Skip to content

Midas Rex MR8 MR8-AS07-RRecalls & reported problems

Medtronic Ps Medical, Inc. · ATT MR8-AS07-R MR8 7CM STRT 2.4MM REFURB · Class 2

All models of Midas Rex MR8 (41)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Midas Rex MR8 attachment is a surgical drill attachment used in bone procedures. The 59 reports mostly describe the attachment overheating, coming apart, vibrating, or its bearings wearing out and failing, with no patient harm reported in any of them. There are 48 reports in the last 12 months compared with 8 in the prior 12 months, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
62
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 62 adverse event reports that list this exact device (DI 00763000063597), received between Aug 28, 2024 and Sep 30, 2026 — classified as 62 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (26 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 4 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 2 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 5 reportsJun 2026: 35 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 3 reportsOct 2026: 0 reports20242025202635
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component26
  • Overheating of Device23
  • Vibration6
  • Corroded3
  • Naturally Worn3
  • No Apparent Adverse Event3
  • Difficult to Insert1
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions57
  • Insufficient Information5

Recent adverse event reports

The 25 most recent of 62 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026No Apparent Adverse Event

    It was reported that the customer had dissatisfaction with the product. There was no patient involvement or complications as a result of the event. Additional information received that loose bearing at tube distal was observed during evaluation .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942940 ↗
  • MalfunctionSep 29, 2026Insufficient Device Problem Information

    It was reported that pre-operatively the device was broken. There was no patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933201 ↗
  • MalfunctionSep 24, 2026Difficult to Insert

    It was reported that it was difficult to insert the tool / blade into the device during preventive maintenance (PM). There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884853 ↗
  • MalfunctionAug 27, 2026Overheating of Device

    It was reported that the device got hot. No additional information was provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50483791 ↗
  • MalfunctionJun 30, 2026Difficult to Remove

    IT WAS REPORTED THAT PRIOR TO USE BRISTLES STUCK IN THE ATTACHMENT. AT THE TIME OF REPORT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT MISALIGNED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25666634 ↗
  • MalfunctionJun 26, 2026Overheating of Device

    IT WAS REPORTED THAT THE DEVICE WAS HEATING UP DURING A PROCEDURE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25632195 ↗
  • MalfunctionJun 25, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE ATTACHMENT BEARING WAS LOOSE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25617785 ↗
  • MalfunctionJun 25, 2026Overheating of Device

    IT WAS REPORTED THAT THE DEVICE WAS OVERHEATING PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARINGS DAMAGED WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25617516 ↗
  • MalfunctionJun 25, 2026Vibration

    IT WAS REPORTED THAT DURING PRE-OP, THE ATTACHMENT WAS VIBRATING WHEN USING AND NOT DRILLING. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25617474 ↗
  • MalfunctionJun 24, 2026Overheating of Device

    IT WAS REPORTED THAT DURING TESTING, THE ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25605005 ↗
  • MalfunctionJun 24, 2026Naturally Worn

    IT WAS REPORTED THAT DURING USE BEARINGS WERE FAILED. AT THE TIME OF REPORT NO PATIENT WAS INVOLVED. ADDITIONAL INFORMATION RECEIVED THAT DURING INSPECTION BEARINGS FOUND TO BE LOOSE AND DAMAGED AT THE DISTAL TUBE .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25604047 ↗
  • MalfunctionJun 24, 2026Overheating of Device

    IT WAS REPORTED THAT THE ATTACHMENT WAS ALLEGED TO HAVE EXPERIENCED OVERHEATING DURING USE.IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25603740 ↗
  • MalfunctionJun 23, 2026Overheating of Device

    IT WAS REPORTED THAT THE DEVICE WAS OVERHEATING PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING CHIPPED WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25591566 ↗
  • MalfunctionJun 23, 2026Overheating of Device

    IT WAS REPORTED THAT DURING MAINTENANCE, THE ATTACHMENT WAS FOUND OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25590200 ↗
  • MalfunctionJun 23, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT ATTACHMENT SMALL RINGS CAME OUT WHILE DRILLING. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25589963 ↗
  • MalfunctionJun 23, 2026Overheating of Device

    IT WAS REPORTED THAT THE ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25589951 ↗
  • MalfunctionJun 19, 2026Corroded; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING INSPECTION THE BEARING FELL OUT/DETACHED AND THERE WAS RUST/CORROSION. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25561305 ↗
  • MalfunctionJun 19, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE ATTACHMENT BEARING WAS LOOSE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25558914 ↗
  • MalfunctionJun 16, 2026Loss of Power

    IT WAS REPORTED THAT THE DURING USE ATTACHMENT BALL BEARING WAS NOT WORKING PROPERLY. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25516737 ↗
  • MalfunctionJun 16, 2026Overheating of Device

    IT WAS REPORTED THAT PRIOR TO USE, THE ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25514766 ↗
  • MalfunctionJun 16, 2026Vibration; Naturally Worn

    IT WAS REPORTED THAT PRIOR TO USE, THE DEVICE WAS TAGGED FOR REPAIR DUE TO BEING DESCRIBED AS "WOBBLY" AND IN NEED OF NEW BEARINGS. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25514044 ↗
  • MalfunctionJun 15, 2026Overheating of Device; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING INSPECTION THE BEARING FELL OUT/WAS MISSING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25503503 ↗
  • MalfunctionJun 15, 2026Detachment of Device or Device Component

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25503244 ↗
  • MalfunctionJun 15, 2026Corroded; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING INSPECTION THE BEARING WAS DETACHED AND THERE WAS CORROSION. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25501203 ↗
  • MalfunctionJun 15, 2026Vibration

    IT WAS REPORTED THAT DURING TESTING, THE ATTACHMENT TOOL WOBBLE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25499135 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices

All DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page