Skip to content

HarmonyAIR A-Series Lighting System (REF LAS01IN01)Recalls & reported problems

Steris Corporation · The A-Series Dual Light 43-inch Light, 35.5-inch Light, 13-inch Drop Tube is used with the HarmonyAIR A-Series Lighting System. · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00724995188467), received between Oct 19, 2022 and Jul 10, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 15, 2026Open, Classified

    The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.

    Z-2449-2026 · Root cause: Material/Component Contamination · FDA record ↗

  2. May 15, 2026Open, Classified

    There is a potential for paint delamination (chipping/peeling) on the lighting system's light heads.

    Z-2448-2026 · Root cause: Process change control · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Flaked1
  • Insufficient Device Problem Information1
  • Peeled/Delaminated1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 10, 2026Insufficient Device Problem Information

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25785536 ↗
  • MalfunctionMar 18, 2026Detachment of Device or Device Component

    THE USER FACILITY REPORTED THE LIGHT INTENSITY BUTTON FELL OFF THE HARMONYAIR A-SERIES SURGICAL LIGHTING SYSTEM DURING A PROCEDURE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24632240 ↗
  • MalfunctionJan 10, 2025Peeled/Delaminated; Peeled/Delaminated

    THE USER FACILITY REPORTED THAT PAINT ON THEIR HARMONYAIR A-SERIES DUAL LIGHT IS CHIPPING AND PEELING. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21129522 ↗
  • MalfunctionSep 19, 2024Use of Device Problem; Use of Device Problem

    THE USER FACILITY REPORTED THAT PAINT ON MULTIPLE LIGHTHEADS TO THEIR HARMONYAIR A-SERIES SURGICAL LIGHTING SYSTEMS ARE CHIPPING AND PEELING. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20259922 ↗
  • MalfunctionDec 28, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THE LIGHT HANDLE OF THEIR HARMONYAIR A-SERIES LIGHTING SYSTEM IS DETACHING AND FALLING DURING USE. NO INJURIES OR PROCEDURE DELAYS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16062815 ↗
  • MalfunctionOct 19, 2022Flaked; Flaked

    THE USER FACILITY REPORTED THAT PAINT ON MULTIPLE LIGHTHEADS TO THEIR DUAL HARMONY SURGICAL LIGHTING SYSTEMS IS BUBBLING AND CHIPPING. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15635650 ↗

Other Light, surgical, ceiling mounted devices

All Light, surgical, ceiling mounted devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.