GORE VIABAHN VBX Balloon Expandable Endo BXA075901ARecalls & reported problems
W. L. Gore & Associates, Inc. · VIABAHN BX BALLOON EXP ENDO 7mmX59mm 7Fr 80cmCath Heparin · Class 3 · Implant
All models of GORE VIABAHN VBX Balloon Expandable Endo (111)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00733132637287), received between Apr 26, 2023 and Apr 9, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Collapse” (1 report).
1 other recall cover devices cleared under the same FDA submission (P160021); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Collapse1
- Detachment of Device or Device Component1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Thrombosis/Thrombus1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 9, 2024Collapse; Obstruction of Flow; Collapse; Obstruction of Flow
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2024, THE PATIENT WAS IMPLANTED WITH A GORE® EXCLUDER® AAA ENDOPROSTHESIS TO TREAT A RUPTURED ABDOMINAL AORTIC ANEURYSM. GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICES) WERE IMPLANTED IN THE SUPERIOR MESENTERIC ARTERY (SMA) AND THE RIGHT RENAL ARTERY AS SNORKELS DURING THE PROCEDURE. THE PHYSICIAN REPORTED ON THE EVENING OF (B)(6) 2024, THAT THE VBX STENT GRAFT WAS THROMBOSED. IMMEDIATELY, AN ENDOVASCULAR REINTERVENTION OCCURRED AND REPORTEDLY THE PHYSICIAN WAS UNABLE TO ADVANCE THROUGH THE SMA AND VBX STENT GRAFT. THE PROCEDURE WAS ABORTED, AND THE PATIENT TOLERATED THE PROCEDURE. THE PHYSICIAN SUSPECTED THE VBX DEVICE MAY HAVE THROMBOSED BECAUSE THE EYE OF THE TIGER TECHNIQUE WAS PERFORMED INCORRECTLY, CONSEQUENTLY THE GORE® EXCLUDER® AAA ENDOPROSTHESIS MAY HAVE COMPRESSED THE VBX STENT GRAFT. THE PATIENT WILL CONTINUE TO BE MONITORED. NONE OF THE GORE DEVICES WERE EXPLANTED, AND NO ADDITIONAL GORE DEVICES WERE IMPLANTED.
Patient problems: Thrombosis/Thrombus
MAUDE report 19075891 ↗MalfunctionApr 26, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ON (B)(6) 2023, THE PATIENT WAS BEING TREATED FOR A STENOTIC LESION IN THE ILIAC ARTERY USING A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS. THE DEVICE WAS INSERTED SUCCESSFULLY WITH NO ISSUE THROUGH A 7 FR TERUMO SHEATH. THE DEVICE WAS UNABLE TO BE ADVANCED DUE TO STENOSIS. THE VESSEL HAD NOT BEEN DILATED. THE DEVICE WAS ATTEMPTED TO BE WITHDRAWN THROUGH THE SHEATH AND THE STENT GRAFT DETACHED FROM THE DELIVERY SYSTEM CATHETER. THE DEVICE WAS REMOVED WITH A SNARE CATHETER FROM THE PATIENT. THE PATIENTS VESSEL WAS DILATED AND A DIFFERENT GORE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16820642 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160021. One submission often covers many models, so check the product and lot details in each record.
- Z-0077-2022 · Sep 7, 2021 · Open, ClassifiedHeparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diame
Other Iliac Covered Stent, Arterial devices
- GORE VIABAHN VBX Balloon Expandable Endo BXB073902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths19 reports
- iCast Covered Stent System, 6mmX16mmX120cm 42616Atrium Medical Corporation · Iliac Covered Stent, Arterial1 recall15 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB057902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB067902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB075902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall11 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB095902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB077902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB085902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page