Amvisc Plus 60081LRecalls & reported problems
Bausch & Lomb Incorporated · AMVISC PLUS 0.8ML · Class 3
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00757770495672), received between Aug 8, 2024 and Jun 18, 2026 — classified as 5 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Insufficient Device Problem Information2
Patient problems
- Endophthalmitis3
- Hypopyon3
- Loss of Vision3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Eye Pain2
- Swelling/ Edema2
- Toxic Anterior Segment Syndrome (TASS)2
- Blurred Vision1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJun 18, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT 2 DAYS AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) IN THE LEFT EYE, THE PATIENT PRESENTED WITH HYPOPYON AND VITRITIS. BASED ON THE FINDINGS, THE SURGEON CONCLUDED THE CAUSE AS ENDOPHTHALMITIS. IN THE IMPLANTING SURGEON'S OPINION, THE MOST LIKELY CAUSE OF THE EVENT COULD NOT BE DETERMINED. MEDICATION ADMINISTERED DURING ORIGINAL SURGERY: INTRACAMERAL MOXIFLOXACIN. THE FOLLOWING MEDICATIONS WERE PRESCRIBED FOR USE AT HOME POST-OPERATIVELY: PREDNISOLONE 1%: 1 GTT QID X 2 WEEKS, THEN TID X 2 WEEKS (TOTAL DURATION: 1 MONTH) OFLOXACIN 0.3%: 1 GTT QID X 7 DAYS KETOROLAC 0.5%: 1 GTT QID X 2 WEEKS, THEN TID X 2 WEEKS (TOTAL DURATION: 1 MONTH)
Patient problems: Endophthalmitis; Hypopyon
MAUDE report 25551864 ↗InjuryJun 18, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT 4 DAYS AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) IN THE LEFT EYE (OS), THE PATIENT PRESENTED COMPLAINING OF VISION LOSS AND EYE PAIN. SLIT LAMP EXAMINATION IDENTIFIED A FIBRIN REACTION WITH HYPOPYON. BASED ON THE FINDINGS THE SURGEON CONCLUDED THE CAUSE AS ENDOPHTHALMITIS. TWO DAYS LATER, THE PATIENT WAS EVALUATED BY A RETINA SPECIALIST WHO NOTED THE VIEW OF POSTERIOR SEGMENT WAS LIMITED DUE TO VITREOUS OPACITY. A VITREOUS TAP WAS PERFORMED, AND AN INJECTION OF VANCOMYCIN AND CEFTAZIDIME WAS ADMINISTERED. THE PATIENT'S BEST CORRECTED VISUAL ACUITY (BCVA) DECREASED TO HAND MOTION. IN THE IMPLANTING SURGEON'S OPINION, THE MOST LIKELY CAUSE OF THE EVENT COULD NOT BE DETERMINED. THE PATIENT¿S OUTCOME IS UNKNOWN, AS THEY HAVE NOT RETURNED FOR FOLLOW-UP AND CANNOT BE REACHED. THE FOLLOWING MEDICATIONS WERE PRESCRIBED FOR USE AT HOME POST-OPERATIVELY: PREDNISOLONE 1%: 1 GTT QID X 2 WEEKS, THEN TID X 2 WEEKS (TOTAL DURATION: 1 MONTH), OFLOXACIN 0.3%: 1 GTT QID X 7 DAYS, KETOROLAC 0.5%: 1 GTT QID X 2 WEEKS, THEN TID X 2 WEEKS (TOTAL DURATION: 1 MONTH).
Patient problems: Endophthalmitis; Eye Pain; Hypopyon; Loss of Vision; Vitreous Floaters
MAUDE report 25550059 ↗InjuryOct 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE HEALTHCARE FACILITY REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANTATION IN THE RIGHT EYE, THE PATIENT HAD PAIN AND BLURRED VISION. TWO (2) DAYS POST PROCEDURE, THE PATIENT PRESENTED FOR EMERGENCY CARE AT ANOTHER HEALTHCARE FACILITY WITH SEVERE PAIN AND SYMPTOMS OF ENDOPHTHALMITIS. THE PATIENT WAS HOSPITALIZED AND ON THE SECOND DAY, A VITRECTOMY WITH SILICONE OIL ADMINISTRATION WAS PERFORMED. A MICROBIOLOGICAL CULTURE IDENTIFIED ENTEROCOCCUS FAECALIS INFECTION. ANTIBIOTIC THERAPY (INCLUDING VANCOMYCIN) WAS INITIATED. FOUR (4) DAYS POST-SECONDARY PROCEDURE, THE PATIENT WAS DISCHARGED FROM THE HOSPITAL WITH PRESERVED VISION OF 1/50, WITHOUT PAIN AND THE IOL REMAINS IMPLANTED.
