MiniMed® 670G MMT-1780KPK (UDI 00763000283520)Recalls & reported problems
Medtronic Minimed, Inc. · PUMP MMT-1780KPK 670G V4.13 MG BC · Class 3
All models of MiniMed 670G (23)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This MiniMed 670G is an automated insulin pump. Its 4,325 adverse event reports mostly describe power problems, blank displays and high or low blood sugar, and some describe a loose retainer ring. The reports include 10 deaths and 2,143 injuries. 3 open recalls name this device, covering over- or under-delivery when the pump is raised or lowered, shorter battery life after impacts, and worn battery caps that can cut power.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 4,325
- Deaths reported
- 10
- Injuries reported
- 2,143
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 4,325 adverse event reports that list this exact device (DI 00763000283520), received between Oct 29, 2019 and Jul 7, 2025 — classified as 10 deaths, 2,143 injuries, 2,172 malfunctions.
The most frequently reported specific device problem is “Power Problem” (781 reports).
3 FDA recall records name this device's identifier, 3 still open.
12 other recalls cover devices cleared under the same FDA submission (P160017); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 13, 2026Class IIOpen, Classified
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Z-1739-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Jul 31, 2024Class IOpen, Classified
Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.
Z-0003-2025 · Root cause: Device Design · Recall details · FDA record ↗
- May 23, 2022Class IIOpen, Classified
Due to battery cap deterioration, this may potentially result in an incomplete battery circuit (similar to when battery cap is not attached) and a loss of power/functionality of the insulin infusion pumps.
Z-1355-2022 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1,765
- No Apparent Adverse Event977
- Power Problem781
- No Display/Image369
- Patient Device Interaction Problem319
- Mechanical Problem301
- Obstruction of Flow261
- Device Difficult to Program or Calibrate247
Patient problems
- No Consequences Or Impact To Patient1,908
- Hyperglycemia1,732
- Hypoglycemia667
- Elevated Ketones/ Diabetic Ketoacidosis135
- Nausea99
- Vomiting83
- No Clinical Signs, Symptoms or Conditions62
- Blood Loss50
Recent adverse event reports
The 25 most recent of 4,325 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED LOW BLOOD GLUCOSE LEVELS. THE CUSTOMER ALSO RECEIVED A CALIBRATION NOT ACCEPTED ALERTS AND CHANGE SENSOR ALARMS. THE HYPOGLYCEMIA WAS TREATED WITH FOOD. THE BLOOD GLUCOSE VALUE AT THE TIME OF THE EVENT WAS 67 MG/DL. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-7020A AND MMT-1780KPK. TROUBLESHOOTING WAS PARTIALLY PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. NO PRODUCT RETURN IS REQUIRED FOR ANALYSIS MMT-7020A. NO PRODUCT RETURN IS REQUIRED FOR ANALYSIS MMT-1780KPK.
Patient problems: Hypoglycemia
MAUDE report 22409630 ↗MalfunctionMar 13, 2024Break; Excess Flow or Over-Infusion; Break; Excess Flow or Over-Infusion
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS DELIVERING TOO MUCH INSULIN AND RETAINER RING WAS DAMAGED. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. CUSTOMER WILL CONTINUE DISCONTINUE THE USAGE OF THE PUMP AND THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18893054 ↗MalfunctionAug 22, 2023Obstruction of Flow; Obstruction of Flow
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED INSULIN FLOW BLOCKED ALARM DURING FILL TUBING. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE USING THE DEVICE AND THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17593967 ↗MalfunctionJul 3, 2023Computer Software Problem; Computer Software Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP ALARMED WITH PUMP ERROR 53 AND THE ALARM IS RESOLVED. NO TREATMENT WAS NECESSARY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS SUCCESSFULLY PERFORMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17244647 ↗MalfunctionJun 30, 2023Unexpected Therapeutic Results; Unexpected Therapeutic Results
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED THAT PUMP WAS SUSPENDING ON LOW. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER HAD CONTINUED TO USE THE DEVICE. THE INSULIN PUMP WILL NOT BE RETURNED FOR THE PRODUCT ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17236056 ↗InjuryMar 6, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
MEDTRONIC LEGAL RECEIVED INFORMATION REGARDING THE CUSTOMER¿S HYPOGLYCEMIC EVENT FROM THE ATTORNEY OF THE FAMILY. THE INFORMATION RECEIVED ON MAY 5, 2020 STATED THE FOLLOWING - REPORTER ALLEGES: IN OR ABOUT (B)(6) 2019 THE CUSTOMER BEGAN USING A 670G INSULIN PUMP. BEFORE THE HYPOGLYCEMIC EVENT, AND AFTER RECEIVING THE (B)(6) 2019 FIELD SAFETY NOTIFICATION, THE CUSTOMER FOUND THE RETAINER RING ON HIS PUMP FULLY DISPLACED AND DANGLING ON THE INFUSION SET TUBING. THE CUSTOMER WAS ABLE TO POP IT BACK IN AND CONTINUED PUMP THERAPY. ON OR ABOUT (B)(6) 2020, THE CUSTOMER SUFFERED A SEVERE HYPOGLYCEMIC EVENT FOR NO REASON HE COULD SEE, BUT NOTICED THAT THE RETAINER RING WAS FULLY DISPLACED AGAIN. THE CUSTOMER HAD PREVIOUSLY CALLED IN ABOUT THE RECALL AND STATED THE RESERVOIR WAS LOCKING IN PLACE.
