Skip to content

AMSCO 7053HP (REF FH19072)Recalls & reported problems

Corporation Steris Canada · The AMSCO 7053HP Washer-Disinfector with Steam Heat delivers a validated and cleared process for cleaning reusable da Vinci EndoWrist X/Xi, S/Si, and Single-Site instruments and staplers. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00724995176655), received between Sep 13, 2023 and Jul 2, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 19, 2025Open, Classified

    Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

    Z-0910-2026 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Use of Device Problem3
  • Insufficient Device Problem Information2
  • Adverse Event Without Identified Device or Use Problem1
  • Improper or Incorrect Procedure or Method1
  • Leak/Splash1

Patient problems

  • Bruise/Contusion3
  • No Clinical Signs, Symptoms or Conditions2
  • Abrasion1
  • Pain1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 2, 2026Insufficient Device Problem Information

    THE USER FACILITY REPORTED THAT AN EMPLOYEE OBTAINED AN INJURY WHILE OPERATING A CART TO THEIR AMSCO 7053HP WASHER/DISINFECTOR. THE EMPLOYEE SOUGHT MEDICAL TREATMENT HOWEVER, IT IS UNKNOWN THE TREATMENT THAT MAY HAVE BEEN ADMINISTERED.

    Patient problems: Pain

    MAUDE report 25695394 ↗
  • MalfunctionMay 1, 2026Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT AN EMPLOYEE OBTAINED A SHOULDER BRUISE WHILE ATTEMPTING TO PUSH A RACK INTO TO THEIR AMSCO 7053 HP WASHER/DISINFECTOR. IT IS UNKNOWN IF MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED.

    Patient problems: Bruise/Contusion

    MAUDE report 25060160 ↗
  • MalfunctionJan 16, 2026Improper or Incorrect Procedure or Method

    THE USER FACILITY REPORTED THAT AFTER AN EMPLOYEE LOADED A CONTAINER ONTO THE RACK IN THEIR AMSCO 7053HP WASHER/DISINFECTOR, THE HANDLE OF THE CONTAINER WAS "STICKING OUT". WHEN THE EMPLOYEE COMMANDED THE WASHER DOOR TO CLOSE, THE HANDLE CONTACTED THE DOOR SUBSEQUENTLY CAUSING THE INTERIOR TEMPERED GLASS TO BREAK. THE GLASS WAS BROKEN IN A SPIDER-LIKE FASHION BUT REMAINED INTACT. A USER FACILITY EMPLOYEE RECEIVED A SMALL CUT ON HER HAND WHEN SHE CONTACTED THE DOOR. NO MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED AND THE EMPLOYEE CONTINUED WORKING.

    Patient problems: Abrasion

    MAUDE report 24094388 ↗
  • MalfunctionSep 4, 2025Leak/Splash; Leak/Splash

    THE USER FACILITY REPORTED THAT WATER WAS LEAKING FROM THEIR AMSCO 7053 HP WASHER/DISINFECTOR ONTO THE FLOOR. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22965137 ↗
  • MalfunctionOct 3, 2024Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information

    THE USER FACILITY REPORTED THAT "BLACK FLAKES" WERE OBSERVED IN THE CHAMBER OF THEIR AMSCO 7053HP WASHER/DISINFECTOR. USER FACILITY PERSONNEL FOLLOWED THEIR POST-INSPECTION PROCEDURES ACCORDING TO THEIR GOOD HOSPITAL PRACTICES AND INSPECTED ALL INSTRUMENTS PRESENT IN THE CHAMBER. PROCEDURE DELAYS WERE REPORTED AS THE INSTRUMENTS HAD TO BE REPROCESSED. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20362471 ↗
  • MalfunctionJul 12, 2024Use of Device Problem; Use of Device Problem

    THE USER FACILITY REPORTED THAT AN EMPLOYEE BRUISED THEIR ARM WHILE OPERATING AN AMSCO 7053HP WASHER/DISINFECTOR. THE EMPLOYEE DID NOT SEEK PROFESSIONAL MEDICAL ATTENTION AND INSTEAD SELF-TREATED BY APPLYING AN ICE PACK TO THEIR ARM.

    Patient problems: Bruise/Contusion

    MAUDE report 19732679 ↗
  • MalfunctionSep 13, 2023Use of Device Problem; Use of Device Problem

    THE USER FACILITY REPORTED AN EMPLOYEE OBTAINED AN INJURY TO THEIR HAND WHILE OPERATING THEIR AMSCO 7053HP WASHER/DISINFECTOR. THE EMPLOYEE SELF-APPLIED AN ICE PACK, TOOK PAIN MEDICATION AND RETURNED TO WORK THE SAME DAY.

    Patient problems: Bruise/Contusion

    MAUDE report 17739049 ↗

Other Disinfector, medical devices devices

All Disinfector, medical devices devices →

Sources and updates