CapSureFix Novus MRI™ SureScan™ 407652 (UDI 00763000014643)Recalls & reported problems

Medtronic, Inc. · LEAD 407652 CAPSUREFIX MRI US RCMCRD · Class 3 · Implant

All models of CapSureFix Novus MRI SureScan (29)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a Medtronic CapSureFix Novus MRI SureScan 407652 lead, a permanent pacemaker electrode implanted in the heart. The 27 reports mostly describe the lead under-sensing or over-sensing the heartbeat signal, sometimes causing the pacemaker to pace when it shouldn't or to lose capture; leads also moved out of position. Two reports describe death and 12 describe injury. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
27
Deaths reported
2
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 00763000014643), received between Dec 7, 2017 and Aug 13, 2026 — classified as 2 deaths, 12 injuries, 13 malfunctions.

The most frequently reported specific device problem is “Under-Sensing” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Under-Sensing10
  • Over-Sensing7
  • Device Dislodged or Dislocated5
  • Adverse Event Without Identified Device or Use Problem2
  • Failure to Capture2
  • High Capture Threshold2
  • Pacing Problem2
  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • No Known Impact Or Consequence To Patient13
  • No Clinical Signs, Symptoms or Conditions7
  • Bacteremia1
  • Bradycardia1
  • Cardiac Arrest1
  • Cardiac Perforation1
  • Dyspnea1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 13, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced bacteraemia. The Implantable Pulse Generator (IPG) system was removed. No further patient complications have been reported as a result of this event.

    Patient problems: Bacteremia

    MAUDE report 50232429 ↗
  • MalfunctionJun 25, 2026Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING. THE PATIENT EXPERIENCED SYNCOPE AND WAS STATED TO BE IN-PATIENT. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Syncope/Fainting

    MAUDE report 25630518 ↗
  • MalfunctionJun 22, 2026Pacing Problem; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING RESULTING IN INAPPROPRIATE PACING ALONG WITH POSSIBLE IN TERMITTENT LOSS OF CAPTURE. IT WAS ALSO REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) RESULTING IN UNNECESSARY PACING AT TIME. BOTH, THE RV AND RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25573986 ↗
  • DeathNov 6, 2024Under-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS UNDERSENSING IN ELECTROGRAMS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20622674 ↗
  • MalfunctionOct 9, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED AN ATRIAL LEAD POSITION CHECK FAILURE. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20406114 ↗
  • MalfunctionSep 2, 2024Over-Sensing; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED INTERMITTENT FAR FIELD R-WAVE (FFRW) OVERSENSING ON STORED ELECTROGRAMS (EGM), RESULTING IN INAPPROPRIATE MODE SWITCH, ATRIAL EVENTS APPEAR TO BE FALLING IN BLANKING AT TIMES POSSIBLY TRIGGERING ATRIAL TAC HYCARDIA/ATRIAL FIBRILLATION (AT/AF) DEVICE CLASSIFIED TERMINATION OF AT/AF DETECTED EVENT(S) AND DOES NOT APPEAR TO IMPACT THE DEVICE FUNCTION OR PERFORMANCE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20117278 ↗
  • MalfunctionAug 27, 2024Over-Sensing; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING ON THE HIGH RATE EPISODES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20074247 ↗
  • DeathFeb 18, 2022Intermittent Capture; Intermittent Capture

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED BRADYCARDIA. IT WAS FURTHER REPORTED THAT INTERMITTENT VENTRICULAR CAPTURE OF THE RIGHT VENTRICULAR (RV) LEAD WAS NOTED ON PRESENTING ELECTROGRAMS (EGM) AND THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING. THE RV AND RA LEADS REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bradycardia; Cardiac Arrest

    MAUDE report 13557296 ↗
  • MalfunctionDec 18, 2020Under-Sensing; Under-Sensing

    IT WAS FURTHER REPORTED THAT ATRIAL UNDERSENSING WAS TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENT(S) AT TIMES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11035790 ↗
  • InjuryJun 25, 2020Under-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNRELIABLE SENSING, UNDER-SENSING, LOW/SMALL P WAVES AND POSSIBLE INAPPROPRIATE TRACKING. THE RA LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10194666 ↗
  • MalfunctionJan 3, 2020Over-Sensing; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9544716 ↗
  • MalfunctionDec 16, 2019Over-Sensing; Over-Sensing

    IT WAS FURTHER REPORTED THAT THE SHORT V-V INTERVALS OCCURRED DURING A PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9473111 ↗
  • InjuryApr 30, 2019Over-Sensing; Pacing Problem; High Capture Threshold; Over-Sensing; Pacing Problem; High Capture Threshold

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED SYNCOPE. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS AND OVERSENSING RESULTING IN PACING INHIBITION. THE RV LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Syncope

