Skip to content

HarmonyAIR A-Series Lighting System (REF LAS01IN01)Recalls & reported problems

Steris Corporation · The A-Series Dual Light 43-inch Light, 35.5-inch Light, 13-inch Drop Tube is used with the HarmonyAIR A-Series Lighting System. · Class 2

All models of HarmonyAIR (240)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00724995188467), received between Oct 19, 2022 and Jul 10, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 15, 2026Class IIOpen, Classified

    The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.

    Z-2449-2026 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

  2. May 15, 2026Class IIOpen, Classified

    There is a potential for paint delamination (chipping/peeling) on the lighting system's light heads.

    Z-2448-2026 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Flaked1
  • Insufficient Device Problem Information1
  • Peeled/Delaminated1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 10, 2026Insufficient Device Problem Information

    THE USER FACILITY REPORTED VIA MEDWATCH REPORT # 2100490000-2026-8005 THAT, \"OR LIGHTS IN OR #2 ARE TURNING OFF AND ON DURING CASE. OVERALL SURGERY TIME EXTENDED DUE TO LACK OF LIGHT.\" NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25785536 ↗
  • MalfunctionMar 18, 2026Detachment of Device or Device Component

    THE USER FACILITY REPORTED THE LIGHT INTENSITY BUTTON FELL OFF THE HARMONYAIR A-SERIES SURGICAL LIGHTING SYSTEM DURING A PROCEDURE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24632240 ↗
  • MalfunctionJan 10, 2025Peeled/Delaminated; Peeled/Delaminated

    THE USER FACILITY REPORTED THAT PAINT ON THEIR HARMONYAIR A-SERIES DUAL LIGHT IS CHIPPING AND PEELING. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21129522 ↗
  • MalfunctionSep 19, 2024Use of Device Problem; Use of Device Problem

    THE USER FACILITY REPORTED THAT PAINT ON MULTIPLE LIGHTHEADS TO THEIR HARMONYAIR A-SERIES SURGICAL LIGHTING SYSTEMS ARE CHIPPING AND PEELING. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20259922 ↗
  • MalfunctionDec 28, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THE LIGHT HANDLE OF THEIR HARMONYAIR A-SERIES LIGHTING SYSTEM IS DETACHING AND FALLING DURING USE. NO INJURIES OR PROCEDURE DELAYS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16062815 ↗
  • MalfunctionOct 19, 2022Flaked; Flaked

    THE USER FACILITY REPORTED THAT PAINT ON MULTIPLE LIGHTHEADS TO THEIR DUAL HARMONY SURGICAL LIGHTING SYSTEMS IS BUBBLING AND CHIPPING. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15635650 ↗

FDA clearance and classification

The FDA regulates product code FSY (light, surgical, ceiling mounted) under 21 CFR 878.4580 as a Class II device. 21 CFR 878.4580 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the HarmonyAIR A-Series Lighting System

Has the HarmonyAIR A-Series Lighting System been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-2449-2026 (Class II), began on May 15, 2026: The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the HarmonyAIR A-Series Lighting System?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Oct 19, 2022 and Jul 10, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (2), “Flaked” (1) and “Peeled/Delaminated” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the HarmonyAIR A-Series Lighting System?

None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the HarmonyAIR A-Series Lighting System FDA approved?

Devices of this type (product code FSY) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Steris Corporation.

Who makes the HarmonyAIR A-Series Lighting System, and is it still sold?

The FDA's device database lists Steris Corporation as the company responsible for it (brand HarmonyAIR, model A-Series Lighting System). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Light, surgical, ceiling mounted devices

All Light, surgical, ceiling mounted devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page