NA CB10G17RRecalls & reported problems
Medtronic, Inc. · CUSTOM PACK CB10G17R E INFANT · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000076153), received between Jul 13, 2022 and Jul 13, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
10 other recalls cover devices cleared under the same FDA submission (K891687); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak2
- Misassembled During Installation2
Patient problems
- Hemorrhage/Blood Loss/Bleeding2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 13, 2022Fluid/Blood Leak; Misassembled During Installation; Fluid/Blood Leak; Misassembled During Installation
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED THAT TUBING IN RACE WAY OF PUMP STARTED TO LEAK. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. IT WAS ALSO REPORTED THAT THE AFFECTED TUBING WAS FOUND TO BE STANDARD TUBING, NOT TYGON TUBING AS THE SPECIFICATION SUGGESTS.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 15000595 ↗MalfunctionJul 13, 2022Fluid/Blood Leak; Misassembled During Installation; Fluid/Blood Leak; Misassembled During Installation
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED THAT THE RACEWAY TUBING WAS LEAKING. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. IT WAS ALSO REPORTED THAT THE AFFECTED TUBING WAS FOUND TO BE STANDARD TUBING, NOT TYGON TUBING AS THE SPECIFICATION SUGGESTS. THERE WAS NO AIR LEAK IN THE SYSTEM. BLOOD LOSS DUE TO THE LEAK WAS MINIMAL BUT NOT QUANTIFIED BY THE ACCOUNT. NO TRANSFUSION WAS NEEDED AS A RESULT OF THIS LEAK.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 14998807 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K891687. One submission often covers many models, so check the product and lot details in each record.
- Z-2495-2020 · Aug 8, 2018 · TerminatedExtra corporeal circuit with bio-active surface.
- Z-1023-2016 · Feb 4, 2016 · TerminatedMedtronic Custom Perfusion Tubing Pack, Rx, Sterile EO. Terumo Cardiovascular Systems (Terumo CVS) has distributed an Urgent Safety Alert regarding the CD I H/S Cuvette. The cuvet
- Z-2736-2015 · Aug 18, 2015 · TerminatedMEDTRONIC PERFUSION TUBING PACKS (with potentially affected Hemoconcentrators). Model # 7U65R2, 8C72R4, 6N21R2, BB7M38R11, TL2Q68R17, HY8R66R3, HY2U43R9, 8C72R4, BB5J59R19, 6B65R3
- Z-2105-2013 · Jul 17, 2013 · TerminatedMedtronic Custom Perfusion System with Carmeda BioActive Surface. Model Numbers: CB1D82R12 and CB5N73R7. Sterilized by Ethylene Oxide, Do no Reuse. Product Usage: Usage: Th
- Z-0618-2013 · Nov 19, 2012 · TerminatedMedtronic Intersept Custom Tubing Pack with or without coating ( Carmeda BioActive Surfact, Carmeda BioActive Surfact and Trillium Biosurface,or Balance Biosurface) with the follow
- Z-0813-2012 · Dec 14, 2011 · TerminatedCustom Packs/Mexico, Medtronic Cardio-Kit, Sterilized Using Ethylene Oxide, Nonpyrogenic, Rx Only. Model # DLP6N48R3 & DLP6N48R4. Medtronic Intersept Custom Tubing Pack with T
- Z-1920-2008 · May 9, 2008 · TerminatedMedtronic Extracorporeal Circuit with bio-active surface; Cardiopulmonary Bypass Oxygenator, Catalog Numbers: CB0B90R3, CB0L44R15, CB0X97R10, CB1V99R2, CB1Y06R14, CB1Z42R15, CB
- Z-1227-2007 · Aug 1, 2007 · TerminatedMedtronic Intersept 3608 Sterile tubing (8') 1/2 x 3/32 in. Quantity 10, Lot 7626621, Use by 2009-07-31. Manufacturer: Medtronic, Inc. Minneapolis, MN 55432-5604 USA. Diffe
- Z-0583-2007 · Jan 2, 2007 · TerminatedMedtronic 2T10R1 Intersept Custom Tubing Pack. Pack 2T10R1 consists of two PVC lines; one of them is 132 inch in length and is capped. The second line is 102 inch in length and
- Z-1489-04 · Sep 15, 2004 · TerminatedMedtronic DLP VAD Venous Cannula for Ventricular Assist, Models 95036 and CB95036.
Other OXYGENATOR, CARDIOPULMONARY BYPASS devices
- HLS Set BEQ-HLS 7050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls13 deaths90 reports
- Affinity Fusion® BB811Medtronic, Inc. · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls90 reports
- HLS Set BEQ-HLS 5050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls4 reports
- NA HLS Set Advanced 5.0, Small Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS3 recalls0 reports
- A.M.G. PMP Sterile A.M.G. Module PMP NO T.P. SterileEurosets Srl · Oxygenator, Cardiopulmonary Bypass2 deaths106 reports
- CAPIOX®FX CX*FX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass2 deaths102 reports
- CAPIOX®RX CX*RX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass1 death102 reports
- QUADROX-iD QUADROX-iD Adult, SOFTLINE Coating, without FilterMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.