Stellaris BL14334Recalls & reported problems
Bausch & Lomb Incorporated · Stellairs PC Combined System with Version 4+ Software · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00757770519996), received between Nov 22, 2024 and Jun 19, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Unexpected Shutdown” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Unexpected Shutdown2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 19, 2025Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED BY THE USER FACILITY IN MALAYSIA THAT THE SYSTEM WAS SHOWING VFM14 DURING THE PROCEDURE HOWEVER FLUID LEVEL IN CASSETTE WAS NOT FULL. WRONGLY DETECTED FLUIDIC LEVEL EVEN AFTER A NEW CASSETTE WAS INSERTED. THE FIELD SERVICE ENGINEER REPORTED THAT THE SYSTEM STOPPED UNINTENTIONALLY AFTER THE ERROR. THERE WAS A MINIMAL DELAY TO RESTART AND REINSERT THE CASSETTE. NO MEDICAL INTERVENTION OR TREATMENT WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292074 ↗MalfunctionNov 22, 2024Unexpected Shutdown; Unexpected Shutdown
THE USER FACILITY IN SLOVENIA REPORTED THAT DURING VITRECTOMY, THE INFUSION DID NOT FLOW. THE DEVICE DID NOT RECOGNIZE THE CASSETTE AND NOTICED THAT IT WAS FULL, HOWEVER, IT WAS NOT. TOWARDS THE END OF THE PROCEDURE, THE SYSTEM HAD SHUT DOWN. THERE WAS NO IMPACT OR HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20762321 ↗
Other Instrument, Vitreous Aspiration And Cutting, Ac-Powered devices
- CONSTELLATION 8065751150Alcon Laboratories, Inc. · Instrument, vitreous aspiration and cutting, ac-powered1 recall1,246 reports
- CONSTELLATION 8065753041Alcon Laboratories, Inc. · Instrument, vitreous aspiration and cutting, ac-powered503 reports
- Centurion, OZil 8065751761Alcon Laboratories, Inc. · Instrument, vitreous aspiration and cutting, ac-powered475 reports
- UNITY 00380650002969Alcon Laboratories, Inc. · Instrument, Vitreous Aspiration And Cutting, Ac-Powered271 reports
- Ozil 8065750469Alcon Laboratories, Inc. · Instrument, vitreous aspiration and cutting, ac-powered261 reports
- Infiniti 8065752160Alcon Laboratories, Inc. · Instrument, vitreous aspiration and cutting, ac-powered54 reports
- Bausch + Lomb SE5523WVBausch & Lomb Incorporated · Instrument, Vitreous Aspiration And Cutting, Ac-Powered1 recall4 reports
- Bausch + Lomb SE5525WVBausch & Lomb Incorporated · Instrument, Vitreous Aspiration And Cutting, Ac-Powered1 recall2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.