GORE VIABAHN VBX Balloon Expandable Endo BXB097902ARecalls & reported problems
W. L. Gore & Associates, Inc. · BX2 HEP REDUCED PROFILE BX2 9MMX79MM 7Fr 135CM Cath · Class 3 · Implant
All models of GORE VIABAHN VBX Balloon Expandable Endo (111)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00733132658848), received between Oct 30, 2024 and May 1, 2026 — classified as 6 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (P160021); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 1, 2024Class IIOpen, Classified
Due to packaging defects, sterility assurance and heparin activity may be compromised.
Z-2945-2024 · Root cause: Package design/selection · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Dislodged or Dislocated4
- Device Endoleak1
- Migration1
- Partial Blockage1
- Patient-Device Incompatibility1
Patient problems
- Clinical Endoleak5
- Device Embedded In Tissue or Plaque1
- Stenosis1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryMay 1, 2026Device Dislodged or Dislocated
ON (B)(6) 2025, THE PATIENT UNDERWENT ENDOVASCULAR REPAIR OF A THORACOABDOMINAL AORTIC ANEURYSM USING THE GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) SYSTEM, AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) DEVICES IN THE BRANCH VESSELS. THE PATIENT TOLERATED THE PROCEDURE WITHOUT REPORTED COMPLICATIONS. ON (B)(6) 2026, A TYPE 1C ENDOLEAK WAS IDENTIFIED AT THE LEVEL OF THE SUPERIOR MESENTERIC ARTERY. ON (B)(6) 2026, THE PATIENT UNDERWENT ENDOVASCULAR REINTERVENTION WITH PLACEMENT OF AN ADDITIONAL VBX DEVICE (9 MM × 59 MM) TO TREAT THE ENDOLEAK. THE REINTERVENTION SUCCESSFULLY RESOLVED THE EVENT, AND THE DEVICE WAS PATENT AT THE CONCLUSION OF THE PROCEDURE. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL ON (B)(6) 2026.
Patient problems: Clinical Endoleak
MAUDE report 25057342 ↗InjuryMar 30, 2026Partial Blockage
THE FOLLOWING INFORMATION WAS REPORTED THROUGH THE AAA 24-01 TAMBE POST APPROVAL STUDY: ON (B)(6) 2026, THE PATIENT UNDERWENT ENDOVASCULAR PROCEDURE TO TREAT A THORACOABDOMINAL AORTIC ANEURYSM USING A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) DEVICE SYSTEM AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE). ON (B)(6) 2026, OCCLUSION WAS IDENTIFIED IN THE CELIAC ARTERY VBX DEVICE. NO REINTERVENTION HAS BEEN PERFORMED AT THIS TIME.
Patient problems: Stenosis
MAUDE report 24727712 ↗InjuryFeb 18, 2026Migration
ON (B)(6) 2025, A PATIENT WAS TREATED FOR AN EMERGENT THORACOABDOMINAL AORTIC RUPTURE USING A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) SYSTEM WITH GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) DEVICES IN THE VISCERAL VESSELS. THE RENAL ARTERIES WERE PLUGGED USING ABBOTT AMPLATZER¿ VASCULAR PLUG II (AVP II). THE AVP II PLUGS WERE PLACED WITHIN THE VBX DEVICES. THE PROCEDURE CONCLUDED SUCCESSFULLY. ON AN UNKNOWN DATE, THE PATIENT BECAME HYPOTENSIVE POST OPERATION. IT WAS DETERMINED THE VBX DEVICE HAD MOVED FROM THE SUPERIOR MESENTERIC ARTERY (SMA) PORTAL OF THE TAMBE DEVICE, FORMING A POST-PROCEDURE TYPE III EDNOLEAK. IT'S UNKNOWN WHY THE VBX DEVICE MOVED. A REINTERVENTION TOOK PLACE TO RELINE THE SMA PORTAL. TWO DAYS POST OPERATION, THE PATIENT EXPIRED, DUE TO THE PATIENT'S FAMILY DECIDED TO DISCONTINUE CARE. ACCORDING TO THE PHYSICIAN, THE PATIENT HAD GI BLEEDING. THE PATIENT HAD MALPERFUSION TO SOME OF THEIR VISCERAL VESSELS AND IT WAS NOT UNEXPECTED THAT THE PATIENT HAD GI BLEEDING. THE PHYSICIAN DID NOT STATE THAT GORE DEVICES CAUSED OR CONTRIBUTED TO THE MALPERFUSION - THE PATIENT DIDN¿T HAVE BLOOD FLOW TO BEGIN WITH. THE MALPERFUSION O
Patient problems: Clinical Endoleak
MAUDE report 24380748 ↗InjuryNov 4, 2025Patient-Device Incompatibility; Device Endoleak; Patient-Device Incompatibility; Device Endoleak
THE FOLLOWING INFORMATION WAS REPORTED THROUGH THE (B)(6) TAMBE POST APPROVAL STUDY: ON (B)(6) 2025, THE PATIENT UNDERWENT ENDOVASCULAR PROCEDURE TO TREAT A PARARENAL AORTIC ANEURYSM USING A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) DEVICE SYSTEM WITH GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICES). THE PATIENT TOLERATED THE PROCEDURE AND DISCHARGED HOME ON (B)(6) 2025. ON (B)(6) 2025, A COMPUTED TOMOGRAPHY ANGIOGRAPHY (CTA) WAS DONE AND AN AORTIC ANEURYSM ENLARGEMENT OF 10MM WAS DETERMINED. ADDITIONALLY, ON (B)(6) 2025, AN ENDOLEAK OF AN UNKNOWN ETIOLOGY WAS DETERMINED. ACCORDING TO DR. (B)(6) THE ENDOLEAK CAUSED THE ANEURYSM ENLARGEMENT. THE SITE UPDATED THE STUDY DATA FOR: ON OCTOBER 11, 2025, A TYPE II ENDOLEAK AND A TYPE IIIC ENDOLEAK WERE DETERMINED. THE PATIENT RECEIVED VASOACTIVE MEDICATIONS PENDING VASCULAR SURGERY CONSULT. ON (B)(6) 2025, THE ADDITIONAL PROCEDURE WAS DONE WHERE DR. (B)(6) DID THE FOLLOWING: STENTED THE SUPERIOR MESENTERIC AND RIGHT RENAL ARTERIES AS WELL AS RETROGRAFTED THE AORTIC SAC ACCESS WITH A STENTING OF RIGHT RENAL WITH GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX) DEVICE AND DID A COIL EMBOLIZAT
