Skip to content

HarmonyAIR A-Series Lighting System (REF LAS02IN01)Recalls & reported problems

Steris Corporation · The A-Series Dual Light with Single Monitor Mount 51-inch Single Monitor Mount, 43-inch Light, 35.5-inch Light, 13-inch Drop Tube is used with the HarmonyAIR A-Series Lighting System. · Class 2

All models of HarmonyAIR (240)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00724995187989), received between Nov 16, 2023 and Jul 16, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 15, 2026Open, Classified

    The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.

    Z-2449-2026 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

  2. May 15, 2026Open, Classified

    There is a potential for paint delamination (chipping/peeling) on the lighting system's light heads.

    Z-2448-2026 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 16, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THEIR LIGHT HANDLE ASSEMBLY DETACHED FROM A HARMONYAIR A-SERIES LIGHT HEAD PRIOR TO A PROCEDURE. THE HANDLE ASSEMBLY DID NOT FALL BUT RATHER REMAINED CONNECTED TO AND SUPPORTED BY THE LIGHT HEAD WIRE HARNESS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22523363 ↗
  • MalfunctionNov 16, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE, ONE OF THE LIGHTHEADS FROM THEIR HARMONYAIR A-SERIES SURGICAL LIGHTING SYSTEM DISCONNECTED FROM THE YOKE ARM. USER FACILITY PERSONNEL WERE ABLE TO SECURE THE LIGHTHEAD AND MOVE IT OUT OF THE WAY. A PROCEDURE DELAY OCCURRED DUE TO THE RE-SETUP; THE PATIENT PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18151578 ↗

Other Light, surgical, ceiling mounted devices

All Light, surgical, ceiling mounted devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page