GORE VIABAHN VBX Balloon Expandable Endo BXB087901ARecalls & reported problems
W. L. Gore & Associates, Inc. · BX2 HEP REDUCED PROFILE BX2 8MMX79MM 7Fr 80CM Cath · Class 3 · Implant
All models of GORE VIABAHN VBX Balloon Expandable Endo (111)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00733132658756), received between Jul 25, 2025 and Jun 27, 2026 — classified as 3 injuries.
The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (P160021); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 1, 2024Class IIOpen, Classified
Due to packaging defects, sterility assurance and heparin activity may be compromised.
Z-2945-2024 · Root cause: Package design/selection · Recall details · FDA record ↗
Most reported problems
Device problems
- Deformation Due to Compressive Stress1
- Device Endoleak1
- Partial Blockage1
Patient problems
- Clinical Endoleak1
- Stenosis1
- Thrombosis/Thrombus1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJun 27, 2026Deformation Due to Compressive Stress
ON (B)(6) 2026, THE PATIENT UNDERWENT A COVERED ENDOVASCULAR RECONSTRUCTION OF THE AORTIC BIFURCATION (CERAB) PROCEDURE UTILIZING GORE® VIABAHN® VBX BALLOON-EXPANDABLE ENDOPROSTHESES (VBX DEVICES) DEVICES. DURING LESION PREPARATION, PRE-DILATION WAS PERFORMED USING A 7¿8 MM SHOCKWAVE INTRAVASCULAR LITHOTRIPSY (IVL) DEVICE. THE BALLOON REPORTEDLY RUPTURED DURING USE, LIKELY DUE TO THE PRESENCE OF HEAVILY CALCIFIED STENOTIC DISEASE. THE VBX DEVICE DELIVERY SYSTEMS WERE ADVANCED VIA BILATERAL FEMORAL ACCESS SITES USING 7F AND 8F COOK FLEXOR ANSEL INTRODUCER SHEATHS OVER A 0.035" TERUMO GLIDEWIRE ADVANTAGE® HYDROPHILIC COATED GUIDEWIRE TO THE TARGET TREATMENT SITES. THREE ENDOPROSTHESES WERE SUCCESSFULLY DELIVERED AND DEPLOYED WITHOUT REPORTED COMPLICATIONS. THE 11 MM VBX DEVICE IN THE AORTA WAS POST-DILATED TO 16 MM TO OPTIMIZE APPOSITION. THE BILATERAL 8 MM ILIAC KISSING VBX STENTS WERE NOT POST-DILATED. ALL BALLOONS DEFLATED AS INTENDED, AND THE DELIVERY SYSTEMS WERE WITHDRAWN WITHOUT RESISTANCE OR BALLOON STICTION. INTRAVASCULAR ULTRASOUND (IVUS) DEMONSTRATED ADEQUATE EXPANSION AND LUMINAL PATENCY AT THE COMPLETION OF THE PROCEDURE. HEMOSTASIS WAS ACHIEVED USING A TERUMO ANGIO-SEAL
Patient problems: Stenosis
MAUDE report 25645954 ↗InjuryDec 15, 2025Device Endoleak; Device Endoleak
THE FOLLOWING WAS REPORTED FROM A STUDY: ON (B)(6) 2025, A STUDY PATIENT WAS TREATED FOR A COMMON ILIAC ARTERY ANEURYSM IN AN IBE PROCEDURE. DURING THE PROCEDURE, A GORE EXCLUDER ILIAC BRANCH ENDOPROSTHESIS INTERNAL ILIAC COMPONENT (IBE-IIC) AND A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) WERE IMPLANTED IN AN OVERLAPPED FASHION IN THE PATIENT'S LEFT HYPOGASTRIC ARTERY. AN ADJUNCTIVE PROCEDURE WAS ALSO PERFORMED: EMBOLIZATION USING COILS. ON (B)(6) 2025, AN INTERVENTION WAS PERFORMED DUE TO A TYPE IIIB ENDOLEAK. A VBX DEVICE WAS IMPLANTED IN THE LEFT HYPOGASTRIC ARTERY TO EXCLUDE THE ENDOLEAK. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Clinical Endoleak
MAUDE report 23811653 ↗InjuryJul 25, 2025Partial Blockage; Partial Blockage
THE FOLLOWING WAS REPORTED TO GORE ON JULY 8, 2025, FROM THE IMAGING DATABASE: THE FOLLOWING DESCRIPTION SUMMARY WAS PROVIDED FROM INFORMATION EXTRACTED FROM THE CSD BY THE CCS TEAM: ON (B)(6) 2025, THIS 68-YEAR-OLD MALE PATIENT UNDERWENT ENDOVASCULAR TREATMENT AS A PLANNED ENDOVASCULAR PROCEDURE FOR AN AORTOILIAC ANEURYSM. THE PATIENT WAS TREATED WITH TWO GORE® EXCLUDER® ENDOPROSTHESIS DEVICES AND THREE GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE). THE GORE® DEVICES WERE SUCCESSFULLY NAVIGATED TO THE INTENDED LOCATIONS AND SUCCESSFULLY DEPLOYED. THE SITES WERE ACCESSED PERCUTANEOUSLY. DEVICE CATHETERS WERE SUCCESSFULLY REMOVED AFTER SUCCESSFUL ACCESS, DELIVERY AND DEPLOYMENT OF THE DEVICES. THE PATIENT WAS DISCHARGED HOME ON (B)(6) 2025. ON (B)(6) 2025, IT WAS NOTED THE PATIENT HAD A CTA. THERE IS NO LOSS OF PATENCY NOTED. IT IS NOTED BY IMAGINING THAT THERE IS DEVICE INFOLDING OR COMPRESSION OF THE GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS (IBC). IT IS ALSO NOTED THAT THERE IS MILD THROMBUS PRESENT IN THE VBX DEVICES IN THE RCI AND RIIA. (B)(6): -OTHER USED TO CAPTURE THROMBOSIS.
Patient problems: Thrombosis/Thrombus
MAUDE report 22612666 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160021. One submission often covers many models, so check the product and lot details in each record.
- Z-0077-2022 · Sep 7, 2021 · Open, ClassifiedHeparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diame
Other Iliac covered stent, arterial devices
- GORE VIABAHN VBX Balloon Expandable Endo BXB073902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths19 reports
- iCast Covered Stent System, 6mmX16mmX120cm 42616Atrium Medical Corporation · Iliac Covered Stent, Arterial1 recall15 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB057902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB067902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB075902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall11 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB095902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB077902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB085902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page