GORE VIABAHN VBX Balloon Expandable Endo BXB073901ARecalls & reported problems
W. L. Gore & Associates, Inc. · BX2 HEP REDUCED PROFILE BX2 7MMX39MM 6Fr 80CM Cath · Class 3 · Implant
All models of GORE VIABAHN VBX Balloon Expandable Endo (111)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00733132658633), received between Apr 22, 2025 and Apr 22, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Complete Blockage” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (P160021); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 1, 2024Class IIOpen, Classified
Due to packaging defects, sterility assurance and heparin activity may be compromised.
Z-2945-2024 · Root cause: Package design/selection · Recall details · FDA record ↗
Most reported problems
Device problems
- Complete Blockage1
Patient problems
- Thrombosis/Thrombus1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 22, 2025Complete Blockage; Complete Blockage
ON (B)(6) 2025, A PATIENT UNDERWENT TREATMENT FOR PLAQUE IN THE ABDOMINAL AORTA USING A GORE® EXCLUDER® AAA ENDOPROSTHESIS WHERE A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) WAS USED IN THE LEFT COMMON ILIAC ARTERY. IT WAS REPORTED THE PATIENT RETURNED TO THE HOSPITAL ON (B)(6) 2025, WITH WORSENING PAIN AND NEW NUMBNESS IN THE FOOT. A CT SCAN REVEALED THROMBOSIS ON LEFT LIMB, PROMPTING A THROMBECTOMY PROCEDURE. AT THIS TIME, THE PHYSICIAN IMPLANTED THREE ADDITIONAL VBX DEVICES TO REINFORCE BOTH LIMBS UP THE GORE® EXCLUDER® AAA ENDOPROSTHESIS BIFURCATION. A FINAL ANGIOGRAM WAS PERFORMED AND EVERYTHING LOOKED PATENT. ADDITIONALLY, AN INTRAVASCULAR ULTRASOUND (IVUS) WAS PERFORMED ON BOTH LIMBS, WITH RESULTS SHOWING NO ABNORMALITIES. PHYSICIAN SUSPECTED THE ISSUE WAS POSSIBLY DUE TO POST DILATION AFTER INITIAL PROCEDURE, WITH THE RIGHT LIMB EXERTING PRESSURE ON THE LEFT LIMB. THE PATIENT IS CURRENTLY DOING WELL.
Patient problems: Thrombosis/Thrombus
MAUDE report 21885914 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160021. One submission often covers many models, so check the product and lot details in each record.
- Z-0077-2022 · Sep 7, 2021 · Open, ClassifiedHeparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diame
Other Iliac covered stent, arterial devices
- GORE VIABAHN VBX Balloon Expandable Endo BXB073902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths19 reports
- iCast Covered Stent System, 6mmX16mmX120cm 42616Atrium Medical Corporation · Iliac Covered Stent, Arterial1 recall15 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB057902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB067902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB075902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall11 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB095902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB077902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB085902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page