SUBSEQUENTLY REPORTED THE INJURY (GRAZE AND/OR BRUISING AND SWELLING), AND CALLED IN SICK FOR THEIR NEXT DAY SCHEDULED SHIFT.
Patient problems: Abrasion; Angioedema
MAUDE report 24921725 ↗MalfunctionDec 11, 2025Fire; Fire; Overheating of Device
GETINGE RECEIVED INFORMATION THAT THE HEATING WIRE IN THE ULTRASONIC CAVITY FOR THIS DEVICE TURNED RED DURING OPERATION. UPON INSPECTION, IT WAS FOUND THAT THE SHORTING WIRE OF THE HEATING ELEMENT HAD BURNT OUT AND REQUIRED REPLACEMENT. IT WAS REPORTED THAT THE FUSE WAS NOT IMPACTED. THERE WERE NO INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23776775 ↗MalfunctionMay 26, 2025Fire; Fire
GETINGE BECAME AWARE OF AN ISSUE INVOLVING THE CM320-SERIES WASHER DISINFECTOR. INITIAL INFORMATION INDICATED THAT A FIRE OCCURRED UPON POWERING ON THE EQUIPMENT IN THE MORNING. ACCORDING TO THE INFORMATION PROVIDED BY THE SALES AND SERVICE UNITS, THE CAUSE OF THE INCIDENT COULD NOT BE CONCLUSIVELY DETERMINED AT THE TIME. THE FIRE APPEARS TO HAVE ORIGINATED IN THE AREA BETWEEN THE ULTRASOUND CHAMBER (SECOND CHAMBER) AND THE CLEANING AND DISINFECTION CHAMBER (THIRD CHAMBER), WITH VISIBLE DAMAGE OBSERVED BENEATH THE THIRD CHAMBER. THE FACILITY SPRINKLER SYSTEM WAS ACTIVATED AND FUNCTIONED AS INTENDED, WHICH SUCCESSFULLY CONTAINED THE FIRE. AT THE TIME OF THAT INITIAL INFORMATION RECEIVED THE CAUSE OF THE FIRE WAS NOT KNOWN TO BE DEVICE RELATED. AN INVESTIGATION WAS INITIATED, AND THE RETURN OF RELEVANT PARTS WAS REQUESTED TO SUPPORT CAUSE ANALYSIS. AT THE TIME OF THIS REPORT, NO SERIOUS INJURIES HAVE BEEN REPORTED IN CONNECTION WITH THE INCIDENT, HOWEVER WE HAVE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL OF HARM IF THE ISSUE WAS TO REOCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22083718 ↗
Other Disinfector, Medical Devices devices
- GETINGE 88-SERIES 88-5Getinge Disinfection AB · Disinfector, Medical Devices3 recalls6 reports
- GETINGE 88-SERIES 88-5Getinge Disinfection AB · Disinfector, Medical Devices3 recalls0 reports
- GETINGE CM320-SERIESGetinge Disinfection AB · Disinfector, Medical Devices2 recalls1 report
- GETINGE 86-SERIES S-8668Getinge Disinfection AB · Disinfector, Medical Devices1 recall37 reports
- Getinge Aquadis 56 56MGetinge Disinfection AB · Disinfector, Medical Devices1 recall11 reports
- AMSCO 7052HPCorporation Steris Canada · Disinfector, medical devices1 recall10 reports
- GETINGE 9100-SERIES 9128EGetinge Disinfection AB · Disinfector, Medical Devices1 recall10 reports
- AMSCO 7053HPCorporation Steris Canada · Disinfector, medical devices1 recall10 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.