Patient problems: Clinical Endoleak
MAUDE report 23466752 ↗InjuryMay 12, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON (B)(6) 2025, A PATIENT WAS TREATED USING A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESES (VBX DEVICE) DEVICES IN THE BRANCH VESSELS. IT WAS REPORTED DURING THE PROCEDURE, TWO 9 MM X 79 MM VBX DEVICES WERE IMPLANTED WITHOUT INCIDENCE IN THE SUPERIOR MESENTERIC ARTERY (SMA). AT THE END OF THE PROCEDURE THERE WERE NO ENDOLEAKS NOTED, AND THE SMA WAS PATENT. ON AN UNKNOWN DATE, A COUPLE OF DAYS LATER, A COMPUTED TOMOGRAPHY (CT) SCAN SHOWED A ENDOLEAK BETWEEN THE TWO VBX DEVICES. THE PROXIMAL VBX DEVICE WAS STILL POSITIONED IN THE TAMBE PORTAL AND THE DISTAL VBX DEVICE WAS IN THE SMA. HOWEVER, IT WAS REPORTED THERE WAS A GAP BETWEEN THE TWO DEVICES. ON (B)(6) 2025, THERE WAS A REINTERVENTION PROCEDURE PERFORMED TO EXCLUDE THE ENDOLEAK. ANOTHER VBX DEVICE WAS IMPLANTED, BRIDGING THE PREVIOUSLY PLACED DEVICES. IT WAS REPORTED THE SMA WAS PATENT THROUGHOUT. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Clinical Endoleak
MAUDE report 22004498 ↗InjuryMay 12, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON (B)(6) 2025, A PATIENT WAS TREATED USING A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESES (VBX DEVICE) DEVICES IN THE BRANCH VESSELS. IT WAS REPORTED DURING THE PROCEDURE, TWO 9 MM X 79 MM VBX DEVICES WERE IMPLANTED WITHOUT INCIDENCE IN THE SUPERIOR MESENTERIC ARTERY (SMA). AT THE END OF THE PROCEDURE THERE WERE NO ENDOLEAKS NOTED, AND THE SMA WAS PATENT. ON AN UNKNOWN DATE, A COUPLE OF DAYS LATER, A COMPUTED TOMOGRAPHY (CT) SCAN SHOWED AN ENDOLEAK BETWEEN THE TWO VBX DEVICES. THE PROXIMAL VBX DEVICE WAS STILL POSITIONED IN THE TAMBE PORTAL AND THE DISTAL VBX DEVICE WAS IN THE SMA. HOWEVER, IT WAS REPORTED THERE WAS A GAP BETWEEN THE TWO DEVICES. ON (B)(6) 2025, THERE WAS A REINTERVENTION PROCEDURE PERFORMED TO EXCLUDE THE ENDOLEAK. ANOTHER VBX DEVICE WAS IMPLANTED, BRIDGING THE PREVIOUSLY PLACED DEVICES. IT WAS REPORTED THE SMA WAS PATENT THROUGHOUT. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Clinical Endoleak
MAUDE report 22003975 ↗MalfunctionOct 30, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2024, A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) WAS INTENDED FOR IMPLANTATION INTO THE LEFT HYPOGASTRIC ARTERY AS AN EXTENSION TO A GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS (IBE) IN AN ENDOVASCULAR PROCEDURE TO REPAIR ANEURYSMS. CONTRALATERAL ACCESS WAS GAINED, AND IT WAS REPORTED THAT THERE WAS AN INABILITY TO ADVANCE THE VBX DEVICE DELIVERY CATHETER (DUE TO EXCESSIVELY TORTUOUS ANATOMY) THROUGH A 12 F GORE® DRYSEAL FLEX INTRODUCER SHEATH. REPORTEDLY, DURING REMOVAL, THE VBX DEVICE DELIVERY CATHETER BROKE AT THE HUB AND THE STENT PORTION DISLODGED FROM THE BALLOON AND INTO THE LEFT COMMON ILIAC ARTERY (LCIA). THE VBX DEVICE STENT PORTION WAS LEFT IN THE LCIA AS ENDOTRASH. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 20566572 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160021. One submission often covers many models, so check the product and lot details in each record.
- Z-0077-2022 · Sep 7, 2021 · Open, ClassifiedHeparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diame
Other Iliac covered stent, arterial devices
- GORE VIABAHN VBX Balloon Expandable Endo BXB073902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths19 reports
- iCast Covered Stent System, 6mmX16mmX120cm 42616Atrium Medical Corporation · Iliac Covered Stent, Arterial1 recall15 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB057902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB067902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB075902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall11 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB095902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB077902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB085902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page