OSOM hCG Urine 50 Test Kit 101Recalls & reported problems
Sekisui Diagnostics, LLC · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00742860001017), received between Jan 8, 2025 and Jan 8, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “False Negative Result” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- False Negative Result1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJan 8, 2025False Negative Result; False Negative Result
ON (B)(6) 2024, THE PATIENT'S FIRST MORNING URINE WAS TESTED WITH AN OSOM HCG URINE TEST, AND THE RESULT WAS NEGATIVE. NO SERUM QUANTITATIVE HCG WAS PERFORMED. ON THE SAME DAY, THE PATIENT HAD AN ABDOMINAL SURGICAL PROCEDURE. A MASS WAS DISCOVERED DURING THE PROCEDURE WHICH TURNED OUT TO BE A THIRD TRIMESTER PREGNANCY. THE SURGERY WAS HALTED, AND THE PATIENT WAS ADMITTED TO HOSPITAL. ADDITIONAL DETAILS OF THE EVENT (SUCH AS WHICH PROCEDURE THE PATIENT HAD, PATIENT WEIGHT, ETC.) WERE NOT PROVIDED. DETAILS REGARDING THE OUTCOME OF THE PATIENT OR THE PREGNANCY WERE ALSO NOT PROVIDED. THE CUSTOMER STATED THEY HAD BEEN USING THE SAME LOT FOR SOME TIME, AND THAT THEY ARE A HIGH-VOLUME FACILITY (APPROX. 50 HCG PER DAY). THE FACILITY REPORTED NO OTHER FALSE NEGATIVES HAD BEEN OBTAINED USING THIS LOT.
Patient problems: Insufficient Information
MAUDE report 21107248 ↗
Other Visual, pregnancy hCG, prescription use devices
- McKesson 5001Mckesson Medical-Surgical Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls165 reports
- CARDINAL HEALTH B1077-23Cardinal Health 200, LLC · Visual, pregnancy hCG, prescription use1 recall149 reports
- McKesson 5000Mckesson Medical-Surgical Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls90 reports
- McKesson 5002Mckesson Medical-Surgical Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls33 reports
- Sure-Vue 4581405021Fisher Scientific Company L.L.C. · Visual, pregnancy hCG, prescription use1 recall83 reports
- Henry Schein 9004072Henry Schein, Inc. · Visual, pregnancy hCG, prescription use1 recall79 reports
- AlereTM 92210Alere San Diego, Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls23 reports
- MEDLINE MPH12025HCGMedline Industries, Inc. · Visual, pregnancy hCG, prescription use1 recall69 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.