GORE DRYSEAL FLEX Introducer Sheath DSF1465Recalls & reported problems
W. L. Gore & Associates, Inc. · DRYSEAL FLEX INTRODUCER SHEATH 65cm 14Fr Hydrophilic Coating · Class 2
All models of GORE DRYSEAL FLEX Introducer Sheath (29)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00733132628308), received between Apr 1, 2022 and Dec 22, 2023 — classified as 1 death, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
1 other recall cover devices cleared under the same FDA submission (K160254); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Detachment of Device or Device Component1
- Difficult to Remove1
- Material Separation1
Patient problems
- Aneurysm1
- No Clinical Signs, Symptoms or Conditions1
- Stroke/CVA1
- Vascular Dissection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 22, 2023Break; Break
THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2023, THE PATIENT UNDERWENT AN ENDOVASCULAR PROCEDURE UTILIZING A GORE® DRYSEAL FLEX INTRODUCER SHEATH AS AN ACCESSORY. REPORTEDLY, NEAR THE END OF THE PROCEDURE, THE DRYSEAL SHEATH WAS BROUGHT UP AND THRU THE VENOUS BIFURCATION WHERE THEY WERE UP AND OVER WITH EVERYTHING BEING FINE AT THAT POINT. THEN THEY BROUGHT UP THE PENUMBRA DEVICE WHICH WENT STRAIGHT THRU TO INFERIOR VENA CAVA (IVC) AS THEY WERE TREATING A CLOT AND THOUGHT IT WENT THRU THE SIDE WALL OF THE SHEATH BUT DETERMINED THAT THE SHEATH SPLIT IN HALF AT THE BIFURCATION AND WAS BEING HELD BY THE SHEATH COIL. THE PHYSICIAN USED A BALLOON TO PULL THE TWO PIECES (HALVES) OF THE SHEATH TOGETHER BY BRIDGING THEM AS ONE UNIT AND PULLED IT OUT OF THE PATIENT SAFELY. THERE WAS NO INJURY REPORTED TO THE PATIENT AND PHYSICIAN USED ANOTHER GORE SHEATH TO COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WAS NO PATIENT ANATOMY (TORTUOUS) ISSUE AND PHYSICIAN WAS SURPRISED THAT THE SHEATH BROKE LIKE THAT. THE FIELD SALES ASSOCIATE ARRIVED NEAR THE COMPLETION OF THE PROCEDURE AND PATIENT TOLERATED THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18387266 ↗DeathApr 1, 2022Difficult to Remove; Material Separation; Detachment of Device or Device Component; Difficult to Remove; Material Separation; Detachment of Device or Device Component
ON (B)(6) 2022, THE PATIENT UNDERWENT AN EMERGENCY T- BRANCH THOROABDOMINAL ENDOVASCULAR PROCEDURE USING A GORE® VIABAHN® ENDOPROSTHESIS AND A GORE® DRYSEAL FLEX INTRODUCER SHEATH WITH THE ACCESS FROM THE AXILLARY ARTERY, THROUGH THE LEFT SUBCLAVIAN ARTERY (LSA) INTO THE DESCENDING AORTA. REPORTEDLY, THE SHEATH UNRAVELED INSIDE THE PATIENT AND BROKE WHEN THE PHYSICIAN TRIED TO TAKE IT OUT OF THE PATIENT. PART OF THE SHEATH REMAINED IN THE PATIENT AND EXTENDED FROM THE THORACIC AORTA TO THE ARCH. THE BROKEN PART OF THE SHEATH WAS SNARED AND PULLED OUT THROUGH THE SHEATH IN THE GROIN. IN ADDITION, THE LSA TORE, AND IT WAS REPAIRED BY A 10 X 4 VIABAHN. THE PATIENT ALSO DEVELOPED WHAT THE PHYSICIAN BELIEVED INITIALLY WAS A RETROGRADE DISSECTION FROM THE GRAFT TO THE LSA BUT UPON FURTHER REVIEW THE BROKEN SHEATH/ WIRE BRAID MAY HAVE CAUSED THE DISSECTION. THE PROCEDURE TOOK PROLONGED TIME. THE PROBLEM WAS DETECTED BY X-RAY BEING USED DURING THE ENDOVASCULAR PROCEDURE. THE PROCEDURE WAS COMPLETED BY STENTING A 9MM ATRIUM IN THE RENAL ARTERY AND WITH A DIFFERENT NON-GORE SHEATH. IT WAS REPORTED THAT THE DEVICE WAS PRE-TESTED PRIOR TO USE AND NO DAMAGE WAS NOTICED. THE SHEATH WAS USED ACCO
Patient problems: Aneurysm; Stroke/CVA; Vascular Dissection
MAUDE report 13972941 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K160254. One submission often covers many models, so check the product and lot details in each record.
- Z-1962-2020 · Nov 18, 2019 · TerminatedGore DrySeal Flex Introducer Sheath, Catalog #DSF1433, 33cm 14FR, Sterile. The responsible firm on the label is W. L. Gore & Associates, Inc., Flagstaff, AZ.
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page