HarmonyAIR A-Series Lighting System (REF LAH02IN02)Recalls & reported problems

Steris Corporation · The A-Series Dual High Density Light with Single Monitor Mount 51-inch Single Monitor Mount, 43-inch Light, 35.5-inch Light, 20-inch Drop Tube is used with the HarmonyAIR A-Series Lighting System. · Class 2

All models of HarmonyAIR (240)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00724995186074), received between Jan 15, 2025 and Aug 8, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 15, 2026Open, Classified

    There is a potential for paint delamination (chipping/peeling) on the lighting system's light heads.

    Z-2448-2026 · Root cause: Process change control · Recall details · FDA record ↗

  2. May 15, 2026Open, Classified

    The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.

    Z-2449-2026 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Intermittent Loss of Power1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 8, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE LIGHTHEAD OF THEIR HARMONYAIR A-SERIES SURGICAL LIGHTING SYSTEM DETACHED FROM THE YOKE ASSEMBLY. THE LIGHTHEAD DID NOT FALL BUT REMAINED CONNECTED TO AND SUPPORTED BY THE WIRES WITHIN THE ASSEMBLY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY FOLLOWING A SHORT DELAY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22750730 ↗
  • MalfunctionJan 15, 2025Intermittent Loss of Power; Intermittent Loss of Power

    THE USER FACILITY REPORTED DURING A PATIENT PROCEDURE THEIR HARMONY AIR A-SERIES SURGICAL LIGHTING SYSTEM EXPERIENCED A LOSS OF CONTROL OF THE LIGHTS AND HAD TO RESTART THE SYSTEM AT THE WALL CONTROL RESULTING IN A SHORT PROCEDURE DELAY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21160925 ↗

Other Light, surgical, ceiling mounted devices