MiniMed®670G MMT-1780KPKRecalls & reported problems
Medtronic Minimed, Inc. · PUMP MMT-1780KPK 670G V4.10 EN MG · Class 3
All models of MiniMed 670G (23) · MiniMed670G (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This is a Medtronic MiniMed 670G insulin pump. Its reports mostly describe high or low blood glucose, rapid battery drain, repeated battery alarms and connection problems. There are 28,789 reports, including 39 deaths and 5,925 injuries. Three recalls are open, covering dosing errors when the pump is raised or lowered, reduced battery life after drops, and a deteriorating battery cap that can cut power.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 28,789
- Deaths reported
- 39
- Injuries reported
- 5,925
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 28,789 adverse event reports that list this exact device (DI 00763000090203), received between Apr 16, 2018 and Jul 7, 2025 — classified as 39 deaths, 5,925 injuries, 22,825 malfunctions.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (7,942 reports).
3 FDA recall records name this device's identifier, 3 still open.
12 other recalls cover devices cleared under the same FDA submission (P160017); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 13, 2026Class IIOpen, Classified
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Z-1739-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Jul 31, 2024Class IOpen, Classified
Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.
Z-0003-2025 · Root cause: Device Design · Recall details · FDA record ↗
- May 23, 2022Class IIOpen, Classified
Due to battery cap deterioration, this may potentially result in an incomplete battery circuit (similar to when battery cap is not attached) and a loss of power/functionality of the insulin infusion pumps.
Z-1355-2022 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Patient Device Interaction Problem7,942
- Device Alarm System5,558
- Adverse Event Without Identified Device or Use Problem4,402
- No Apparent Adverse Event3,559
- Appropriate Device Problem Term/Code Not Available2,157
- Connection Problem1,772
- No Display/Image1,724
- Circuit Failure1,701
Patient problems
- No Consequences Or Impact To Patient18,741
- Hyperglycemia5,614
- No Clinical Signs, Symptoms or Conditions2,425
- Hypoglycemia2,327
- Elevated Ketones/ Diabetic Ketoacidosis396
- Blood Loss227
- Nausea144
- Vomiting131
Recent adverse event reports
The 25 most recent of 28,789 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED HIGH BLOOD GLUCOSE LEVELS AFTER A FLIGHT. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1780KPK. TROUBLESHOOTING WAS NOT PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS NOT RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 22409512 ↗MalfunctionJul 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED HYPOGLYCEMIA DURING THE FLIGHT. THE CUSTOMER ALSO REPORTED A CALIBRATION NOT ACCEPTED ALERT. THE BLOOD GLUCOSE VALUE AT THE TIME OF THE EVENT WAS 56 MG/DL. THE LOW BLOOD GLUCOSE WAS TREATED WITH CARB INTAKE. THE EVENT INVOLVED PRODUCT(S) MMT-7020A, MMT-1780KPK. TROUBLESHOOTING WAS PERFORMED. IT WAS UNKNOWN WHETHER THE CUSTOMER WAS USING THE AUTO MODE/SMART GUARD FEATURE AT THE TIME OF THE EVENT. IT WAS UNKNOWN WHETHER THE CUSTOMER WAS USING THE INSULIN PUMP SYSTEM WITHIN 48 HOURS OF THE REPORTED EVENT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. MMT-7020A NO PRODUCT RETURN IS REQUIRED FOR ANALYSIS. MMT-1780KPK NO PRODUCT RETURN IS REQUIRED FOR ANALYSIS.
