GORE VIABAHN VBX Balloon Expandable Endo BXA087902ARecalls & reported problems

W. L. Gore & Associates, Inc. · VIABAHN BX BALLOON EXP ENDO 8mmX79mm 8Fr 135cmCath Heparin · Class 3 · Implant

All models of GORE VIABAHN VBX Balloon Expandable Endo (111)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00733132637799), received between Nov 9, 2020 and Mar 25, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Device Stenosis” (1 report).

1 other recall cover devices cleared under the same FDA submission (P160021); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1
  • Device Stenosis1
  • Material Perforation1
  • Separation Problem1

Patient problems

  • Clinical Endoleak2
  • No Consequences Or Impact To Patient1
  • Stenosis1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2024, A PATIENT UNDERWENT A THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) PROCEDURE. DURING THIS PROCEDURE, GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) WERE USED FOR BRANCH DEVICES. AS REPORTED, THE PROCEDURE WENT WELL WITH TECHNICAL SUCCESS. ON (B)(6) 2024, THE PHYSICIAN REVIEWED THE POST PROCEDURE IMAGES AGAIN AND NOTICED A POSSIBLE SMALL ENDOLEAK. IT APPEARED THE DEVICE PURCHASE IN THE SUPERIOR MESENTERIC ARTERY (SMA) ARTERY MAY NOT HAVE BEEN SUFFICIENT. ON (B)(6) 2024, AN INTERVENTION WAS PERFORMED. THE VBX DEVICE WAS EXTENDED WITH A NEW 8 X 59 VBX DEVICE THE INTERVENTION HAD A GOOD OUTCOME AND THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: Clinical Endoleak

    MAUDE report 21691686 ↗
  • InjuryFeb 28, 2023Material Perforation; Appropriate Device Problem Term/Code Not Available; Material Perforation; Appropriate Device Problem Term/Code Not Available

    ON (B)(6) 2022, A STUDY PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF A THORACOABDOMINAL AORTIC ANEURYSM IN THE GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) CLINICAL TRIAL. THE VBX STENTS WERE IMPLANTED AS BRANCH COMPONENTS IN THE CELIAC, SUPERIOR MESENTERIC ARTERY (SMA), LEFT AND RIGHT RENAL ARTERIES. DURING PATIENT FOLLOW-UP VISIT ON (B)(6) 2022, CTA SHOWED A BRISK ENDOLEAK, WHICH LOOKED TO BE ASSOCIATED WITH VBX STENT IN THE SMA AND LIKELY REPRESENTED A TYPE 1C AS THERE WAS LITTLE EXTENSION OF THE VBX STENT INTO THE SMA. AS REPORTED, THE PHYSICIAN STATED THIS WAS A DEFINITIVE LEAK FROM THE VBX MID STENT, SUSPECTED TO BE FROM A FABRIC TEAR OF THE VBX STENT. PATIENT RETURNED FOR TREATMENT ON (B)(6) 2023. ANGIOGRAPHIC FINDINGS SHOWED NO ENDOLEAK DEMONSTRATED IN THE SMA. HOWEVER, GIVEN THE MINIMAL STENT OVERLAP INTO THE SMA, THE VBX STENT WAS EXTENDED WITH A 7X 59MM GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS. COMPLETION ARTERIOGRAM DEMONSTRATED NO LEAK. THE ENDOLEAK WAS CONSIDERED RESOLVED. THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: Clinical Endoleak

    MAUDE report 16453315 ↗
  • InjuryOct 31, 2022Device Stenosis; Device Stenosis

    THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2021, A STUDY PATIENT UNDERWENT A AAA PROCEDURE USING A FENESTRATED DEVICE. SEVERAL GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) WERE IMPLANTED AS SIDE BRANCHES. ONE VSX DEVICE WAS IMPLANTED IN THE SUPERIOR MESENTERIC ARTERY AS FIRST DISTAL EXTENSION. ON (B)(6) 2022, STENOSIS IN THE SUPERIOR MESENTERIC ARTERY WAS OBSERVED. ON (B)(6) 2022, A BALLOON ANGIOPLASTY AND STENTING WERE COMPLETED DUE TO THE STENOSIS. PATENCY WAS RESTORED. THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: Stenosis

    MAUDE report 15703255 ↗
  • InjuryNov 9, 2020Separation Problem; Separation Problem

    ON (B)(6) 2020 A PATIENT UNDERWENT AN ENDOVASCULAR PROCEDURE IN THE (B)(6) STUDY. GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESES WERE UTILIZED IN THE BRANCH VESSELS. ON (B)(6) 2020 IMAGING REVEALED THAT THE TWO VBX DEVICES WHICH WERE DEPLOYED IN THE SUPERIOR MESENTERIC ARTERY HAD BECOME SEPARATED. ON (B)(6) 2020 A REINTERVENTION TO TREAT THE TYPE III ENDOLEAK WAS PERFORMED. AN ADDITIONAL VBX WAS DEPLOYED TO BRIDGE THE SEPARATION AND THE ENDOLEAK WAS CONSIDERED RESOLVED. THE PATIENT TOLERATED THE PROCEDURE . ON (B)(6) 2020 THE PATIENT WAS DISCHARGED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10813680 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160021. One submission often covers many models, so check the product and lot details in each record.

  • Z-0077-2022 · Sep 7, 2021 · Open, Classified
    Heparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diame

Other Iliac Covered Stent, Arterial devices

All Iliac Covered Stent, Arterial devices →