Midas Rex MR8 MR8-AF01-RRecalls & reported problems

Medtronic Ps Medical, Inc. · ATT MR8-AF01-R MR8 FTD F1 B-5 2.4MM RFRB · Class 2

All models of Midas Rex MR8 (41)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000063542), received between Jan 2, 2024 and Jun 19, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Naturally Worn” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Naturally Worn2
  • Detachment of Device or Device Component1
  • Material Twisted/Bent1
  • No Apparent Adverse Event1
  • Noise, Audible1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 19, 2026Naturally Worn

    IT WAS REPORTED THAT DURING SERVICING THE BEARINGS WERE RUNNING ROUGH . AT THE TIME OF REPORT NO PATIENT IMPACT IS BEEN OBSERVED .ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25563522 ↗
  • MalfunctionJun 10, 2026Naturally Worn

    IT WAS REPORTED THAT THE ATTACHMENT HAS BEARING FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25452815 ↗
  • MalfunctionMay 27, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT ATTACHMENT BEARING WAS LOOSE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25296106 ↗
  • MalfunctionJan 2, 2024Material Twisted/Bent; No Apparent Adverse Event; Noise, Audible; Material Twisted/Bent; No Apparent Adverse Event; Noise, Audible

    ON FOLLOW UP IT WAS REPORTED THAT THE ENGINEER INDICATED FAULTY BEARINGS WHICH HAD FAILED, NOT A BENT CRANIOTOME HEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18424227 ↗