KYPHON Xpander KyphoPak Tray KPT2002Recalls & reported problems
Medtronic Sofamor Danek, Inc. · KIT KPT2002 KYPHOPAK TRAY 20/3 FF W/OIS · Class 2
All models of KYPHON Xpander KyphoPak Tray (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000031763), received between Apr 16, 2021 and Jun 4, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
5 other recalls cover devices cleared under the same FDA submission (K041454); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Deflation Problem1
- Material Deformation1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 4, 2024Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING SPINAL PRODUCT THAT IS USED IN DECOMPRESSION AND ROD REPLACEMENT REVISION SURGERY FOR PRIMARY OSTEOPOROSIS. IT WAS REPORTED THAT AFTER BALLOON INFLATION, THE TRIGGER WAS SQUEEZED, AND THE HANDLE WAS FULLY PULLED, BUT IT WAS NOT COLLECTED, AND THERE WAS AN ISSUE. AFTER SEVERAL REPETITIONS, IT APPEARED TO HAVE DEFLATED. NO FURTHER COMPLICATIONS OR SYMPTOMS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THERE WERE NO FRAGMENTS LEFT INSIDE PATIENT BODY AND NO PATIENT SYMPTOMS REPORTED AS A RESULT. SINCE THE BALLOON DID NOT DEFLATE, IT SEEMS THAT IT WAS BROKEN WITH A CURETTE AND DEFLATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19457071 ↗MalfunctionSep 5, 2022Deflation Problem; Deflation Problem
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING BALLOON KYPHOP LASTY FOR IMPRESSION FRACTURE L2. IT WAS REPORTED THAT THE BALLOON WAS DILATED WITHOUT ANY PROBLEMS, BUT THE PRESSURE COULDN'T RETURN. THE SURGEON HAD TO TRY A NUMBER OF THINGS SO THAT THE BALLOON COULD BE REMOVED FROM THE VERTEBRAL BODY. THE PROCEDURE WAS COMPLETED WITH ORIGINAL PRODUCT. THERE WAS 30 MINS PROCEDURAL DELAY. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15355120 ↗MalfunctionApr 16, 2021Material Rupture; Material Deformation; Material Rupture; Material Deformation
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT SPINAL THERAPY FOR VFC. IT WAS REPORTED THAT DURING THE TREATMENT OF A VERTEBRAL FRACTURE OF L1, THE WORKING CANULA WAS FOUND DETERIORATED WITH THE PASSAGE OF ONE OF THE DM PROVIDED IN THE KIT. THE SURGEON HAD DIFFICULTY INTRODUCING THE FOLLOWING DM INTO THE WORKING CANULA. THIS DEFECTIVE WORKING CANULA DAMAGED THE KYPHOPLASTY BALLOON AND DID NOT ALLOW THE OPERATIVE PROCEDURE TO BE CARRIED OUT AS WELL AS POSSIBLE DUE THE LEAKAGE OF RADIOPAQUE PRODUCT AND THE IMPOSSIBILITY OF INFLATING THE BALLOON PROPERLY. THE DEFECTIVE WORKING CANULA DAMAGED THE KYPHOPLASTY BALLOON. THERE WAS PATIENT INVOLVED IN THE EVENT AND NO SYMPTOMS WERE REPORTED. ON (B)(6) 2021 (B)(4) (REP): ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THE EVENT OCCURRED ON (B)(6) 2021. THE SURGEON IS (B)(6). THE PATIENT IS (B)(6) YEARS OLD, (B)(6). ON (B)(6) 2021 (B)(4) (REP): ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THE LOT NUMBER IS 220866360. THE REPORTED CANNULA CAME IN CONTACT WITH THE PATIENT. THERE IS NO REVISION AND NO PATIENT COMPLICATIONS INVOLVED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11677268 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K041454. One submission often covers many models, so check the product and lot details in each record.
- Z-2390-2020 · Apr 24, 2020 · TerminatedMedtronic Bone Void Filler, Small - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Kit c
- Z-2391-2020 · Apr 24, 2020 · TerminatedMedtronic Bone Void Filler, Large - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Kit c
- Z-2389-2020 · Apr 24, 2020 · TerminatedExtremities, Large Kit Assy, US, 15/3 - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Ki
- Z-3175-2011 · Mar 23, 2011 · TerminatedKyphon brand KyphoPak Express Tray First Fracture, 15/2 with KyphX Osteo Introducer System; Catalog Number: KPE1003; Product is manufactured and distributed by Medtronic Spine LLC
- Z-1209-2009 · Feb 4, 2009 · TerminatedKyphX Xpander Inflatable Bone Tamp, size 15/3, Catalog Number K09A, Manufactured by Kyphon Sarl, Switzerland. For use in spinal surgery.
Other Arthroscope devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page