Skip to content

GORE VIABAHN VBX Balloon Expandable Endo BXB097902ARecalls & reported problems

W. L. Gore & Associates, Inc. · BX2 HEP REDUCED PROFILE BX2 9MMX79MM 7Fr 135CM Cath · Class 3 · Implant

All models of GORE VIABAHN VBX Balloon Expandable Endo (111)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
6
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00733132658848), received between Oct 30, 2024 and May 1, 2026 — classified as 6 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (P160021); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 1, 2024Class IIOpen, Classified

    Due to packaging defects, sterility assurance and heparin activity may be compromised.

    Z-2945-2024 · Root cause: Package design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Dislodged or Dislocated4
  • Device Endoleak1
  • Migration1
  • Partial Blockage1
  • Patient-Device Incompatibility1

Patient problems

  • Clinical Endoleak5
  • Device Embedded In Tissue or Plaque1
  • Stenosis1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 1, 2026Device Dislodged or Dislocated

    ON (B)(6) 2025, THE PATIENT UNDERWENT ENDOVASCULAR REPAIR OF A THORACOABDOMINAL AORTIC ANEURYSM USING THE GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) SYSTEM, AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) DEVICES IN THE BRANCH VESSELS. THE PATIENT TOLERATED THE PROCEDURE WITHOUT REPORTED COMPLICATIONS. ON (B)(6) 2026, A TYPE 1C ENDOLEAK WAS IDENTIFIED AT THE LEVEL OF THE SUPERIOR MESENTERIC ARTERY. ON (B)(6) 2026, THE PATIENT UNDERWENT ENDOVASCULAR REINTERVENTION WITH PLACEMENT OF AN ADDITIONAL VBX DEVICE (9 MM × 59 MM) TO TREAT THE ENDOLEAK. THE REINTERVENTION SUCCESSFULLY RESOLVED THE EVENT, AND THE DEVICE WAS PATENT AT THE CONCLUSION OF THE PROCEDURE. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL ON (B)(6) 2026.

    Patient problems: Clinical Endoleak

    MAUDE report 25057342 ↗
  • InjuryMar 30, 2026Partial Blockage

    THE FOLLOWING INFORMATION WAS REPORTED THROUGH THE AAA 24-01 TAMBE POST APPROVAL STUDY: ON (B)(6) 2026, THE PATIENT UNDERWENT ENDOVASCULAR PROCEDURE TO TREAT A THORACOABDOMINAL AORTIC ANEURYSM USING A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) DEVICE SYSTEM AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE). ON (B)(6) 2026, OCCLUSION WAS IDENTIFIED IN THE CELIAC ARTERY VBX DEVICE. NO REINTERVENTION HAS BEEN PERFORMED AT THIS TIME.

    Patient problems: Stenosis

    MAUDE report 24727712 ↗
  • InjuryFeb 18, 2026Migration

    ON (B)(6) 2025, A PATIENT WAS TREATED FOR AN EMERGENT THORACOABDOMINAL AORTIC RUPTURE USING A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) SYSTEM WITH GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) DEVICES IN THE VISCERAL VESSELS. THE RENAL ARTERIES WERE PLUGGED USING ABBOTT AMPLATZER¿ VASCULAR PLUG II (AVP II). THE AVP II PLUGS WERE PLACED WITHIN THE VBX DEVICES. THE PROCEDURE CONCLUDED SUCCESSFULLY. ON AN UNKNOWN DATE, THE PATIENT BECAME HYPOTENSIVE POST OPERATION. IT WAS DETERMINED THE VBX DEVICE HAD MOVED FROM THE SUPERIOR MESENTERIC ARTERY (SMA) PORTAL OF THE TAMBE DEVICE, FORMING A POST-PROCEDURE TYPE III EDNOLEAK. IT'S UNKNOWN WHY THE VBX DEVICE MOVED. A REINTERVENTION TOOK PLACE TO RELINE THE SMA PORTAL. TWO DAYS POST OPERATION, THE PATIENT EXPIRED, DUE TO THE PATIENT'S FAMILY DECIDED TO DISCONTINUE CARE. ACCORDING TO THE PHYSICIAN, THE PATIENT HAD GI BLEEDING. THE PATIENT HAD MALPERFUSION TO SOME OF THEIR VISCERAL VESSELS AND IT WAS NOT UNEXPECTED THAT THE PATIENT HAD GI BLEEDING. THE PHYSICIAN DID NOT STATE THAT GORE DEVICES CAUSED OR CONTRIBUTED TO THE MALPERFUSION - THE PATIENT DIDN¿T HAVE BLOOD FLOW TO BEGIN WITH. THE MALPERFUSION O

    Patient problems: Clinical Endoleak

    MAUDE report 24380748 ↗
  • InjuryNov 4, 2025Patient-Device Incompatibility; Device Endoleak; Patient-Device Incompatibility; Device Endoleak