Patient problems: Hypoglycemia
MAUDE report 22409502 ↗MalfunctionJul 3, 2025Battery Problem; Battery Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED RAPID BATTERY DEPLETION, WITH THE BATTERY LASTING LESS THAN ONE WEEK DESPITE USING NEW AA DURACELL ALKALINE BATTERIES. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1780KPK. TROUBLESHOOTING WAS PERFORMED WHICH INCLUDED REVIEWING ALARM HISTORY, VERIFYING PUMP SETTINGS, AND SUGGESTING MANUAL SCREEN TIMEOUT, BUT THE ISSUE PERSISTED. THE CUSTOMER WAS ADVISED TO REVERT TO THEIR BACKUP PLAN PER HEALTHCARE PROFESSIONAL'S INSTRUCTIONS AND PLACE THE PUMP IN STORAGE MODE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22392483 ↗MalfunctionJun 25, 2025Battery Problem; Battery Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED "REPLACE BATTERY NOW" ALERTS EVERY THREE DAYS AND OCCASIONAL "POWER LOSS" ALARMS, DESPITE USING DURACELL AND ENERGIZER BATTERIES. THE CUSTOMER ALSO REPORTED SHORT BATTERY LIFE; THE BATTERY LASTED ONLY FOR THREE DAYS. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1780KPK. TROUBLESHOOTING WAS PERFORMED WHICH INCLUDED REVIEWING ALARM HISTORY AND PERFORMING A PUMP RESTART IN A PREVIOUS CALL, BUT THE ISSUE PERSISTED. THE CUSTOMER WAS ADVISED TO MONITOR THE PUMP AND REVERT TO THEIR BACKUP PLAN PER HCP INSTRUCTIONS IF FURTHER ISSUES AROSE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22326213 ↗MalfunctionJun 25, 2025Computer Software Problem; Battery Problem; Computer Software Problem; Battery Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED REPEATED BATTERY FAILED ALARMS AND RECEIVED A THE CUSTOMER RECEIVED PUMP ERROR 53 (SOFTWARE ERROR DETECTED) ALARM. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1780KPK. TROUBLESHOOTING WAS PERFORMED FOR THE BATTERY FAILED ALARM, AND THE CUSTOMER STATED THAT THE NO BATTERY COMPARTMENT AND THE BATTERY WAS DAMAGED. TROUBLESHOOTING WAS PERFORMED FOR THE PUMP ERROR, AND THE CUSTOMER WAS ABLE TO CLEAR THE ALARM AND ABLE TO REWIND THE PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22325781 ↗MalfunctionJun 25, 2025Battery Problem; Battery Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED FREQUENT LOW BATTERY AND REPLACE BATTERY NOW ALERTS, REQUIRING BATTERY REPLACEMENT EVERY TWO DAYS. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1780KPK. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS REPLACED AND RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22325752 ↗MalfunctionJun 25, 2025Battery Problem; Wireless Communication Problem; Battery Problem; Wireless Communication Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED FREQUENT REPLACE BATTERY NOW ALARMS, ALONG WITH RAPID BATTERY DEPLETION WITHIN 3-4 DAYS, AND ISSUES WITH SENSOR CONNECTIVITY. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-7020A AND MMT-1780KPK. TROUBLESHOOTING WAS PERFORMED FOR BOTH BATTERY AND SENSOR ISSUES, BUT THE CUSTOMER CONTINUED TO EXPERIENCE THE SAME PROBLEMS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCTS (PUMP AND SENSORS) WERE REPLACED, AND THE PUMP WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22325563 ↗MalfunctionSep 22, 2024Break; Moisture or Humidity Problem; Power Problem; Break; Moisture or Humidity Problem; Power Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED THE INSULIN PUMP SUDDENLY SHUT DOWN AFTER EXPOSED THE PUMP TO MOISTURE. IT WAS ALSO REPORTED THAT THE INSULIN PUMP HAD A CRACK ON THE BATTERY COMPARTMENT. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1780KPK. TROUBLESHOOTING WAS PERFORMED BUT THE ISSUE WAS NOT RESOLVED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. ADVISED THE CUSTOMER TO DISCONNET THE PUMP. THE PRODUCT MMT-1780KPK WAS EXPECTED TO RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20276068 ↗MalfunctionJun 14, 2023Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED PHYSICAL DAMAGE TO THE PUMP. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER DISCONTINUED USING THE DEVICE AND DEVICE WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17123691 ↗InjuryFeb 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED VIA PHONE CALL THAT THE CUSTOMER HAD EXPERIENCED HYPERGLYCEMIA. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 110 MG/DL. THE CUSTOMER DECLINED TO TROUBLESHOOT. THE CUSTOMER WAS TREATED WITH INTRAVENOUS DRIPS. IT WAS UNKNOWN WHETHER THE AUTO MODE FEATURE WAS ACTIVE OR NOT AT THE TIME OF THE INCIDENT. IT WAS UNKNOWN WHETHER THE CUSTOMER WAS AT THE PUMP WITHIN 48 HOURS OF THE REPORTED EVENT. THE PUMP WILL NOT BE RETURNED FOR ANALYSIS. THE RESERVOIR AND INFUSION SET WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 16384071 ↗MalfunctionAug 4, 2022Break; Patient Device Interaction Problem; Break; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15172243 ↗MalfunctionAug 4, 2022Break; Connection Problem; Break; Connection Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP RECEIVED INSULIN FLOW BLOCK ALARM. CUSTOMER STATED THAT THE EVENT WAS RESOLVED BY COMPLETE SET CHANGE. THE INSULIN PUMP WAS CRACK AND DAMAGED. THE LOCATION OF DAMAGE WAS THE RESERVOIR LIP RING. THE RETAINER RING WAS LOOSE, MISSING AND DAMAGED. THE RESERVOIR WAS NOT ABLE TO LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED, THE CUSTOMER WILL DISCONTINUE THE USE OF DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15162246 ↗MalfunctionAug 2, 2022Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS CRACKED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE TO USE THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15147338 ↗MalfunctionAug 2, 2022Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT CLEAR RING WAS BROKEN FROM BLACK PART ON THE INSULIN PUMP. CUSTOMER STATED THAT THE LOCATION OF THE DAMAGE WAS RESERVOIR COMPARTMENT