Stellaris BL1433Recalls & reported problems

Bausch & Lomb Incorporated · Stellaris PC Combined System · Class 2

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00757770497836), received between Jul 19, 2021 and Apr 14, 2026 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Communication or Transmission Problem1
  • Decrease in Suction1
  • Power Problem1
  • Pressure Problem1
  • Unexpected Shutdown1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Eye Injury1
  • Inflammation1
  • Insufficient Information1
  • Intraocular Pressure Increased1
  • Prolapse1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 14, 2026Power Problem

    THE USER FACILITY IN RUSSIA REPORTS EXPERIENCING AN UNEXPECTED POWER INTERRUPTION TWO TO FOUR TIMES A DAY. IT WAS REPORTED THERE WAS NO MEDICAL INTERVENTION/TREATMENT REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24873465 ↗
  • MalfunctionNov 11, 2024Unexpected Shutdown; Unexpected Shutdown

    A USER FACILITY IN THE CANARY ISLANDS REPORTED THAT THE SYSTEM SHUTDOWN BY ITSELF DURING SURGERY. THERE WAS NO REPORTED PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20657036 ↗
  • InjuryDec 20, 2023Decrease in Suction; Decrease in Suction

    THE USER FACILITY IN FRANCE REPORTED THAT ASPIRATION WAS LOST DURING THE PROCEDURE. THE PATIENT WAS MOVED TO A DIFFERENT OPERATION ROOM WHERE ANOTHER SYSTEM WAS USED TO COMPLETE THE SURGERY. THE PROCEDURE WAS EXTENDED BY 55 MINUTES. CURRENTLY, THERE IS NO IMPACT ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18364991 ↗
  • InjurySep 28, 2023Pressure Problem; Pressure Problem

    IT WAS REPORTED BY THE USER FACILITY IN THE UNITED KINGDOM THAT DURING THE CASE THE EYE BECAME HARD DUE TO HIGH PRESSURE CAUSING THE IRIS TO PROLAPSE. WHEN CHECKED THE INFUSION PRESSURE WAS SET TO 35MMHG. THE IRIS WAS REDEPOSITED SUCCESSFULLY AND THE ANTERIOR CHAMBER REFORMED. SURGERY WAS COMPLETED SAFELY, BUT THE IOP FLUCTUATIONS DID COMPROMISE THE CORNEAL VIEW TO SOME EXTENT MAKING THE LATER STAGES OF THE MEMBRANE PEEL MORE CHALLENGING. I WASN¿T TIMING THE SURGERY, BUT THE VITRECTOMY CUTTER IS INEFFICIENT AND THE GEL TRIMMING TOOK LONGER THAN IT SHOULD HAVE. A SMALL VOLUME OF ANESTHESIA TOP-UP WAS NEEDED. APART FROM A SLIGHTLY DILATED PUPIL, EVEN FOLLOWING MIOCHOL-E, THE EYE IS QUIET WITH A NORMAL IOP.

    Patient problems: Eye Injury; Intraocular Pressure Increased; Prolapse

    MAUDE report 17839341 ↗
  • InjuryMay 9, 2023Communication or Transmission Problem; Communication or Transmission Problem

    A USER FACILITY IN THE UNITED KINGDOM REPORTED THAT THEY TRIED TO USE THE FOOT PEDAL AND HAD NO RESPONSE FROM IT DURING SEGMENT REMOVAL. THE OPERATION WAS STOPPED AND THE PATIENT WAS SENT BACK TO WARD FROM THEATRE WITH ONLY THE CAPSULOROHEXIS HAVING BEEN DONE AND PHACO NOT STARTED. THIS CAUSED SOME CORNEAL INFLAMMATION. THE PATIENT WAS NOT LEFT APHAKIC. AFTER A 90 MINUTE DELAY, WHEN THE SYSTEM WAS WORKING AGAIN THE OPERATION WAS COMPLETED THE SAME AFTERNOON. AN ADDITIONAL TOP UP OF TOPICAL ANESTHESIA WAS GIVEN. A FOLLOW UP WITH THE PATIENT SHOWS THEY ARE DOING WELL WITH NO CONCERNS REGARDING VISUAL OUTCOME.

    Patient problems: Inflammation; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16899284 ↗
  • InjuryJul 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A USER FACILITY IN (B)(6) REPORTED THAT DURING SURGERY THE ASPIRATION DID NOT WORK WITH THE ANTERIOR VITRECTOMY. A NEW VITRECTOMY WAS TRIED AND THE ASPIRATION STILL DID NOT WORK. THE PROCEDURE WAS PROLONGED BY TWO (2) HOURS AND ADDITIONAL ANESTHESIA WAS REQUIRED. THE PATIENT WAS TRANSFERRED TO ANOTHER HOSPITAL TO COMPLETE THE SURGERY. THE PATIENT IS CURRENTLY RECOVERING.

    Patient problems: Insufficient Information

    MAUDE report 12191820 ↗

Other Instrument, Vitreous Aspiration And Cutting, Ac-Powered devices

All Instrument, Vitreous Aspiration And Cutting, Ac-Powered devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.