KYPHON Xpander KyphoPak Tray KPT1002Recalls & reported problems

Medtronic Sofamor Danek, Inc. · KIT KPT1002 KYPHOPAK TRAY 10/3 FF W/OIS · Class 2

All models of KYPHON Xpander KyphoPak Tray (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00763000031657), received between Mar 3, 2022 and Aug 26, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (4 reports).

5 other recalls cover devices cleared under the same FDA submission (K041454); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture4
  • Leak/Splash3
  • Device Damaged Prior to Use1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Extravasation1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2024Leak/Splash; Device Damaged Prior to Use; Leak/Splash; Device Damaged Prior to Use

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING BKP FOR L2 VERT EBRAL COMPRESSION FRACTURE FOR PRIMARY OSTEOPOROSIS. IT WAS REPORTED THAT THE 3-WAY STOPCOCK WAS DAMAGED WHEN THE KIT WAS UNBOXED. HOWEVER, SETTING WAS PERFORMED BECAUSE THERE SEEMED TO BE NO PROBLEMS WITH THE IBT ITSELF. THERE WERE NO PROBLEMS WITH NEGATIVE PRESSURE AT THE TIME OF SETTING, BUT CONTRAST AGENT LEAKAGE WAS CONFIRMED FROM THE CONNECTOR OF THE TIP OF THE INFLATION SYRINGE AND THE IBT CONNECTION WHEN THE IBT WAS PLACED IN THE VERTEBRAL BODY AND EXTENDED. A NEW KPT1002 WAS UNBOXED AND ANOTHER SET ASSEMBLED, AND BALLOON INFLATION WAS RESUMED AFTER A DELAY OF ABOUT 5 MINUTES. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20062016 ↗
  • MalfunctionJul 31, 2023Leak/Splash; Material Rupture; Leak/Splash; Material Rupture

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY FOR PRE-OP DIAGNOSIS OF OSTEOPOROSIS WITH COMPRESSION FRACTURE. IT WAS REPORTED THAT INFLATABLE BONE TAMP WAS RUPTURED DURING INFLA TION. CONTRAST LEAKAGE WAS OBSERVED, SO THE USE OF THE CONTRAST MEDIUM WAS STOPPED, AND AFTER SALINE RINSING WAS PERFORMED. THE CONTRAST MEDIUM WAS REPLACED WITH AN ALTERNATIVE AND THE PROCEDURE WAS CONTINUED. THE PRESSURE IN IBT WAS 240PSI BEFORE IT RUPTURED.  THE PROCEDURE OR TECHNIQUE PERFORMED WAS TH10 BALLOON KYPHOPLASTY. NO TIME DELAY IN PROCEDURE REPORTED AS A RESULT. THERE WAS NO FRAGMENT OF RUPTURED BALLOON WAS LEFT INSIDE PATIENT BODY. THERE WERE NO PATIENT SYMPTOMS REPORTED. NO ADDITIONAL TREATMENT OR SURGERY PERFORMED AS A RESULT. NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17422905 ↗
  • MalfunctionMar 1, 2023Material Rupture; Material Rupture

    INFORMATION WAS RECEIVED FROM A HEALTHCARE FACILITY (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING SPINAL PRODUCT WHICH IS USED IN T12 BALLOON KYPHOPLASTY (BKP) SPINAL THERAPY FOR PRIMARY OSTEOPOROSIS. IT WAS REPORTED THAT WHEN THE BALLOON WAS INFLATED AND THE STYLET WAS REMOVED TO TAKE A FRONTAL VIEW WITH THE C-ARM, CONTRAST MEDIA STARTED LEAKING FROM THE CONNECTION PART, POSSIBLY CAUSING BALLOON BREAKAGE AND THE BALLOON WAS RUPTURED, SO THE BALLOON WAS DEFLATED AND REMOVED. THERE WERE NO BROKEN FRAGMENTS LEFT INSIDE THE PATIENT AND NO FURTHER COMPLICATIONS OR SYMPTOMS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16458564 ↗
  • MalfunctionFeb 3, 2023Material Rupture; Material Rupture

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING BALLOON KYPHOPL ASTY DUE TO UNKNOWN INDICATION. IT WAS REPORTED THAT THE CONTRAST MEDIA LEAKED INSIDE THEVERTEBRAL BODY DURING BALLOON INFLATION (BALLOON BREAKAGE). THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: Extravasation

    MAUDE report 16289674 ↗
  • MalfunctionMar 3, 2022Leak/Splash; Material Rupture; Leak/Splash; Material Rupture

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH PRE-OP DIAGNOSIS- PRIMARY OSTEOPOROSIS, COMPRESSION FRACTURE INVOLVED IN THE BALLOON KYPHOPLASTY PROCEDURE. IT WAS REPORTED THAT INTRA-OPERATIVELY, DUE TO THE BREAKAGE OF THE 10M BALLOON, A NEW BALLOON KIT WAS OPENED AND INFLATED. THE LEAKED CONTRAST MEDIA WAS WASHED WITH SALINE AS MUCH AS POSSIBLE. IBT RUPTURED DURING INFLATION. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. THERE WAS NO DELAY IN OVERALL PROCEDURE TIME. THE PROCEDURE WAS PERFORMED SUCCESSFULLY BY USING A NEW PRODUCT. PATIENT WAS NOT HOSPITALIZED PROLONG AS A RESULT OF THIS EVENT. EVENT WAS ASSOCIATED WITH THE PATIENT. THE REPORTED PRODUCT WAS DISCARDED FOR THE PATIENT SUSPECTED OF BEING INFECTED. ON 2022-FEB-14, RECEIVED ADDITIONAL INFORMATION THAT CONTRAST MEDIA LEAKED INSIDE THE PATIENT'S BODY, WHEN CEMENT WAS INJECTED, THE CEMENT INJECTION WAS NOT SEEN CLEAR BECAUSE OF THE LEAKAGE. THIS PRODUCT CAME IN CONTACT WITH THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13657982 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K041454. One submission often covers many models, so check the product and lot details in each record.

  • Z-2389-2020 · Apr 24, 2020 · Terminated
    Extremities, Large Kit Assy, US, 15/3 - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Ki
  • Z-2390-2020 · Apr 24, 2020 · Terminated
    Medtronic Bone Void Filler, Small - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Kit c
  • Z-2391-2020 · Apr 24, 2020 · Terminated
    Medtronic Bone Void Filler, Large - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Kit c
  • Z-3175-2011 · Mar 23, 2011 · Terminated
    Kyphon brand KyphoPak Express Tray First Fracture, 15/2 with KyphX Osteo Introducer System; Catalog Number: KPE1003; Product is manufactured and distributed by Medtronic Spine LLC