Skip to content

Vectris™ 977D160Recalls & reported problems

Medtronic, Inc. · LEAD 977D160 TRIALING MRICS SBCMPCT EMAN · Class 3 · Implant

All models of Vectris (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000006563), received between Jul 27, 2021 and Jul 27, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “High impedance” (1 report).

9 other recalls cover devices cleared under the same FDA submission (P840001); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • High impedance1
  • Impedance Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 27, 2021High impedance; Impedance Problem; High impedance; Impedance Problem

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING A TRIAL PATIENT WHO WAS USING AN EXTERNAL NEUROSTIMULATOR (ENS) FOR UNKNOWN INDICATIONS FOR USE. IT WAS REPORTED THAT REP REPORTS HCP DOING A TRIAL THIS MORNING BUT REP GOT OOR, AND IMPEDANCE SEEMS TO STAY INCONSISTENTLY HIGH. REP SAID INITIALLY CONNECTIVITY WAS ALL RED AS WELL AS IMPEDANCE CHECK, HCP SWITCH LEAD TO ANOTHER WENS CHANNEL BUT GOT THE SAME RESULT. HCP THEN USED ANOTHER WENS BUT GOT THE SAME RESULT AS WELL, EVENTUALLY A COUPLE ELECTRODES CAME BACK WITH IMPEDANCE AROUND 2300 OHMS. HCP THEN USED ANOTHER LEAD AND HE GOT SIMILAR RESULTS AGAIN. REP SAID HCP WOULD MOVE THE LEAD WITHIN THE WENS EVER SO SLIGHTLY AND IMPEDANCE WOULD CHANGE, AND AT ONE POINT HCP GOT UP TO 3 ELECTRODES TO COME IN THAT ISN'T RED. REP SAID HCP JUST NEVER GOT RELIABLE CONNECTIVITY. REP SAID HE WAS ABLE TO GET SOME STIMULATION DOWN THE LEG. HCP EVENTUALLY DECIDED THE IMPEDANCE WASN'T CONSISTENT ENOUGH AND HE ABANDONED THE TRIAL. REP SAID HCP USED LOSS OF RESISTANCE WITH AIR TO INSERT THE NEEDLE. DISCUSSED HOW AIR CAN CAUSE IMPEDANCE TO FLUCTUATE. REP DID RUN STIMULATION BUT THERE WAS NOT ENOUGH CHANGE IN IMPEDANCE. REP TO SEND WENS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12229657 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P840001. One submission often covers many models, so check the product and lot details in each record.

  • Z-0372-2025 · Sep 18, 2024 · Open, Classified
    Medtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used wi
  • Z-1691-2024 · Mar 28, 2024 · Open, Classified
    Intellis AdaptiveStim Product Number 97715, Spinal Cord Stimulator
  • Z-0780-2024 · Dec 19, 2023 · Open, Classified
    Model A710 Intellis Clinician Programmer Application, version 2.0.97
  • Z-2159-2021 · Jul 9, 2021 · Open, Classified
    Medtronic Vectris SureScan MRI, Catalog Numbers: a) 977A260 b) 977A275 c) 977A290
  • Z-1926-2016 · May 3, 2016 · Terminated
    Medtronic RestoreSensor, Model 37714; and, RestoreSensor SureScan, (MRI), Model 97714, Implantable Neurostimulators. Used to treat chronic, intractable pain of the trunk and/or li
  • Z-1147-2015 · Jan 9, 2015 · Terminated
    The Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or m
  • Z-1524-2014 · Feb 27, 2014 · Terminated
    Medtronic neurostimulators. Restore (37711), RestoreADVANCED (37713), RestoreADVANCED SureScan MRI (97713), RestoreSENSOR (37714), RestoreSENSOR SureScan MRI (97714), PrimeAdvanced
  • Z-0278-2014 · Sep 30, 2013 · Terminated
    Medtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation.
  • Z-1536-06 · Aug 2, 2006 · Terminated
    Medtronic 3777 Lead Kit, 1x8 Low Impedance Lead Kit for Spinal Cord Stimulation (SCS). Manufactured at: Medtronic, Inc., Villalba, Puerto Rico USA. Medtronic, Inc., Minneapolis

Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices

All STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices →