GORE VIABAHN VBX Balloon Expandable Endo BXA112902ARecalls & reported problems
W. L. Gore & Associates, Inc. · VIABAHN BX BALLOON EXP ENDO 11mmX29mm 8Fr135cmCath Heparin · Class 3 · Implant
All models of GORE VIABAHN VBX Balloon Expandable Endo (111)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00733132637706), received between May 24, 2019 and May 6, 2022 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Material Separation” (2 reports).
1 other recall cover devices cleared under the same FDA submission (P160021); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Separation2
- Difficult or Delayed Separation1
- Improper or Incorrect Procedure or Method1
- Inflation Problem1
- Malposition of Device1
- Material Split, Cut or Torn1
- No Flow1
- Patient-Device Incompatibility1
Patient problems
- Aneurysm1
- Blood Loss1
- Calcium Deposits/Calcification1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
- Obstruction/Occlusion1
- Occlusion1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryMay 6, 2022Inflation Problem; Malposition of Device; Premature Separation; Inflation Problem; Malposition of Device; Premature Separation
THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2022, A PATIENT UNDERWENT TREATMENT FOR THE RIGHT COMMON FEMORAL ARTERY UTILIZING A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX) DEVICE DURING A TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVR) PROCEDURE. IT WAS REPORTED THAT THE TECHNICIAN PREPARED THE DEVICE WITH A SHORT, NEGATIVE PRESSURE ON THE BALLOON AND THEN THE VBX DEVICE WAS ADVANCED TO THE RIGHT COMMON FEMORAL ARTERY FROM THE CONTRA-LATERAL SIDE USING AN 8 FR ANSEL SHEATH. THE PHYSICIAN REPORTED THAT THE VBX DEVICE WAS DIFFICULT TO ADVANCE THROUGH THE SHEATH, ALTHOUGH IT WAS ULTIMATELY ABLE TO BE ADVANCED TO THE INTENDED LOCATION. UPON INFLATION, THE BALLOON ON THE STENT REPORTEDLY INFLATED ON THE PROXIMAL END, BUT NOT THE DISTAL END. THE STENT THEN DISLODGED FROM THE BALLOON CATHETER WITH ENOUGH FORCE TO PUSH IT INTO THE DISTAL THIRD OF THE SUPERFICIAL FEMORAL ARTERY (SFA). AN ATTEMPT WAS MADE TO USE A SEPARATE BALLOON (ARMADA 14MM X 20MM), BUT THE STENT WAS UNABLE TO BE FULLY DEPLOYED. THE PATIENT WAS THE TAKEN TO SURGERY AND AN OPEN CUT-DOWN WAS PERFORMED TO FIX THE LEAK IN THE COMMON FEMORAL ARTERY. A SHEATH AND GUIDEWIRE WERE THEN USED TO ACCESS THE STENT IN TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14310177 ↗InjuryApr 22, 2021No Flow; No Flow
'OUTCOMES OF SMALL RENAL ARTERY TARGETS IN PATIENTS TREATED BY FENESTRATED- BRANCHED ENDOVASCULAR AORTIC REPAIR' JUSSI M. KÄRKKÄINEN, EMANUEL R. TENORIO, KEOUNA PATHER, BERNARDO C. MENDES, THANILA A. MACEDO, JEAN WIGHAM, ALISA DIDERRICH, GUSTAVO S. ODERICH. THE AIM WAS TO EVALUATE RENAL RELATED OUTCOMES IN PATIENTS WHO HAD INCORPORATION OF A SMALL (<4.0 MM) RENAL ARTERY (RA) DURING FENESTRATED-BRANCHED ENDOVASCULAR AORTIC REPAIR (F-BEVAR). A TOTAL OF 215 CONSECUTIVE PATIENTS ENROLLED IN A PROSPECTIVE F-BEVAR TRIAL WERE REVIEWED. THE KAPLANE-MEIER ESTIMATES OF ONE AND TWO YEAR RA RELATED OUTCOMES ARE PRESENTED IN TABLE 4. AT ONE YEAR, PRIMARY PATENCY WAS 79 ± 9% IN THE STUDY GROUP AND 94 ± 1% IN THE CONTROL GROUP (P < .001); SECONDARY PATENCY WAS 84 ± 8% VS. 97 ± 1% (P < .001), RESPECTIVELY (FIG. 3A). FREEDOM FROM RA INSTABILITY WAS 79 ± 9% VS. 93 ± 2% AT ONE YEAR (P ¼ .005, FIG. 3B), AND FREEDOM FROM RENAL FUNCTION DETERIORATION WAS 90 ± 7% VS. 90 ± 3% (P ¼ .09), RESPECTIVELY. THERE WERE NO DIFFERENCES IN FREEDOM FROM REINTERVENTION AND SURVIVAL BETWEEN THE GROUPS. NO DEVICE ALLEGATIONS OF DEFICIENCIES SPECIFIC FOR GORE DEVICES.
Patient problems: Obstruction/Occlusion
MAUDE report 11709214 ↗MalfunctionDec 9, 2020Material Separation; Material Separation
ON (B)(6) 2020, THE PATIENT WAS UNDERGOING A TRANSCAROTID CUT DOWN FLOW REVERSAL PROCEDURE IN THE INNOMINATE ARTERY USING A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS. THE DEVICE WAS ADVANCED ANTEGRADE AND DEPLOYED WITH NO ISSUE. AFTER THE SUCCESSFUL DEPLOYMENT, THE CATHETER WAS BEING WITHDRAWN THROUGH THE SHEATH. IT WAS OBSERVED THE BALLOON SEPARATED FROM THE CATHETER. THE BALLOON REMAINED INSIDE THE PORTION OF THE SHEATH THAT WAS LOCATED INSIDE OF THE PATIENT. THE BALLOON WAS WITHDRAWN FROM THE PATIENT ALONG WITH THE SHEATH. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10977163 ↗MalfunctionNov 2, 2020Material Separation; Material Separation
ON (B)(6) 2020, A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS(VBX) WAS BEING PLACED AS A INNOMINATE STENT IN CONJUNCTION WITH TRANSCAROTID ARTERY RE-VASCULARIZATION (TCAR) PROCEDURE. DURING THE ADVANCEMENT OF THE DEVICE THE PHYSICIAN HAD TO USE EXCESSIVE FORCE TO GET THE DEVICE INTO PLACE. AFTER THE DEVICE WAS SUCCESSFULLY DEPLOYED, THE PHYSICIAN WAS REMOVING THE DEVICE CATHETER THROUGH THE 8FR TCAR INTRODUCER SHEATH, AND FELT RESISTANCE. AFTER THE RESISTANCE WAS FELT HE CONTINUED TO WITHDRAW THE CATHETER, AND THE CATHETER BROKE. THE PHYSICIAN REMOVED THE REMAINDER OF THE CATHETER AND SHEATH IN TANDEM. WHEN IT WAS REMOVED, IT WAS SUSPECTED THAT THE BALLOON PORTION OF THE CATHETER HUNG UP ON THE SHEATH TIP. PATIENT DID NOT SUFFER ANY ADVERSE EFFECTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10775046 ↗InjuryDec 3, 2019Patient-Device Incompatibility; Patient-Device Incompatibility
ON (B)(6) 2019 A PATIENT WAS UNDERGOING TREATMENT OF A COMMON ILIAC ARTERY STENOTIC LESION WITH GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESES. THERE WAS ALSO A DISTAL AORTIC ANEURYSM, BUT IT WAS DECIDED THIS ANEURYSM WOULD NOT BE TREATED AT THIS TIME. FOLLOWING ADVANCEMENT OF BOTH VBX DEVICES IN KISSING TECHNIQUE, THE DEVICES WERE DEPLOYED WITH SIMULTANEOUS BALLOONING, SUCCESSFULLY. THE BALLOONS WERE DEFLATED AND REMOVED. THE ANESTHESIOLOGIST NOTICED A DROP IN THE PATIENT'S BLOOD PRESSURE SO THE BLOOD PRESSURE CUFF WAS CHECKED AND REAPPLIED. HOWEVER THE BLOOD PRESSURE READING REMAINED LOW. A PIGTAIL CATHETER WAS THEN ADVANCED AND AORTIC RUPTURE WAS REALIZED. AN OCCLUSION BALLOON WAS ADVANCED FOR HEMORRHAGE CONTROL AND A COOK MAIN BODY DEVICE WAS IMPLANTED TO TREAT THE AORTIC RUPTURE. THE LEFT LIMB OF THE MAIN BODY WAS TREATED SUCCESSFULLY, PRESERVING THE LEFT ILIAC ARTERY. HOWEVER, THE RIGHT SIDE COULD NOT BE TREATED. THE RIGHT SIDE WAS PLUGGED WITH AN AMPLATZ DEVICE. FOR THIS REASON, A FEMORAL TO FEMORAL ARTERY BYPASS WAS NECESSARY AND THIS WAS CARRIED OUT BY ANOTHER SURGEON. THE PATIENT REQUIRED REPLACEMENT BLOOD PRODUCTS, HOWEVER WAS STABLE AT THE END OF THE PROCEDURE.
Patient problems: Aneurysm; Blood Loss
MAUDE report 9405105 ↗MalfunctionMay 24, 2019Improper or Incorrect Procedure or Method; Material Split, Cut or Torn; Difficult or Delayed Separation; Improper or Incorrect Procedure or Method; Material Split, Cut or Torn; Difficult or Delayed Separation
THE FOLLOWING WAS REPORTED TO THE GORE FSA BY THE DOCTOR: THE PATIENT PRESENTED FOR A FLOW REVERSAL CASE. THE GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS WAS ADVANCED THROUGH THE 8FR SILK ROAD MEDICAL SHEATH AND DEPLOYED AS EXPECTED. AS THE BALLOON WAS BEING WITHDRAWN, IT GOT STUCK IN THE INTRODUCER SHEATH AND CAME OFF THE CATHETER WHEN THE DOCTOR PULLED BACK (AFTER IT WAS STUCK IN THE SHEATH). THE INTRODUCER AND CATHETER WERE REMOVED TOGETHER WITH NO FURTHER INCIDENT. THE PATIENT WAS REPORTED TO BE FINE. THE CATHETER WAS DISCARDED AT THE HOSPITAL.
Patient problems: Calcium Deposits/Calcification; Occlusion
MAUDE report 8640991 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160021. One submission often covers many models, so check the product and lot details in each record.
- Z-0077-2022 · Sep 7, 2021 · Open, ClassifiedHeparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diame
Other Iliac Covered Stent, Arterial devices
- GORE VIABAHN VBX Balloon Expandable Endo BXB073902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths19 reports
- iCast Covered Stent System, 6mmX16mmX120cm 42616Atrium Medical Corporation · Iliac Covered Stent, Arterial1 recall15 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB057902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB067902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB075902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall11 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB095902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB077902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB085902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page