Skip to content

DeRoyal 25-3001Recalls & reported problems

Deroyal Industries, Inc. · SKIN STAPLER · Class 2

Adverse event reports
6
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00749756024372), received between Sep 30, 2020 and Dec 23, 2025 — classified as 5 injuries, 1 malfunction.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 4 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem5
  • Detachment of Device or Device Component1
  • Mechanical Problem1

Patient problems

  • Hair Loss5
  • No Clinical Signs, Symptoms or Conditions1
  • Rash1
  • Skin Inflammation/ Irritation1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 23, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    A USER FACILITY REPORTED, "PATIENT REPORTED WOUND HEALING ISSUES AT THE STAPLE SITE AND HAIR LOSS ALLEGEDLY CAUSED BY THE STAPLER."

    Patient problems: Hair Loss; Rash; Skin Inflammation/ Irritation

    MAUDE report 23876852 ↗
  • InjuryAug 9, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    A USER FACILITY REPORTED, "HELLO - I WOULD LIKE TO REPORT A QUALITY ISSUE WITH A DEROYAL PRODUCT. I RECENTLY SWITCHED PRODUCTS TO DEROYAL FOR STAPLES IN THE SCALP DURING SURGERY. I HAVE NOW HAD FOUR INSTANCES OF A SEVERE ALOPECIA AREATA OR INFLAMMATORY ALOPECIA LIKE REACTION AROUND THE STAPLES WHICH I HAVE NEVER SEEN BEFORE WITH ANY STAPLES. THEY ALL APPEAR TO BE FROM THE SAME LOT NUMBER AND TIME FRAME. ALL I CAN THINK OF IS THAT PERHAPS THERE IS SOMETHING IN THE COATING OF THE STAPLES. THE LOT NUMBER IS 73B2300556. REF 25-3001."

    Patient problems: Hair Loss

    MAUDE report 19951737 ↗
  • InjuryAug 9, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    A USER FACILITY REPORTED, "HELLO - I WOULD LIKE TO REPORT A QUALITY ISSUE WITH A DEROYAL PRODUCT. I RECENTLY SWITCHED PRODUCTS TO DEROYAL FOR STAPLES IN THE SCALP DURING SURGERY. I HAVE NOW HAD FOUR INSTANCES OF A SEVERE ALOPECIA AREATA OR INFLAMMATORY ALOPECIA LIKE REACTION AROUND THE STAPLES WHICH I HAVE NEVER SEEN BEFORE WITH ANY STAPLES. THEY ALL APPEAR TO BE FROM THE SAME LOT NUMBER AND TIME FRAME. ALL I CAN THINK OF IS THAT PERHAPS THERE IS SOMETHING IN THE COATING OF THE STAPLES. THE LOT NUMBER IS 73B2300556. REF 25-3001."

    Patient problems: Hair Loss

    MAUDE report 19951734 ↗
  • InjuryAug 9, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    A USER FACILITY REPORTED, "HELLO - I WOULD LIKE TO REPORT A QUALITY ISSUE WITH A DEROYAL PRODUCT. I RECENTLY SWITCHED PRODUCTS TO DEROYAL FOR STAPLES IN THE SCALP DURING SURGERY. I HAVE NOW HAD FOUR INSTANCES OF A SEVERE ALOPECIA AREATA OR INFLAMMATORY ALOPECIA LIKE REACTION AROUND THE STAPLES WHICH I HAVE NEVER SEEN BEFORE WITH ANY STAPLES. THEY ALL APPEAR TO BE FROM THE SAME LOT NUMBER AND TIME FRAME. ALL I CAN THINK OF IS THAT PERHAPS THERE IS SOMETHING IN THE COATING OF THE STAPLES. THE LOT NUMBER IS 73B2300556. REF 25-3001."

    Patient problems: Hair Loss

    MAUDE report 19951732 ↗
  • InjuryAug 9, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    A USER FACILITY REPORTED, "HELLO - I WOULD LIKE TO REPORT A QUALITY ISSUE WITH A DEROYAL PRODUCT. I RECENTLY SWITCHED PRODUCTS TO DEROYAL FOR STAPLES IN THE SCALP DURING SURGERY. I HAVE NOW HAD FOUR INSTANCES OF A SEVERE ALOPECIA AREATA OR INFLAMMATORY ALOPECIA LIKE REACTION AROUND THE STAPLES WHICH I HAVE NEVER SEEN BEFORE WITH ANY STAPLES. THEY ALL APPEAR TO BE FROM THE SAME LOT NUMBER AND TIME FRAME. ALL I CAN THINK OF IS THAT PERHAPS THERE IS SOMETHING IN THE COATING OF THE STAPLES. THE LOT NUMBER IS 73B2300556. REF 25-3001."

    Patient problems: Hair Loss

    MAUDE report 19951731 ↗
  • MalfunctionSep 30, 2020Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component

    WHILE USING THE STAPLER TO CLOSE THE PATIENT'S SURGICAL INCISION, MULTIPLE STAPLES FELL OUT OF THE STAPLE GUN ONTO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10608613 ↗

Other Stapler, surgical devices

All Stapler, surgical devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.