GORE VIABAHN VBX Balloon Expandable Endo BXB052902ARecalls & reported problems
W. L. Gore & Associates, Inc. · BX2 HEP REDUCED PROFILE BX2 5MMX29MM 6Fr 135CM Cath · Class 3 · Implant
All models of GORE VIABAHN VBX Balloon Expandable Endo (111)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00733132658305), received between Jan 27, 2025 and Jan 27, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Complete Blockage” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (P160021); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 1, 2024Class IIOpen, Classified
Due to packaging defects, sterility assurance and heparin activity may be compromised.
Z-2945-2024 · Root cause: Package design/selection · Recall details · FDA record ↗
Most reported problems
Device problems
- Complete Blockage1
Patient problems
- Obstruction/Occlusion1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJan 27, 2025Complete Blockage; Complete Blockage
THE FOLLOWING WAS RECEIVED FROM A STUDY: ON (B)(6) 2023, A STUDY PATIENT WAS IMPLANTED WITH A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE DEVICE) FOR AORTIC ANEURYSM TREATMENT. DURING THIS PROCEDURE, GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS WERE USED AS BRANCH DEVICES IN THE VISCERAL ARTERIES. TWO VBX DEVICES WERE IMPLANTED IN THE LEFT ARTERY DURING THE INDEX PROCEDURE. ON (B)(6) 2024, STENOSIS WAS OBSERVED IN THE LEFT RENAL ARTERY. ON (B)(6) 2024, BALLOON ANGIOPLASTY AND AN ADDITIONAL VBX DEVICE WAS IMPLANTED IN THE LEFT RENAL ARTERY. AS REPORTED, THE REINTERVENTION WAS PERFORMED TO PREVENT EVENTUAL LOSS OF PATENCY. AT THE PATIENT'S 12 MONTH FOLLOW-UP, CTA IMAGING AND OBSERVATION NOTED THE LEFT RENAL ARTERY SIDE BRANCH COMPONENTS HAD LOST PATENCY. IT WAS ASSUMED ALL THREE VBX DEVICES WERE OCCLUDED. NO INTERVENTION WAS PERFORMED OR PLANNED AT THIS TIME.
Patient problems: Obstruction/Occlusion
MAUDE report 21244640 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160021. One submission often covers many models, so check the product and lot details in each record.
- Z-0077-2022 · Sep 7, 2021 · Open, ClassifiedHeparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diame
Other Iliac covered stent, arterial devices
- GORE VIABAHN VBX Balloon Expandable Endo BXB073902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths19 reports
- iCast Covered Stent System, 6mmX16mmX120cm 42616Atrium Medical Corporation · Iliac Covered Stent, Arterial1 recall15 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB057902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB067902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB075902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall11 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB095902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB077902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB085902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page