    MAUDE report 8563663 ↗
  • MalfunctionDec 19, 2018Under-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD HAS UNDERSENSING NOTED ON STORED ELECTROCARDIOGRAMS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8182553 ↗
  • MalfunctionSep 26, 2018Under-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7912180 ↗
  • MalfunctionMay 10, 2018Over-Sensing; Under-Sensing; Over-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING AND UNDERSENSING DURING EPISODES OF VENTRICULAR TACHYCARDIA (VT) AND NON-SUSTAINED VENTRICULAR TACHYCARDIA WHICH OCCURRED DURING MITRAL VALVE REPAIR SURGERY. ALL LEAD MEASUREMENTS WERE WITHIN NORMAL LIMITS AT A SUBSEQUENT INTERROGATION. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7503469 ↗
  • InjuryMay 9, 2018Failure to Capture; Device Dislodged or Dislocated; Failure to Capture; Device Dislodged or Dislocated

    IT WAS REPORTED THAT POST-OPERATIVELY, THE RIGHT ATRIAL (RA) LEAD BECAME DISLODGED AND WAS CONFIRMED BY X-RAY TO BE IN THE RIGHT VENTRICLE. NO CAPTURE WAS EXHIBITED IN THE ATRIUM. THE RA LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7499627 ↗
  • InjuryMay 4, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) AND RIGHT ATRIAL (RA) LEADS DISLODGED ONE DAY AFTER IMPLANT. IT WAS NOTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) DEVICE WAS NOT SUTURED AND MOVED A SIGNIFICANT AMOUNT, WHICH CAUSED THE LEADS TO DISLODGE. THE RV LEAD WAS EXPLANTED AND REPLACED. THE RA LEAD WAS REPOSITIONED AND REMAINS IN USE. THE IPG DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7489166 ↗
  • InjuryApr 25, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT COMPLAINED OF DYSPNEA ON EXERTION, AND FURTHER EXAMINATION REVEALED POSSIBLE POOR BLOOD FLOW ACROSS THE TRICUSPID VALVE DUE TO THE PLACEMENT OF THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD WAS EXPLANTED AND NOT REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Dyspnea; Tricuspid Insufficiency

    MAUDE report 7459698 ↗
  • MalfunctionApr 10, 2018Failure to Sense; Under-Sensing; Failure to Sense; Under-Sensing

    IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THERE WAS NO SENSING ON THE RIGHT ATRIAL (RA) LEAD. THE LEAD WAS ATTEMPTED AND NOT USED. ANOTHER LEAD WAS USED. IT WAS ALSO REPORTED UPON SLITTING THE PHYSICIAN LOST THE SLITTER ENGAGEMENT OF THE CATHETER. WHEN RE-ENGAGING THE SLITTER AND ATTEMPTING TO SLIT THE CATHETER PORTION CONTAINING THE HUB SLIT OFF AND THERE WAS NO "HANDLE" LEFT TO PULL ON THE CATHETER. ANOTHER CATHETER WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7416289 ↗
  • InjuryApr 10, 2018Device Disinfection Or Sterilization Issue; Device Disinfection Or Sterilization Issue

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD VEGETATION. THE LEADS WERE EXTRACTED AND REPLACED WITH A TEMPORARY LEAD. THE DEVICE REMAINS IN USE AND THE SYSTEM WILL BE REPLACED IN THE FUTURE WITH A LEADLESS IMPLANTABLE PULSE GENERATOR (IPG). NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 7416184 ↗
  • InjuryApr 10, 2018Failure to Capture; Adverse Event Without Identified Device or Use Problem; Failure to Capture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED NO CAPTURE. IT WAS DETERMINED THAT THE LEAD HAD PERFORATED THE PATIENT'S CARDIAC TISSUE RESULTING IN A PERICARDIAL EFFUSION. THE LEAD WAS REPOSITIONED AND REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Cardiac Perforation; Pericardial Effusion

    MAUDE report 7415208 ↗
  • InjuryApr 10, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected

    IT WAS REPORTED BY THE PATIENT THAT THEIR LEAD WAS READJUSTED ONE WEEK AFTER IT WAS IMPLANTED. NO INFORMATION COULD BE OBTAINED THROUGH FOLLOW-UP ATTEMPTS WITH THE CLINIC. THE RIGHT ATRIAL (RA) LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7415062 ↗
  • InjuryFeb 8, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING POST IMPLANT, THE RIGHT ATRIAL (RA) LEAD DISLODGED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7255622 ↗
  • MalfunctionFeb 8, 2018Signal Artifact/Noise; Over-Sensing; Under-Sensing; Signal Artifact/Noise; Over-Sensing; Under-Sensing

    IT WAS REPORTED THAT DURING IMPLANT OF THE RIGHT VENTRICULAR (RV) LEAD, THE LEAD EXHIBITED NOISE AND UNDERSENSING. THE LEAD WAS REMOVED AN AN ALTERNATIVE LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7255099 ↗

Other permanent pacemaker Electrode devices

All permanent pacemaker Electrode devices →