GORE VIABAHN VBX Balloon Expandable Endo BXA062901ARecalls & reported problems
W. L. Gore & Associates, Inc. · VIABAHN BX BALLOON EXP ENDO 6mmX29mm 7Fr 80cmCath Heparin · Class 3 · Implant
All models of GORE VIABAHN VBX Balloon Expandable Endo (111)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00733132637461), received between May 29, 2019 and Aug 29, 2019 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
1 other recall cover devices cleared under the same FDA submission (P160021); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated1
- Entrapment of Device1
Patient problems
- Aneurysm1
- Perforation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 29, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE DOCTOR REPORTED THE FOLLOWING TO THE GORE FSA: THE PATIENT PRESENTED FOR A MODIFIED FENESTRATED AORTIC CASE. AFTER THE FENESTRATION WAS CREATED AND THE WIRE PLACED IN THE RIGHT RENAL, THE DOCTOR WAS UNABLE TO ADVANCE THE GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX), USING A 7FR SHEATH, THROUGH THE FENESTRATED HOLE. THE DOCTOR PULLED THE VBX BACK INTO THE SHEATH AND OBSERVED THAT THE STENT HAD DISLODGED FROM THE BALLOON. THE DISLODGED STENT WAS JAILED ABOVE THE CELIAC ARTERY BEHIND THE TEVAR.
Patient problems: Aneurysm
MAUDE report 8946050 ↗InjuryMay 29, 2019Entrapment of Device; Entrapment of Device
ADDITIONAL INFORMATION FROM THE PHYSICIAN WAS OBTAINED. THE BALLOON CATHETER MAY HAVE ENCOUNTERED AN OVATION AAA DEVICE STRUT, CAUSING DAMAGE TO THE VBX DEVICE DURING WITHDRAWAL. THE GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (LOT #18674112) THAT WAS ADVANCED FROM THE ARM THROUGH THE 7FR COOK SHEATH WAS THE ONLY ONE THAT ENCOUNTERED ISSUES DURING THE WITHDRAWAL OF DELIVERY CATHETER. THE FOUR IMPLANTED VBX DEVICES HAD DIFFERENT ENTRY POINTS. THEY WERE NOT ALL ADVANCED FROM THE SAME ENTRY POINTS AS STATED IN ORIGINAL DESCRIPTION.
Patient problems: Perforation
MAUDE report 8651792 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160021. One submission often covers many models, so check the product and lot details in each record.
- Z-0077-2022 · Sep 7, 2021 · Open, ClassifiedHeparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diame
Other Iliac Covered Stent, Arterial devices
- GORE VIABAHN VBX Balloon Expandable Endo BXB073902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths19 reports
- iCast Covered Stent System, 6mmX16mmX120cm 42616Atrium Medical Corporation · Iliac Covered Stent, Arterial1 recall15 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB057902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB067902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB075902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall11 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB095902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB077902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB085902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.