Skip to content

Bausch + Lomb BL5114ARecalls & reported problems

Bausch & Lomb Incorporated · 1.8mm Premium Vacuum Phaco Pack with MICS Angled Needle · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00757770057702), received between Aug 17, 2017 and May 22, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Complete Blockage” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Complete Blockage1
  • Particulates1

Patient problems

  • Eye Burn1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 22, 2025Complete Blockage; Complete Blockage

    THE USER FACILITY IN UNITED KINGDOM REPORTED THAT ON A CASE WITH A PARTICULARLY DENSE CATARACT, TIP BLOCKAGE POSSIBLY OCCURRED CAUSING ASPIRATION TO PAUSE MOMENTARILY. A SMALL CORNEAL BURN RESULTED, REQUIRING A SUTURE AND DELAYING THE SURGERY UP TO 15 MINUTES. A BAUSCH AND LOMB REPRESENTATIVE WAS PRESENT DURING THE OPERATION AND INSTRUCTED SURGEON ON HOW TO PREVENT ANY BURN IF TIP BLOCKS, AND ADJUSTED PHACO SETTINGS.

    Patient problems: Eye Burn

    MAUDE report 22069915 ↗
  • MalfunctionAug 17, 2017Particulates; Particulates

    A REPORT FROM A USER FACILITY IN THE (B)(4) STATED THAT HAIR LIKE STRANDS APPEARED TO COME OUT FROM THE PHACO TIP. THE STRANDS APPEARED RIGHT AT THE START OF SURGERY WHEN THE SURGEON STARTED IN THE SCULPT PHASE AND THEY WERE REMOVED WITH CAPSULORHEXIS FORCEPS. THE PACK, HANDPIECE AND TROLLEY WERE REPLACED. THERE WAS NO OTHER INCIDENCE DURING THE LIST.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6802194 ↗

Other Unit, Phacofragmentation devices

All Unit, Phacofragmentation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page