truFreeze 2000181Recalls & reported problems

US Endoscopy · 16 FR CryoDecompression Tube (CDT) - Low Durometer is used with the truFreeze Spray Cryotherapy System. · Class 2

All models of truFreeze (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00724995199210), received between May 30, 2023 and Feb 4, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem2
  • Break1
  • Difficult to Advance1
  • Insufficient Device Problem Information1
  • Mechanical Problem1

Patient problems

  • Laceration(s) of Esophagus2
  • No Clinical Signs, Symptoms or Conditions2
  • Crepitus1
  • Perforation1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 4, 2026Break

    THE USER FACILITY REPORTED THAT THE WIRE OF THEIR ACTIVE VENTING CRYO DECOMPRESSION TUBE "POKED" THE PATIENT DURING A PATIENT PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH ANOTHER ACTIVE VENTING CRYO DECOMPRESSION TUBE. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24261044 ↗
  • MalfunctionNov 21, 2025Mechanical Problem; Mechanical Problem

    THE USER FACILITY REPORTED THAT THE GUIDEWIRE OF THE ACTIVE VENTING CRYO DECOMPRESSION TUBE WENT THROUGH THE SHEATH CAUSING A SMALL CUT TO THE PATIENT'S ESOPHAGUS. NO MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED. THE PROCEDURE WAS CANCELLED AND RESCHEDULED FOR ANOTHER DAY.

    Patient problems: Laceration(s) of Esophagus

    MAUDE report 23622455 ↗
  • MalfunctionJul 25, 2025Use of Device Problem; Use of Device Problem

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE A STIFFENING WIRE, NORMALLY INTERNAL TO THE ACTIVE VENTING CRYO DECOMPRESSION TUBE, BECAME EXPOSED. AS A RESULT, THE PATIENT OBTAINED MUCOSAL DAMAGE IN THE UPPER ESOPHAGEAL SPHINCTER FROM THE EXPOSED WIRE. THE PATIENT WAS TREATED WITH A LIDOCAINE RINSE. THE PROCEDURE WAS DISCONTINUED AND RESCHEDULED AND COMPLETED SUCCESSFULLY AT A LATER DATE .

    Patient problems: Laceration(s) of Esophagus

    MAUDE report 22612538 ↗
  • MalfunctionApr 17, 2025Difficult to Advance; Difficult to Advance

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE TWO WIRES OF THEIR CRYO DECOMPRESSION TUBE WERE EXPOSED AND RESISTANCE WAS FELT BY USER FACILITY PERSONNEL WHILE OPERATING THE TUBE. USER FACILITY PERSONNEL ATTEMPTED TO STRAIGHTEN OUT THE WIRING OF THE TUBING BUT WERE UNSUCCESSFUL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY FOLLOWING A SHORT DELAY USING ANOTHER DEVICE. UPON COMPLETION OF THE PROCEDURE, CREPITUS WAS NOTICED ON THE BACK OF THE PATIENT'S NECK. MEDICAL TREATMENT WAS ADMINISTERED TO REMOVE THE CREPITUS; HOWEVER, IT IS UNKNOWN WHAT TYPE OF MEDICAL TREATMENT WAS ADMINISTERED.

    Patient problems: Crepitus

    MAUDE report 21857508 ↗
  • MalfunctionJun 8, 2024Use of Device Problem; Use of Device Problem

    THE USER FACILITY REPORTED THAT DURING A TRUFREEZE SPRAY CRYOTHERAPY PROCEDURE, WIRES INSIDE THEIR CRYO DECOMPRESSION TUBE LOOSENED AND BEGAN TO PULL OUT OF THE TUBE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19493596 ↗
  • MalfunctionMay 30, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    THE USER FACILITY REPORTED THAT A PERFORATION WAS DETECTED IN A PATIENT APPROXIMATELY THIRTEEN DAYS AFTER UNDERGOING A PROCEDURE WHICH INCLUDED USE OF TRUFREEZE CRYOSPRAY THERAPY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE PATIENT SOUGHT AND RECEIVED ADDITIONAL MEDICAL INTERVENTION (ANTIBIOTICS, STENT).

    Patient problems: Perforation

    MAUDE report 17021843 ↗

Other UNIT, CRYOSURGICAL, ACCESSORIES devices

All UNIT, CRYOSURGICAL, ACCESSORIES devices →