Bausch + Lomb BL5425WVRecalls & reported problems

Bausch & Lomb Incorporated · TSV 25ga Posterior Vitrectomy Pack with Valves and AFI and Wide Field Illuminator · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00757770058648), received between Jan 25, 2017 and Jul 3, 2025 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Cut9
  • Aspiration Issue1
  • Particulates1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Known Impact Or Consequence To Patient3
  • Insufficient Information2
  • Retinal Injury1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 3, 2025Failure to Cut; Failure to Cut

    A USER FACILITY IN CHINA REPORTED THAT DURING AN OPERATION THE VITRECTOMY CUTTER DID NOT CUT WELL, AND CONTINUED TO ASPIRATE. THE CUTTER WAS REPLACED. THERE WAS PATIENT CONTACT, AND NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22395744 ↗
  • MalfunctionJun 30, 2025Failure to Cut; Failure to Cut

    THE USER FACILITY IN CHINA REPORTED DURING AN OPERATION, THE VITRECTOMY CUTTER WAS NOT SHARP ENOUGH TO CUT WELL. WHILE THE CUTTER WAS STOPPED TO BE REPLACED, ASPIRATION FUNCTION CONTINUED AT A SIGNIFICANTLY REDUCED RATE, WITH MANY BUBBLES IN THE SUCTION PIPE. THERE WAS CONTACT WITH THE PATIENT'S EYE, BUT NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22360864 ↗
  • MalfunctionMar 11, 2025Failure to Cut; Failure to Cut

    A USER FACILITY IN CHINA REPORTS THAT THE VITRECTOMY CUTTER IS NOT CUTTING WHILE STILL ASPIRATING DURING THE OPERATION. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT/INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21577296 ↗
  • MalfunctionJan 8, 2025Failure to Cut; Failure to Cut

    THE USER FACILITY IN CHINA REPORTED THAT THE VITRECTOMY CUTTER WAS NOT CUTTING YET WAS STILL ASPIRATING DURING SURGERY. THERE WAS NO PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 21107350 ↗
  • MalfunctionSep 25, 2024Failure to Cut; Failure to Cut

    THE USER FACILITY IN CHINA REPORTED THE VITRECTOMY CUTTER CAN NOT CUT WHILE STILL ASPIRATING. NO IMPACT TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20300036 ↗
  • MalfunctionJan 10, 2024Failure to Cut; Failure to Cut

    THE USER FACILITY IN CHINA REPORTED A VITRECTOMY CUTTER WAS NOT CUTTING.

    Patient problems: Insufficient Information

    MAUDE report 18483568 ↗
  • MalfunctionDec 13, 2023Failure to Cut; Failure to Cut

    THE USER FACILITY IN CHINA REPORTS THAT THE VITRECTOMY CUTTER WAS NOT CUTTING. IT IS UNKNOWN IF ASPIRATION CONTINUED AND CUTTING STOPPED, OR IF THERE WAS ANY PATIENT IMPACT, AND WHAT THE OUTCOME IS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18319936 ↗
  • MalfunctionJun 8, 2018Particulates; Particulates

    THE DOCTOR REPORTED HEARING EXCESSIVE ¿CHIRPING¿ FROM THE VITRECTOR, AND OBSERVED WHAT APPEARED TO BE A WHITE METALLIC DUST COMING OUT OF THE VITRECTOR. THEY OPENED ANOTHER PACK. THE PARTICULATE WAS REMOVED AND SURGERY COMPLETED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7582075 ↗
  • MalfunctionSep 21, 2017Failure to Cut; Failure to Cut

    THE USER FACILITY IN (B)(6) REPORTED DURING SURGERY THE VITRECTOMY CUTTER DID NOT HAVE ADEQUATE CUTTING ACTION WHEN SET AT OR ABOVE 3500CPM. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6883673 ↗
  • InjuryJun 15, 2017Aspiration Issue; Aspiration Issue

    THE USER FACILITY REPORTED THE VITRECTOMY CUTTER WAS NOT ASPIRATING DURING SURGERY. THE SURGEON MOVED THE CUTTER CAUSING TRACTION ON THE RETINA RESULTING IN A RETINAL DETACHMENT.

    Patient problems: Retinal Injury

    MAUDE report 6642904 ↗
  • MalfunctionJan 25, 2017Failure to Cut; Failure to Cut

    CUSTOMER IS HAVING ISSUES WITH THE CUTTER STOPPING DURING SURGERY. A REPLACEMENT CUTTER WAS USED TO COMPLETE THE PROCEDURE. THE DATE OF EVENT IS NOT AVAILABLE, AND NO PRODUCT WILL BE RETURNED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6277567 ↗

Other Instrument, Vitreous Aspiration And Cutting, Ac-Powered devices

All Instrument, Vitreous Aspiration And Cutting, Ac-Powered devices →