GORE VIABAHN VBX Balloon Expandable Endo BXB087902ARecalls & reported problems

W. L. Gore & Associates, Inc. · BX2 HEP REDUCED PROFILE BX2 8MMX79MM 7Fr 135CM Cath · Class 3 · Implant

All models of GORE VIABAHN VBX Balloon Expandable Endo (111)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00733132658763), received between Jun 13, 2024 and May 1, 2026 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (P160021); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 1, 2024Class IIOpen, Classified

    Due to packaging defects, sterility assurance and heparin activity may be compromised.

    Z-2945-2024 · Root cause: Package design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Complete Blockage1
  • Deformation Due to Compressive Stress1
  • Device Stenosis1
  • Insufficient Device Problem Information1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Obstruction/Occlusion2
  • Stenosis1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 1, 2026Complete Blockage

    THE FOLLOWING WAS REPORTED FROM A STUDY: ON (B)(6) 2026, A STUDY PATIENT UNDERWENT A THORACOABDOMINAL AORTIC ANEURYSM PROCEDURE. GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) WERE IMPLANTED DURING THIS PROCEDURE. DURING THIS PROCEDURE, GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) DEVICES WERE IMPLANTED IN THE VISCERAL ARTERIES AS BRANCH DEVICES. DISSECTION IN THE SUPERIOR MESENTERIC ARTERY (SMA) WAS OBSERVED LATER IN THE EVENING POST PROCEDURE. AS REPORTED, THIS WAS PROCEDURE RELATED AND THE DISSECTION WAS CAUSED BY THE GUIDEWIRE. ON (B)(6) 2026, THE SMA DISSECTION FLAP WITH LUMEN THROMBOSIS WAS STENTED WITH A 7X40 INNOVA NITINOL STENT THAT EXTENDED FROM THE GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) AROUND THE GENU TO TACK UP THE FLAP.

    Patient problems: Obstruction/Occlusion

    MAUDE report 25059176 ↗
  • InjuryApr 21, 2026Insufficient Device Problem Information

    ON (B)(6) 2025, THE PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF A ZONE 5 DESCENDING THORACOABDOMINAL AORTIC ANEURYSM USING A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) IN THE SUPERIOR MESENTERIC ARTERY (SMA). ON (B)(6) 2026, THE PATIENT UNDERWENT A REINTERVENTION DUE TO OCCLUSION OF THE SMA STENT. THE ENDOPROSTHESIS IN THE SMA WAS FOUND TO BE OCCLUDED AND WAS 'BENDING' OUT OF THE PORTAL. HOWEVER, IT WAS NOTED THE COLLATERAL VESSELS WERE FILLING, AND THERE WAS A CONCERN REGARDING ADDITIONAL RADIATION EXPOSURE, THEREFORE NO FURTHER INTERVENTION OF THE SMA WAS PERFORMED. NO ADDITIONAL TREATMENT PLAN IS CURRENTLY IN PLACE.

    Patient problems: Obstruction/Occlusion

    MAUDE report 24942357 ↗
  • InjuryMay 1, 2025Deformation Due to Compressive Stress; Device Stenosis; Deformation Due to Compressive Stress; Device Stenosis

    THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2025, PATIENT UNDERWENT AN ENDOVASCULAR PROCEDURE TO TREAT AN ANEURYSM UTILIZING GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE). ON (B)(6) 2025, THE PATIENT PRESENTED WITH SLIGHT PAIN. A COMPUTED TOMOGRAPHY SCAN WAS TAKEN, AND IT SHOWED THE LEFT RENAL VBX DEVICE WAS COMPRESSED. PATIENT UNDERWENT REINTERVENTION SURGERY IN WHICH A THROMBECTOMY AND PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY WERE PERFORMED ON THE COMPRESSED VBX DEVICE. THE VBX DEVICE WAS THEN RELINED WITH A GORE® VIABAHN® ENDOPROSTHESIS DEVICE. PATIENT TOLERATED THE PROCEDURE. NO FURTHER PATIENT INFORMATION IS AVAILABLE.

    Patient problems: Stenosis

    MAUDE report 21942000 ↗
  • MalfunctionMar 26, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE FOLLOWING WAS REPORTED TO GORE: DURING A THORACOABDOMINAL BRANCH ENDOPROSTHESIS PROCEDURE, A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) WAS TO BE IMPLANTED IN THE LEFT RENAL ARTERY. USING AN 8FR SHEATH (UNSPECIFIED BRAND), AN 8 X 79 VBX DEVICE WAS ADVANCED TO THE INTENDED TREATMENT SITE. HOWEVER, IT WAS REALIZED THE VBX DEVICE WAS TOO LONG. THE PHYSICIAN WAS PULLING THE DEVICE BACK OUT THROUGH THE SHEATH WHEN THE VBX STENT WAS DISLODGED. A GOOSENECK SNARE WAS USED TO CAPTURE AND REMOVE THE STENT FROM THE PATIENT WITH EASE. A NEW 8 X 59 VBX DEVICE WAS THEN ADVANCED AND DEPLOYED WITH NO ISSUES. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21698400 ↗
  • MalfunctionFeb 18, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2025, PATIENT UNDERWENT TREATMENT FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) WHERE A REDUCED PROFILE GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) WAS USED IN THE RIGHT INTERNAL ILIAC. IT WAS REPORTED THAT THE VBX DEVICE DISLODGED FROM THE BALLOON WHILE PHYSICIAN WAS PULLING BACK ON AN UNKNOWN BRAND 12X45 SHEATH AND THE STENT SHEARED OFF THE BALLOON. THE STENT STAYED INSIDE THE PATIENT IN THE TERMINAL AORTA. IT WAS REPORTED THAT THE STENT WAS LEFT AS ENDOTRASH SINCE IT WAS NOT DEPLOYED. A NEW VBX DEVICE WAS USED TO COMPLETE THE PROCEDURE. REPORTEDLY, PHYSICIAN SUSPECTS THE ISSUE WAS DUE TO THE EXTREMELY TORTUOUS ANATOMY. IT WAS NOT A SMOOTH TRANSITION DUE TO THE ANATOMY OF THE PATIENT. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21397590 ↗
  • MalfunctionJun 13, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    THE FOLLOWING WAS REPORTED TO GORE: DURING AN ABDOMINAL ANEURYSM PROCEDURE, TWO GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESES (VBX DEVICES) WERE IMPLANTED TO TREAT THE AORTO-ILIAC ARTERIES USING A KISSING TECHNIQUE. AS REPORTED, THE VBX DEVICES REACHED THE INTENDED TREATMENT SITE AND BOTH WERE DEPLOYED WITH NO ISSUES. HOWEVER, WHEN THE VBX BALLOON CATHETER WAS REMOVED FROM THE PATIENT'S LEFT SIDE ACCESS, IT WAS NOTICED THE BALLOON COVERING HAD BECOME DETACHED ON ONE END. THE REMOVED BALLOON WAS CONFIRMED TO HAVE NO MISSING PIECES. IT IS UNKNOWN WHICH BALLOON WAS PARTIALLY DETACHED AS SAME SIZED VBX DEVICES WERE USED. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19529380 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160021. One submission often covers many models, so check the product and lot details in each record.

  • Z-0077-2022 · Sep 7, 2021 · Open, Classified
    Heparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diame

Other Iliac covered stent, arterial devices

All Iliac covered stent, arterial devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page