Skip to content

GORE® DrySeal Flex Introducer Sheath GDSF1465Recalls & reported problems

W. L. Gore & Associates, Inc. · DrySeal Hydro Flex Sheath IC 65cmx14Fr · Class 2

All models of GORE DRYSEAL FLEX Introducer Sheath (29)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00733132658930), received between Mar 30, 2026 and Mar 30, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (1 report).

1 other recall cover devices cleared under the same FDA submission (K160254); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2026Material Separation

    ON (B)(6) 2026, THE PATIENT UNDERWENT AN PERIPHERAL ARTERY DISEASE PROCEDURE TO TREAT A RIGHT LOWER EXTREMITY ACUTE LIMB ISCHEMIA AND A THROMBOSED ARTERIAL SYSTEM, UTILIZING GORE® DRYSEAL FLEX INTRODUCER SHEATH. ALLEGEDLY, DURING THE PROCEDURE, SHORTLY AFTER INSERTION AND REMOVAL OF THE DILATOR, PHYSICIAN NOTICED SOMETHING APPEARED ABNORMAL. THE PHYSICIAN ATTEMPTED TO ADVANCE A 12 FR PENUMBRA THROMBECTOMY CATHETER THROUGH THE SHEATH BUT WAS UNABLE TO PASS IT ACROSS THE AORTIC BIFURCATION. AN ANGIO SHOT WAS PERFORMED FOR BETTER VISUALIZATION, WHICH REVEALED THAT THE SHEATH WAS KINKED. IT WAS ALSO REPORTED THE PENUMBRA CATHETER WAS REMOVED, AND THE PHYSICIAN ATTEMPTED TO GENTLY WITHDRAW THE SHEATH TO STRAIGHTEN THE KINK. ONLY MINIMAL FORCE WAS APPLIED, HOWEVER THE SHEATH SEPARATED, AND ONLY THE PROXIMAL HALF WAS RETRIEVED. A PTA BALLOON WAS THEN ADVANCED TO FACILITATE THE REMOVAL OF THE REMAINING SHEATH FRAGMENT. THE REMAINDER OF THE SHEATH WAS SUCCESSFULLY EXTRACTED THROUGH THE ACCESS SITE WITHOUT FURTHER COMPLICATION. A NEW DRYSEAL SHEATH WAS THEN PLACED, AND THE CASE PROCEEDED WITHOUT ADDITIONAL ISSUES. THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24737157 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K160254. One submission often covers many models, so check the product and lot details in each record.

  • Z-1962-2020 · Nov 18, 2019 · Terminated
    Gore DrySeal Flex Introducer Sheath, Catalog #DSF1433, 33cm 14FR, Sterile. The responsible firm on the label is W. L. Gore & Associates, Inc., Flagstaff, AZ.

Other Introducer, catheter devices

All Introducer, catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page