Patient problems: Blurred Vision; Endophthalmitis; Eye Pain
MAUDE report 23435574 ↗InjuryAug 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE ONE-DAY POST-OPERATIVE VISIT AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO THE RIGHT EYE, THE SURGEON NOTED THE PATIENT PRESENTED WITH 1-2+ EDEMA, SEIDEL SIGN, AND CELL 2+ FIBRIN IN AC. THIS REACTION WAS DIAGNOSED AS TOXIC ANTERIOR SEGMENT SYNDROME (TASS). THE PATIENT'S BCVA DECREASED FROM 20/150 TO CF @ 1 FT. THE DAY AFTER SYMPTOMS WERE OBSERVED, AN INTRAVITREAL INJECTION OF VANCOMYCIN & CEFTAZIDIME WAS PERFORMED, AND A CULTURE WAS TAKEN. THE RESULT OF THE CULTURE WAS NEGATIVE. IN THE SURGEON'S OPINION, THE MOST LIKELY CAUSE OF EVENT IS POSSIBLY FROM THE VISCO AND LIDO/PHEN. PATIENT OUTCOME IS GOOD. THE FOLLOWING MEDICATIONS WERE PRESCRIBED FOR USE AT HOME POSTOPERATIVELY: PRED-MOXI-BROM 1%, .5%, 0.075% DOSE: 1GTT FREQUENCY: TID 2 WEEKS, BID 1 WEEK, QD 1 WEEK NEOMYCIN & POLYMYXIN & DEXAMETHASONE DOSE: 1 SMALL RIBBON FREQUENCY: ONCE PREDNISOLONE 1% DOSE:1GTT FREQUENCY: QHOURLY DURATION: UNTIL DIRECTED OTHERWISE
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Loss of Vision; Swelling/ Edema; Toxic Anterior Segment Syndrome (TASS)
MAUDE report 19940749 ↗InjuryAug 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE ONE-DAY POST-OPERATIVE VISIT AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO THE LEFT EYE, THE SURGEON NOTED THE PATIENT PRESENTED WITH 1+ EDEMA, 4+ FIBRIN, AND TRACE SLIVER OF HYPOPYON. THIS REACTION WAS DIAGNOSED AS TOXIC ANTERIOR SEGMENT SYNDROME (TASS). IN THE SURGEON'S OPINION, THE MOST LIKELY CAUSE OF EVENT IS POSSIBLY FROM THE VISCO AND LIDO/PHEN. PATIENT OUTCOME IS GOOD. THE FOLLOWING MEDICATIONS WERE PRESCRIBED FOR USE AT HOME POSTOPERATIVELY: NEOMYCIN & POLYMYXIN & DEXAMETHASONE DOSE: 1 SMALL RIBBON FREQUENCY: ONCE PRED-MOXI-BROM 1%, .5%, 0.075% DOSE: 1GTT FREQUENCY: TID 2 WEEKS, BID 1 WEEK, QD 1 WEEK DUREZOL DOSE: 1GTT FREQUENCY: QHOURLY DURATION: TILL DIRECTED OTHERWISE. MOXIFLOXACIN DOSE: 1GTT FREQUENCY: QHOURLY DURATION: TILL DIRECTED OTHERWISE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Hypopyon; Loss of Vision; Swelling/ Edema; Toxic Anterior Segment Syndrome (TASS)
MAUDE report 19940447 ↗
FDA clearance and classification
The FDA regulates product code LZP (aid, surgical, viscoelastic) under 21 CFR 886.4275 as a Class III device. 21 CFR 886.4275 ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Amvisc Plus 60081L
Has the Amvisc Plus 60081L been recalled?
No FDA recall record in our data names this exact device identifier (DI 00757770495672), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Amvisc Plus 60081L?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Aug 8, 2024 and Jun 18, 2026. In those reports, the most common patient problems are “Endophthalmitis” (3), “Hypopyon” (3) and “Loss of Vision” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Amvisc Plus 60081L?
None of the reports are classified as deaths, 5 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Amvisc Plus 60081L, and is it still sold?
The FDA's device database lists Bausch & Lomb Incorporated as the company responsible for it (brand Amvisc Plus, model 60081L). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Aid, Surgical, Viscoelastic devices
- HEALON ENDOCOAT PRO 10300012Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC1 recall1 report
- HEALON DUET PRO II 10300016Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC1 recall0 reports
- HEALON ENDOCOAT VT585UJohnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC22 reports
- HEALON 5 10290051Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC20 reports
- HEALON GV PRO 10240014Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC15 reports
- Amvisc 59081LBausch & Lomb Incorporated · Aid, Surgical, Viscoelastic10 reports
- DUOVISC 8065183150Alcon Laboratories, Inc. · Aid, Surgical, Viscoelastic9 reports
- HEALON DUET PRO 10240016Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC9 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record