GORE DRYSEAL FLEX Introducer Sheath DSF1045Recalls & reported problems

W. L. Gore & Associates, Inc. · DRYSEAL FLEX INTRODUCER SHEATH 45cm 10Fr Hydrophilic Coating · Class 2

All models of GORE DRYSEAL FLEX Introducer Sheath (29)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00733132628094), received between Apr 21, 2025 and Apr 21, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Patient-Device Incompatibility” (1 report).

1 other recall cover devices cleared under the same FDA submission (K160254); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient-Device Incompatibility1

Patient problems

  • Perforation of Vessels1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 21, 2025Patient-Device Incompatibility; Patient-Device Incompatibility

    THE FOLLOWING WAS REPORTED TO GORE: PATIENT PRESENTED FOR TREATMENT DUE TO A POPLITEAL ARTERY ANEURYSM. AS REPORTED, THE PATIENT'S VESSEL WAS HEAVILY CALCIFIED. PRE-BALLOONING WAS NOT PERFORMED. ANTEGRADE ACCESS WAS MADE, A 10FR GORE® DRYSEAL FLEX INTRODUCER SHEATH WAS PLACED AND THE GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN DEVICE) WAS ADVANCED THROUGH THE SHEATH. HOWEVER, THE VIABAHN DEVICE DID NOT CROSS TO THE INTENDED TREATMENT SITE DUE TO THE CALCIFIED AREA AT THE PROXIMAL POPLITEAL ARTERY. DECISION WAS MADE TO WITHDRAW THE CONSTRAINED VIABAHN DEVICE BACK THROUGH THE SHEATH IN ORDER TO BALLOON THE TARGET LESION. HOWEVER, AS THE VIABAHN DEVICE WAS PULLED INTO THE SHEATH, THE DISTAL TIP GOT HUNG UP ON THE EDGE OF THE SHEATH. THE VIABAHN DEVICE WAS PULLED WITH EXTRA FORCE, CAUSING DEVICE DAMAGE. IT WAS REPORTED THE VIABAHN DEVICE WAS MANGLED DUE TO THE FORCE APPLIED. THE DEVICE AND SHEATH WERE REMOVED TOGETHER FROM THE PATIENT. AT THIS TIME IT WAS REALIZED THE COMMON FEMORAL ARTERY WAS PERFORATED. THE PATIENT WAS CONVERTED TO SURGICAL REPAIR. THE PATIENT WAS REPORTEDLY STABLE FOLLOWING THE SURGICAL INTERVENTION.

    Patient problems: Perforation of Vessels

    MAUDE report 21877665 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K160254. One submission often covers many models, so check the product and lot details in each record.

  • Z-1962-2020 · Nov 18, 2019 · Terminated
    Gore DrySeal Flex Introducer Sheath, Catalog #DSF1433, 33cm 14FR, Sterile. The responsible firm on the label is W. L. Gore & Associates, Inc., Flagstaff, AZ.

Other INTRODUCER, CATHETER devices