Skip to content

DeRoyal V71-6510Recalls & reported problems

Deroyal Industries, Inc. · SUCTION CANISTER, 1000cc · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00749756728980), received between Dec 6, 2024 and Dec 6, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Suction Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Suction Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 6, 2024Suction Problem; Suction Problem

    A USER FACILITY REPORTED, "PRIOR TO INTUBATION, THE DECISION WAS MADE TO PLACE A NGT (NASOGASTRIC TUBE). NGT PLACED TO SUCTION. THERE WERE DIFFICULTIES GETTING THE SUCTION TO WORK. PATIENT INTUBATED. RN (REGISTERED NURSE) WAS ABLE TO GET SUCTION TO WORK AND LARGE VOLUME OF BLACK LIQUID SUCTIONED. IT WAS NOTED THE KEY ON THE SIDE OF THE SUCTION HOLDER WAS HORIZONTAL (SUCTION WAS OFF) AND ONCE THE KEY WAS TURNED VERTICAL, IT WAS FUNCTIONAL."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20867995 ↗
  • MalfunctionDec 6, 2024Suction Problem; Suction Problem

    A USER FACILITY REPORTED, "THE DEROYAL SUCTION CANISTER LINERS ARE EXPERIENCING ISSUES WITH SUCTION DUE TO FAULTY ADHESIVES BEING APPLIED. THE END USER MUST PRESS FIRMLY WHEN CONNECTING THE LID TO THE LINER. IF THE ADHESIVE DOESN'T CONNECT ALL THE WAY AROUND, THE SUCTION WILL NOT WORK. THERE ARE ISSUES WITH THE SUCTION DECREASING AND FAILING INTERMITTENT SUCTION. IN BOTH CASES, THE ISSUE IS FIXED BY BEING EXTRA FIRM WHEN CONNECTING THE LID TO THE LINER BEFORE PLACING IT IN THE CANISTER. DEROYAL ROUNDED, TOOK NOTES AND PICTURES AND WILL BE REPORTING BACK ASAP."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20867159 ↗

Other BOTTLE, COLLECTION, VACUUM devices

All BOTTLE, COLLECTION, VACUUM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page