    THE FOLLOWING INFORMATION WAS REPORTED THROUGH THE (B)(6) TAMBE POST APPROVAL STUDY: ON (B)(6) 2025, THE PATIENT UNDERWENT ENDOVASCULAR PROCEDURE TO TREAT A PARARENAL AORTIC ANEURYSM USING A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) DEVICE SYSTEM WITH GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICES). THE PATIENT TOLERATED THE PROCEDURE AND DISCHARGED HOME ON (B)(6) 2025. ON (B)(6) 2025, A COMPUTED TOMOGRAPHY ANGIOGRAPHY (CTA) WAS DONE AND AN AORTIC ANEURYSM ENLARGEMENT OF 10MM WAS DETERMINED. ADDITIONALLY, ON (B)(6) 2025, AN ENDOLEAK OF AN UNKNOWN ETIOLOGY WAS DETERMINED. ACCORDING TO DR. (B)(6) THE ENDOLEAK CAUSED THE ANEURYSM ENLARGEMENT. THE SITE UPDATED THE STUDY DATA FOR: ON OCTOBER 11, 2025, A TYPE II ENDOLEAK AND A TYPE IIIC ENDOLEAK WERE DETERMINED. THE PATIENT RECEIVED VASOACTIVE MEDICATIONS PENDING VASCULAR SURGERY CONSULT. ON (B)(6) 2025, THE ADDITIONAL PROCEDURE WAS DONE WHERE DR. (B)(6) DID THE FOLLOWING: STENTED THE SUPERIOR MESENTERIC AND RIGHT RENAL ARTERIES AS WELL AS RETROGRAFTED THE AORTIC SAC ACCESS WITH A STENTING OF RIGHT RENAL WITH GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX) DEVICE AND DID A COIL EMBOLIZAT

    Patient problems: Clinical Endoleak

    MAUDE report 23466752 ↗
  • InjuryMay 12, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2025, A PATIENT WAS TREATED USING A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESES (VBX DEVICE) DEVICES IN THE BRANCH VESSELS. IT WAS REPORTED DURING THE PROCEDURE, TWO 9 MM X 79 MM VBX DEVICES WERE IMPLANTED WITHOUT INCIDENCE IN THE SUPERIOR MESENTERIC ARTERY (SMA). AT THE END OF THE PROCEDURE THERE WERE NO ENDOLEAKS NOTED, AND THE SMA WAS PATENT. ON AN UNKNOWN DATE, A COUPLE OF DAYS LATER, A COMPUTED TOMOGRAPHY (CT) SCAN SHOWED A ENDOLEAK BETWEEN THE TWO VBX DEVICES. THE PROXIMAL VBX DEVICE WAS STILL POSITIONED IN THE TAMBE PORTAL AND THE DISTAL VBX DEVICE WAS IN THE SMA. HOWEVER, IT WAS REPORTED THERE WAS A GAP BETWEEN THE TWO DEVICES. ON (B)(6) 2025, THERE WAS A REINTERVENTION PROCEDURE PERFORMED TO EXCLUDE THE ENDOLEAK. ANOTHER VBX DEVICE WAS IMPLANTED, BRIDGING THE PREVIOUSLY PLACED DEVICES. IT WAS REPORTED THE SMA WAS PATENT THROUGHOUT. THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: Clinical Endoleak

    MAUDE report 22004498 ↗
  • InjuryMay 12, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2025, A PATIENT WAS TREATED USING A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESES (VBX DEVICE) DEVICES IN THE BRANCH VESSELS. IT WAS REPORTED DURING THE PROCEDURE, TWO 9 MM X 79 MM VBX DEVICES WERE IMPLANTED WITHOUT INCIDENCE IN THE SUPERIOR MESENTERIC ARTERY (SMA). AT THE END OF THE PROCEDURE THERE WERE NO ENDOLEAKS NOTED, AND THE SMA WAS PATENT. ON AN UNKNOWN DATE, A COUPLE OF DAYS LATER, A COMPUTED TOMOGRAPHY (CT) SCAN SHOWED AN ENDOLEAK BETWEEN THE TWO VBX DEVICES. THE PROXIMAL VBX DEVICE WAS STILL POSITIONED IN THE TAMBE PORTAL AND THE DISTAL VBX DEVICE WAS IN THE SMA. HOWEVER, IT WAS REPORTED THERE WAS A GAP BETWEEN THE TWO DEVICES. ON (B)(6) 2025, THERE WAS A REINTERVENTION PROCEDURE PERFORMED TO EXCLUDE THE ENDOLEAK. ANOTHER VBX DEVICE WAS IMPLANTED, BRIDGING THE PREVIOUSLY PLACED DEVICES. IT WAS REPORTED THE SMA WAS PATENT THROUGHOUT. THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: Clinical Endoleak

    MAUDE report 22003975 ↗
  • MalfunctionOct 30, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2024, A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) WAS INTENDED FOR IMPLANTATION INTO THE LEFT HYPOGASTRIC ARTERY AS AN EXTENSION TO A GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS (IBE) IN AN ENDOVASCULAR PROCEDURE TO REPAIR ANEURYSMS. CONTRALATERAL ACCESS WAS GAINED, AND IT WAS REPORTED THAT THERE WAS AN INABILITY TO ADVANCE THE VBX DEVICE DELIVERY CATHETER (DUE TO EXCESSIVELY TORTUOUS ANATOMY) THROUGH A 12 F GORE® DRYSEAL FLEX INTRODUCER SHEATH. REPORTEDLY, DURING REMOVAL, THE VBX DEVICE DELIVERY CATHETER BROKE AT THE HUB AND THE STENT PORTION DISLODGED FROM THE BALLOON AND INTO THE LEFT COMMON ILIAC ARTERY (LCIA). THE VBX DEVICE STENT PORTION WAS LEFT IN THE LCIA AS ENDOTRASH. THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 20566572 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160021. One submission often covers many models, so check the product and lot details in each record.

  • Z-0077-2022 · Sep 7, 2021 · Open, Classified
    Heparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diame

Other Iliac covered stent, arterial devices

All Iliac covered stent, arterial devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page