Patient problems: Hypoglycemia
MAUDE report 16483716 ↗MalfunctionJan 30, 2023Circuit Failure; Circuit Failure
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A HARDWARE LOW LEVEL FAILURE INSULIN PUMP ERROR ALARM. CUSTOMER WAS ABLE TO CLEAR THE ALARM. CUSTOMER IS ABLE TO COMPLETE PUMP REWIND. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. INSULIN PUMP WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16264795 ↗MalfunctionJun 26, 2022Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14824048 ↗MalfunctionJun 26, 2022No Display/Image; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED, BLANK DISPLAY OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14824032 ↗MalfunctionJun 26, 2022No Display/Image; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BLANK DISPLAY, BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14823744 ↗MalfunctionJun 26, 2022No Display/Image; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BLANK DISPLAY, BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14823732 ↗MalfunctionJun 26, 2022Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14823620 ↗MalfunctionJun 24, 2022Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797368 ↗MalfunctionJun 24, 2022Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797360 ↗MalfunctionJun 24, 2022No Display/Image; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED, BLANK DISPLAY OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797303 ↗MalfunctionJun 24, 2022No Display/Image; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED, BLANK DISPLAY OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797299 ↗MalfunctionJun 24, 2022Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797297 ↗MalfunctionJun 24, 2022Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797242 ↗MalfunctionJun 24, 2022Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797115 ↗MalfunctionJun 24, 2022Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797098 ↗MalfunctionJun 24, 2022No Display/Image; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED, BLANK DISPLAY OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797079 ↗MalfunctionJun 24, 2022No Display/Image; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BLANK DISPLAY, BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797059 ↗MalfunctionJun 24, 2022Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797031 ↗MalfunctionJun 24, 2022Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14797026 ↗MalfunctionJun 24, 2022Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14796916 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160017. One submission often covers many models, so check the product and lot details in each record.
- Z-2981-2026 · Jul 13, 2026 · Open, ClassifiedMiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
- Z-1742-2026 · Feb 13, 2026 · Open, ClassifiedMiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
- Z-1265-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 770G (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
- Z-1266-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 780G (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
- Z-1261-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 670G (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
- Z-0193-2023 · Sep 20, 2022 · Open, ClassifiedInsulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742
- Z-0855-2022 · Jan 27, 2022 · Open, ClassifiedMiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782; MiniMed 770G: MMT-1880, MMT-1881
- Z-1771-2021 · Mar 15, 2021 · Open, ClassifiedMiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL; PUMP MMT-1886 MM780G 6.5W BLE MG; KIT MMT-1895WW MM780G 6.5V BLE SF MMOL; KIT MMT-1896WW MM780G 6.5W BLE SF M
- Z-0516-2021 · Apr 30, 2020 · Open, ClassifiedMiniMed 670G, insulin pump, Ref/Model # MMT-1760KPK, MMT-1762KCN, MMT-1762WWK - Product Usage: indicated for continuous or periodic monitoring of glucose levels in the fluid under
- Z-0956-2020 · Nov 21, 2019 · Open, ClassifiedMedtronic MiniMed 670G System with SmartGuard, Ref #s/Model Number(s) - MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, and MM
- Z-2604-2017 · May 8, 2017 · TerminatedMiniMed 640G Insulin Infusion Pump, Catalog No. MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1551, MMT-1751, MMT-1552, MMT-1752.
- Z-2605-2017 · May 8, 2017 · TerminatedMiniMed 670G Insulin Infusion Pump, Catalog No. MMT-1580, MMT-1780, MMT-1782, MMT-1760.
Other Automated insulin dosing device system, single hormonal control devices
- MiniMed® 670G MMT-1780KPKMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls78 deaths55,723 reports
- MiniMed® 670G MMT-1780KPKMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control2 recalls51 deaths33,713 reports
- MiniMed®670G MMT-1780KPKMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls39 deaths28,789 reports
- MiniMed® 670G MMT-1780KLMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls8 deaths8,353 reports
- MiniMed® 670G MMT-1780KMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls7 deaths6,776 reports
- MiniMed™ Reservoir MMT-332AMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control2 deaths3,013 reports
- Guardian™ Sensor (3) MMT-7020AMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control1,304 reports
- MiniMed® 670G MMT-1780KLMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls1 death958 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page