GORE VIABAHN VBX Balloon Expandable Endo BXB072901ARecalls & reported problems
W. L. Gore & Associates, Inc. · BX2 HEP REDUCED PROFILE BX2 7MMX29MM 6Fr 80CM Cath · Class 3 · Implant
All models of GORE VIABAHN VBX Balloon Expandable Endo (111)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00733132658619), received between Jun 12, 2025 and Jun 12, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (P160021); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 1, 2024Class IIOpen, Classified
Due to packaging defects, sterility assurance and heparin activity may be compromised.
Z-2945-2024 · Root cause: Package design/selection · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Rupture2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 12, 2025Material Rupture; Material Rupture
THE FOLLOWING WAS REPORTED TO GORE: ON ()() 2025, A PATIENT WAS TREATED WITH A 4 VESSEL PHYSICIAN MODIFIED ENDOGRAFT. DURING THIS PROCEDURE, A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) WAS IMPLANTED IN THE RIGHT RENAL ARTERY (RRA). A SECOND VBX DEVICE WAS DEPLOYED IN THE RRA BECAUSE THE FIRST VBX DEVICE DID NOT EXTEND FAR ENOUGH INTO THE AORTA. THE PHYSICIAN REPORTED THAT DURING THE DEVICE DEPLOYMENT, BOTH BALLOONS RUPTURED BEFORE REACHING NOMINAL PRESSURE ON THE IN DEFLATOR DEVICE. AS REPORTED, THE BALLOON STARTED INFLATING FROM BOTH ENDS BUT RUPTURED BEFORE FULL EXPANSION. THE PHYSICIAN USED A MUSTANG¿ BALLOON DILATATION CATHETER TO BALLOON EACH VBX DEVICE TO FULL EXPANSION. BOTH VBX CATHETERS WERE REMOVED INTACT. THERE WAS NO ADVERSE OUTCOME FOR THE PATIENT. THE PHYSICIAN REPORTED HE DOES NOT KNOW WHY THE BALLOONS RUPTURED IN THE SAME ANATOMICAL LOCATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22196121 ↗MalfunctionJun 12, 2025Material Rupture; Material Rupture
THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2025, A PATIENT WAS TREATED WITH A 4 VESSEL PHYSICIAN MODIFIED ENDOGRAFT. DURING THIS PROCEDURE, A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) WAS IMPLANTED IN THE RIGHT RENAL ARTERY (RRA). A SECOND VBX DEVICE WAS DEPLOYED IN THE RRA BECAUSE THE FIRST VBX DEVICE DID NOT EXTEND FAR ENOUGH INTO THE AORTA. IT WAS REPORTED BY THE PHYSICIAN THAT TWO BALLOONS RUPTURED BEFORE REACHING NOMINAL PRESSURE ON THE IN DEFLATOR DEVICE. THE BALLOON STARTED INFLATING FROM BOTH ENDS BUT RUPTURED BEFORE FULL EXPANSION. THE VBX CATHETERS WERE REMOVED INTACT. THE PHYSICIAN THEN INSERTED A MUSTANG¿ BALLOON DILATATION CATHETER AND BALLOONED THE VBX DEVICES TO FULL EXPANSION. THERE WAS NO ADVERSE OUTCOME FOR THE PATIENT. THE PHYSICIAN REPORTED HE DOES NOT KNOW WHY THE BALLOONS RUPTURED IN THE SAME ANATOMICAL LOCATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22195865 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160021. One submission often covers many models, so check the product and lot details in each record.
- Z-0077-2022 · Sep 7, 2021 · Open, ClassifiedHeparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diame
Other Iliac covered stent, arterial devices
- GORE VIABAHN VBX Balloon Expandable Endo BXB073902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths19 reports
- iCast Covered Stent System, 6mmX16mmX120cm 42616Atrium Medical Corporation · Iliac Covered Stent, Arterial1 recall15 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB057902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB067902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB075902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall11 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB095902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB077902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB085902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record