AND THE DAMAGE OCCURRED WHILE USING A SILICONE CASE. CUSTOMER ALSO STATED THAT RESERVOIR WAS ABLE TO LOCK IN PLACE WHEN INSERTED INTO INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED; HOWEVER, THE CUSTOMER WILL DISCONTINUE THE USE OF THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15146850 ↗MalfunctionAug 2, 2022Break; Patient Device Interaction Problem; Break; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RETAINER RING HAD HAIRLINE FRACTURE. THE LOCATION OF DAMAGE WAS THE RETAINER RING. THE RESERVOIR WAS ABLE TO LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED, THE CUSTOMER WILL DISCONTINUE THE USE OF DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15146594 ↗MalfunctionAug 2, 2022Patient Device Interaction Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD DAMAGED RETAINER RING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15146363 ↗MalfunctionJul 28, 2022Patient Device Interaction Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS DAMAGED. THE LOCATION OF DAMAGE WAS THE BOTTOM OF THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED, THE CUSTOMER WILL DISCONTINUE THE USE OF DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15120941 ↗MalfunctionJul 28, 2022Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE PLASTIC RING WAS BROKEN AND THE RETAINER RING WAS DAMAGED. THE LOCATION OF DAMAGE WAS THE RESERVOIR. THE RESERVOIR WAS ABLE TO LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED, THE CUSTOMER WILL DISCONTINUE THE USE OF DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15120193 ↗MalfunctionJul 27, 2022Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RETAINER RING WAS BROKEN AND THE RESERVOIR WAS ABLE TO LOCK IN PLACE WHEN INSERTED INTO INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15112075 ↗MalfunctionJul 26, 2022Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15101818 ↗MalfunctionJul 25, 2022Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP CLIP RAILING WAS CRACKED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15093599 ↗MalfunctionJul 25, 2022Break; Break
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15093540 ↗MalfunctionJul 25, 2022Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE BACK OF THE BATTERY COMPARTMENT WAS CRACKED AND SIDE OF THE INSULIN PUMP ALSO CRACKED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15092639 ↗MalfunctionJul 25, 2022Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS CRACKED. THE LOCATION OF DAMAGE WAS THE BATTERY SIDE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED, THE CUSTOMER WILL DISCONTINUE THE USE OF DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15092281 ↗MalfunctionJul 25, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
MEDTRONIC RECEIVED INFORMATION THAT DEVICE TEST FAILED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15091331 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160017. One submission often covers many models, so check the product and lot details in each record.
- Z-2981-2026 · Jul 13, 2026 · Open, ClassifiedMiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
- Z-1742-2026 · Feb 13, 2026 · Open, ClassifiedMiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
- Z-1265-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 770G (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
- Z-1261-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 670G (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
- Z-1266-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 780G (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
- Z-0193-2023 · Sep 20, 2022 · Open, ClassifiedInsulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742
- Z-0855-2022 · Jan 27, 2022 · Open, ClassifiedMiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782; MiniMed 770G: MMT-1880, MMT-1881
- Z-1771-2021 · Mar 15, 2021 · Open, ClassifiedMiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL; PUMP MMT-1886 MM780G 6.5W BLE MG; KIT MMT-1895WW MM780G 6.5V BLE SF MMOL; KIT MMT-1896WW MM780G 6.5W BLE SF M
- Z-0516-2021 · Apr 30, 2020 · Open, ClassifiedMiniMed 670G, insulin pump, Ref/Model # MMT-1760KPK, MMT-1762KCN, MMT-1762WWK - Product Usage: indicated for continuous or periodic monitoring of glucose levels in the fluid under
- Z-0956-2020 · Nov 21, 2019 · Open, ClassifiedMedtronic MiniMed 670G System with SmartGuard, Ref #s/Model Number(s) - MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, and MM
- Z-2604-2017 · May 8, 2017 · TerminatedMiniMed 640G Insulin Infusion Pump, Catalog No. MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1551, MMT-1751, MMT-1552, MMT-1752.
- Z-2605-2017 · May 8, 2017 · TerminatedMiniMed 670G Insulin Infusion Pump, Catalog No. MMT-1580, MMT-1780, MMT-1782, MMT-1760.
Other Artificial pancreas device system, single hormonal control devices
- MiniMed® 670G MMT-1780KPKMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls78 deaths55,723 reports
- MiniMed® 670G MMT-1780KPKMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control2 recalls51 deaths33,713 reports
- MiniMed® 670G MMT-1780KLMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls8 deaths8,353 reports
- MiniMed® 670G MMT-1780KMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls7 deaths6,776 reports
- MiniMed® 670G MMT-1780KPKMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls10 deaths4,325 reports
- MiniMed™ Reservoir MMT-332AMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control2 deaths3,013 reports
- Guardian™ Sensor (3) MMT-7020AMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control1,304 reports
- MiniMed® 670G MMT-1780KLMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls1 death958 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page