GORE® DrySeal Flex Introducer Sheath GDSF2433Recalls & reported problems

W. L. Gore & Associates, Inc. · DrySeal Hydro Flex Sheath Ao 33cmx24Fr · Class 2

All models of GORE DRYSEAL FLEX Introducer Sheath (29)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00733132659036), received between Jan 30, 2026 and May 22, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K160254); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Vascular Dissection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 22, 2026Fluid/Blood Leak

    ON (B)(6) 2026, DURING AN ENDOVASCULAR PROCEDURE, IT WAS REPORTED THAT A GORE® DRYSEAL FLEX INTRODUCER SHEATH (GDSF) WAS ADVANCED IN THE PATIENT AND WHEN TAKING THE DILATOR OUT A LEAK WAS DETERMINED. THE GDSF WOULDN'T HOLD ANY HEP SALINE EVEN AFTER INSERTING MORE WITH THE SYRINGE. THE GDSF WAS REMOVED AND EXCHANGED OUT BEFORE THE PROCEDURE CONTINUED AND BEFORE THE GRAFT WAS INSERTED. THE PATIENT TOLERATED THE PROCEDURE. THE DEFICIENT GDSF WAS DISCARDED AT THE SITE AND NOT AVAILABLE FOR EVALUATION BY GORE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25261274 ↗
  • InjuryApr 2, 2026Adverse Event Without Identified Device or Use Problem

    THE FOLLOWING WAS REPORTED: ON (B)(6) 2026, THE PATIENT WAS IMPLANTED WITH GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS AORTIC EXTENDER AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS) USING A GORE® DRYSEAL FLEX INTRODUCER SHEATH AND A GORE® TRI-LOBE BALLOON CATHETER TO TREAT AN ASCENDING THORACIC DISSECTION. REPORTEDLY DURING THE PROCEDURE WHILE RETRACTING THE DELIVERY CATHETER OF THE GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS AORTIC EXTENDER THE DEVICE MOVED DISTALLY AND PARTIALLY COVERED THE INNOMINATE ARTERY. THE PHYSICIAN IMPLANTED A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS IN THE INNOMINATE ARTERY TO MAINTAIN PATENCY. THEN THE PHYSICIAN IMPLANTED A SECOND GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS AORTIC EXTENDER PROXIMALLY. THE PATIENT TOLERATED THE PROCEDURE. POST PROCEDURE IMAGES REVEALED A TYPE B DISSECTION FROM IMPLANTED GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS AORTIC EXTENDER EXTENDING DOWN INTO THE ILIAC ARTERY. THE PHYSICIAN IS NOT TREATING THE DISSECTION AT THIS TIME. THE EVENT IS ONGOING. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Vascular Dissection

    MAUDE report 24771037 ↗
  • MalfunctionJan 30, 2026Fluid/Blood Leak

    THE FOLLOWING EVENT WAS REPORTED TO GORE: ON (B)(6) 2026, THE PATIENT UNDERWENT AN ENDOVASCULAR PROCEDURE VIA RIGHT INTERNAL JUGULAR ACCESS USING A GORE® DRYSEAL FLEX INTRODUCER SHEATH FOR IMPLANTATION OF A NON-GORE DEVICE (MICRA¿ AV2 LEADLESS PACEMAKER). IT WAS REPORTED THAT THE HEMOSTATIC VALVE WAS PRESSURIZED DURING PREPARATION PRIOR TO INSERTION, WITH NO ISSUES NOTED. FOLLOWING INSERTION OF THE MICRA¿ AV2 DELIVERY SYSTEM INTO THE PATIENT, THE HEMOSTATIC VALVE WAS UNABLE TO MAINTAIN PRESSURE OR TIGHTEN AROUND THE DELIVERY SYSTEM, AND HEPARINIZED SALINE WAS OBSERVED LEAKING. THE REPORTER STATED THE LEAKAGE MAY HAVE BEEN DUE TO A TEAR IN THE HEMOSTATIC VALVE. DUE TO THE INABILITY TO RESTORE PRESSURIZATION, A REPLACEMENT SHEATH WAS USED TO COMPLETE THE PROCEDURE. THE PATIENT TOLERATED THE PROCEDURE, AND NO NEGATIVE PATIENT OUTCOMES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24223142 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K160254. One submission often covers many models, so check the product and lot details in each record.

  • Z-1962-2020 · Nov 18, 2019 · Terminated
    Gore DrySeal Flex Introducer Sheath, Catalog #DSF1433, 33cm 14FR, Sterile. The responsible firm on the label is W. L. Gore & Associates, Inc., Flagstaff, AZ.

Other Introducer, catheter devices

All Introducer, catheter devices →