Bausch + Lomb BLV3220000Recalls & reported problems
Bausch & Lomb Incorporated · Vacuum Phaco Pack with 2.2mm Sleeve · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00757770058402), received between Jul 6, 2017 and Apr 4, 2025 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Particulates” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Particulates10
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 4, 2025Particulates; Particulates
THE USER FACILITY IN AUSTRIA REPORTED THAT DURING SIX PROCEDURES, WHITE PLASTIC PARTICLES ENTERED THE PATIENT'S EYE. THERE WAS NO PATIENT IMPACT REPORTED. THIS REPORT IS FOR THE FIRST PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21774934 ↗MalfunctionApr 3, 2025Particulates; Particulates
THE USER FACILITY IN AUSTRIA REPORTED THAT DURING SIX PROCEDURES, WHITE PLASTIC PARTICLES ENTERED THE PATIENT 'S EYE. THERE WAS NO PATIENT IMPACT REPORTED. THIS REPORT IS FOR THE FIRST PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21767358 ↗MalfunctionApr 3, 2025Particulates; Particulates
THE USER FACILITY IN AUSTRIA REPORTED THAT DURING SIX PROCEDURES, WHITE PLASTIC PARTICLES ENTERED THE PATIENT 'S EYE. THERE WAS NO PATIENT IMPACT REPORTED. THIS REPORT IS FOR THE FIRST PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21766860 ↗MalfunctionApr 3, 2025Particulates; Particulates
THE USER FACILITY IN AUSTRIA REPORTED THAT DURING SIX PROCEDURES, WHITE PLASTIC PARTICLES ENTERED THE PATIENT 'S EYE. THERE WAS NO PATIENT IMPACT REPORTED. THIS REPORT IS FOR THE FIRST PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21766052 ↗MalfunctionApr 3, 2025Particulates; Particulates
THE USER FACILITY IN AUSTRIA REPORTED THAT DURING SIX PROCEDURES, WHITE PLASTIC PARTICLES ENTERED THE PATIENT 'S EYE. THERE WAS NO PATIENT IMPACT REPORTED. THIS REPORT IS FOR THE FIRST PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21764566 ↗MalfunctionApr 2, 2025Particulates; Particulates
THE USER FACILITY IN AUSTRIA REPORTED THAT DURING SIX PROCEDURES, WHITE PLASTIC PARTICLES ENTERED THE PATIENT 'S EYE. THERE WAS NO PATIENT IMPACT REPORTED. THIS REPORT IS FOR THE FIRST PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21755518 ↗MalfunctionDec 20, 2023Particulates; Particulates
THE USER FACILITY REPORTS THAT THE PHACO SLEEVE HAS THREE HOLES IN IT AND ONE OF THE HOLES WAS NOT PUNCHED AND A PART OF THE GREEN SLEEVE GOT INTO THE ANTERIOR CHAMBER. THE DOCTOR WAS ABLE TO ASPIRATE IT OUT. THIS HAPPENED TWICE ON TWO SEPARATE PATIENTS ON THE SAME DAY. THE PRODUCTS WERE DISPOSED OF. THERE WAS NO MEDICAL INTERVENTION/TREATMENT REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18365230 ↗MalfunctionDec 20, 2023Particulates; Particulates
THE USER FACILITY REPORTS THAT THE PHACO SLEEVE HAS THREE HOLES IN IT AND ONE OF THE HOLES WAS NOT PUNCHED AND A PART OF THE GREEN SLEEVE AND GOT INTO THE ANTERIOR CHAMBER. THE DOCTOR WAS ABLE TO ASPIRATE IT OUT. THIS HAPPENED TWICE ON TWO SEPARATE PATIENTS ON THE SAME DAY. THE PRODUCTS WERE DISPOSED OF. THERE WAS NO MEDICAL INTERVENTION/TREATMENT REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18365213 ↗MalfunctionJun 25, 2018Particulates; Particulates
THE USER FACILITY REPORTED AS THE DOCTOR WAS PERFORMING PHACOEMULSIFICATION WHEN A BACK PARTICLE FELL INTO THE PATIENTS EYE. THE SURGEON WAS ABLE TO REMOVE THE PARTICLE, AND COMPLETE THE SURGERY WITH NO IMPACT TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7635467 ↗MalfunctionJul 6, 2017Particulates; Particulates
THE USER FACILITY REPORTED A PIECE OF PLASTIC CAME OUT OF THE SLEEVE DURING THE PROCEDURE. THE DOCTOR WAS UNSURE IF IT IS COMING FROM THE SLEEVE. HE USED A UTRATA FORCEPS TO REMOVE THE FOREIGN MATTER FROM THE EYE. NO ISSUE TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6689705 ↗
Other Unit, Phacofragmentation devices
- Centurion 8065751763Alcon Laboratories, Inc. · Unit, phacofragmentation728 reports
- Centurion 8065753057Alcon Laboratories, Inc. · Unit, phacofragmentation639 reports
- CENTURION® 8065752914Alcon Laboratories, Inc. · Unit, phacofragmentation571 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls398 reports
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751767Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls302 reports
- Constellation, ULTRAVIT, Engauge, 25+ 8065752415Alcon Laboratories, Inc. · Unit, phacofragmentation369 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page