Skip to content

truFreeze 2000177Recalls & reported problems

US Endoscopy · Rapid AV Catheter Kit with 20 FR CDT includes 2.1mm Catheter (Straight Spray), Catheter Introducer, 20 Fr CryoDecompression Tube (CDT), Dual Lumen, Connector, Suction Tubing and is used with the truFreeze Spray Cryotherapy System. · Class 2

All models of truFreeze (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00724995199227), received between Mar 29, 2023 and Aug 29, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information5
  • Adverse Event Without Identified Device or Use Problem1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Pain2
  • Perforation1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 29, 2024Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information

    THE USER FACILITY REPORTED THAT DURING TWO PATIENT PROCEDURES, THE RAPID AV CATHETER BROKE IN THE CHANNEL OF THE SCOPE. FOLLOWING A DELAY, USER FACILITY PERSONNEL UTILIZED ANOTHER CATHETER TO COMPLETE THE PATIENT PROCEDURES. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20099833 ↗
  • MalfunctionJun 11, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    THE USER FACILITY REPORTED THAT THEIR TRUEFREEZE CONSOLE WAS UNABLE TO SCAN THE CATHETER POUCH SUCCESSFULLY. THE CUSTOMER COMPLETED PROCEDURE WITH ANOTHER CATHETER, SUBSEQUENTLY CAUSING THE PATIENT PROCEDURE TO BE DELAYED. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19516180 ↗
  • MalfunctionJul 7, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    THE USER FACILITY REPORTED THAT A PATIENT HAD A LINEAR TEAR NEAR THE MID-ESOPHAGUS A FEW WEEKS AFTER UNDERGOING A PROCEDURE TO TREAT A TUMOR WHICH INCLUDED USE OF TRUFREEZE CRYOSPRAY THERAPY. THE PATIENT HAD DEVELOPED PNEUMOMEDIASTINUM AND SOUGHT AND RECEIVED ADDITIONAL MEDICAL INTERVENTION.

    Patient problems: Pain

    MAUDE report 17276718 ↗
  • MalfunctionJun 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT A PERFORATION WAS DETECTED IN A PATIENT AFTER UNDERGOING A PROCEDURE WHICH INCLUDED USE OF TRUFREEZE CRYOSPRAY THERAPY. AN ADDITIONAL PROCEDURE WAS REQUIRED TO TREAT THE PATIENT.

    Patient problems: Perforation

    MAUDE report 17171018 ↗
  • MalfunctionMay 11, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    THE USER FACILITY REPORTED THAT A PATIENT WAS EXPERIENCING PAIN ONE WEEK AFTER UNDERGOING A PROCEDURE WHICH INCLUDED USE OF TRUFREEZE CRYOSPRAY THERAPY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE PATIENT SOUGHT AND RECEIVED ADDITIONAL MEDICAL INTERVENTION (X-RAY).

    Patient problems: Pain

    MAUDE report 16915309 ↗
  • MalfunctionMar 29, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE RAPID AV SPRAY CATHETER TO THEIR TRUFREEZE CONSOLE SPRAYED BLACK RESIDUE ONTO THE PATIENT'S TISSUE ON THE INITIAL SPRAY. THE PROCEDURE WAS COMPLETED WITH THE SAME CATHETER. NO REPORT OF INJURY. NO REPORT OF PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16638210 ↗

Other UNIT, CRYOSURGICAL, ACCESSORIES devices

All UNIT, CRYOSURGICAL, ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page