Skip to content

GORE VIABAHN VBX Balloon Expandable Endo BXBL082902ARecalls & reported problems

W. L. Gore & Associates, Inc. · BX2 HEP REDUCED PROFILE BX2 8MMX29MM 7Fr 135CM Cath · Class 3 · Implant

All models of GORE VIABAHN VBX Balloon Expandable Endo (111)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00733132658596), received between Apr 15, 2024 and Apr 15, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (P160021); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 1, 2024Class IIOpen, Classified

    Due to packaging defects, sterility assurance and heparin activity may be compromised.

    Z-2945-2024 · Root cause: Package design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 15, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2024, A REDUCED PROFILE GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) WAS IMPLANTED IN THE RENAL ARTERY WITH A COOK FENESTRATED AORTIC GRAFT TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). REPORTEDLY, THE PROCEDURE PRESENTED A CHALLENGE FOR THE PHYSICIAN WITH THE APPROACHED ANGLE ACCESSED FROM THE BRACHIAL ARTERY; THROUGH A COOK FENESTRATED AORTIC GRAFT; TO THE IMPLANT SITE OF THE RENAL ARTERY. THE VBX DEVICE WAS ADVANCED OVER A 0.035¿ TERUMO ADVANTAGE GLIDEWIRE THROUGH A 7.5 F DESTINO STEERABLE SHEATH WITHOUT DIFFICULTY. IT WAS NOTED THE INTRODUCER SHEATH WAS PARKED AT THE OPENING OF THE RENAL ARTERY AND THE VBX DEVICE WAS THEN ADVANCED ONLY ON THE GUIDEWIRE. THE PHYSICIAN NEEDED TO REPOSITION THE VBX DEVICE AND TO DO SO WITHDREW THE VBX DEVICE DELIVERY CATHETER BACK SLIGHTLY INTO THE INTRODUCER SHEATH. CONSEQUENTLY, THE VBX STENT GRAFT CAME OFF OF THE DELIVERY CATHETER AND DISLODGED. THE PHYSICIAN REPORTEDLY REMOVED THE VBX DEVICE DELIVERY CATHETER WITHOUT DIFFICULTY AND WAS ABLE TO USE A PTA BALLOON TO IMPLANT THE VBX STENT GRAFT WHERE IT DISLODGED. THERE WAS NO BRANCH VESSEL COVERAGE. AN ADDITIONAL REDUC

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19110641 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160021. One submission often covers many models, so check the product and lot details in each record.

  • Z-0077-2022 · Sep 7, 2021 · Open, Classified
    Heparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diame

Other Iliac covered stent, arterial devices

All Iliac covered stent, arterial devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page