MiniMed® 670G MMT-1780KL (UDI 00763000166526)Recalls & reported problems

Medtronic Minimed, Inc. · PUMP MMT-1780KL 670G V4.10 BK SF MG · Class 3

All models of MiniMed 670G (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This MiniMed 670G insulin pump appears in 8,353 adverse event reports, including 8 deaths and 1,761 injuries. The reports mostly describe parts breaking or cracking, power problems and blank pump screens. All 3 of its recalls are open, covering insulin over- or under-delivery when the pump is raised or lowered, battery life loss after drops, and battery cap wear.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
8,353
Deaths reported
8
Injuries reported
1,761
Recalls naming it
3

Summary

The FDA's MAUDE database contains 8,353 adverse event reports that list this exact device (DI 00763000166526), received between Mar 7, 2019 and Jul 7, 2025 — classified as 8 deaths, 1,761 injuries, 6,584 malfunctions.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (2,860 reports).

3 FDA recall records name this device's identifier, 3 still open.

12 other recalls cover devices cleared under the same FDA submission (P160017); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 13, 2026Class IIOpen, Classified

    All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

    Z-1739-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Jul 31, 2024Class IOpen, Classified

    Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.

    Z-0003-2025 · Root cause: Device Design · Recall details · FDA record ↗

  3. May 23, 2022Class IIOpen, Classified

    Due to battery cap deterioration, this may potentially result in an incomplete battery circuit (similar to when battery cap is not attached) and a loss of power/functionality of the insulin infusion pumps.

    Z-1355-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Patient Device Interaction Problem2,860
  • Adverse Event Without Identified Device or Use Problem1,606
  • No Apparent Adverse Event1,150
  • Break963
  • Power Problem856
  • No Display/Image648
  • Connection Problem619
  • Moisture or Humidity Problem473

Patient problems

  • No Consequences Or Impact To Patient4,970
  • Hyperglycemia1,562
  • No Clinical Signs, Symptoms or Conditions1,200
  • Hypoglycemia694
  • Elevated Ketones/ Diabetic Ketoacidosis100
  • Nausea92
  • Blood Loss62
  • Vomiting55

Recent adverse event reports

The 25 most recent of 8,353 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA AFTER AIR TRAVEL AND ENCOUNTERED SENSOR ISSUES, INCLUDING "SENSOR UPDATING" ALERTS. BLOOD GLUCOSE VALUE AT THE TIME OF THE EVENT WAS 350 MG/DL, ACCOMPANIED BY SYMPTOMS SUCH AS HEADACHE AND THIRST. THE CUSTOMER VISITED EMERGENCY ROOM FOR THE TREATMENT. THE EVENT INVOLVED PRODUCT(S) MMT-1780KL, MMT-7020A, MMT-864A, AND MMT-332A. TROUBLESHOOTING WAS PERFORMED, WHICH INCLUDED EXPLAINING POSSIBLE CAUSES OF SENSOR ALERTS, ADVISING THE CUSTOMER TO MONITOR BLOOD GLUCOSE LEVELS, AND RECOMMENDING CHANGES TO THE INFUSION SET, RESERVOIR, AND INSULIN. THE CUSTOMER DECLINED TO CHECK FOR LEAKS, SITE/INSULIN ISSUES, OR PERFORM THE HIGH PRESSURE TEST BUT CONFIRMED THAT THE CURRENT SET SEEMED TO BE WORKING FINE. THE CUSTOMER WAS ADVISED TO TURN ON AIRPLANE MODE DURING FLIGHTS, MONITOR BLOOD GLUCOSE LEVELS, AND CALL BACK IF ISSUES PERSISTED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. NO PRODUCT RETURN IS REQUIRED FOR MMT-1780KL, MMT-7020A, MMT-864A, AND MMT-332A.

    Patient problems: Dehydration; Headache; Hyperglycemia

    MAUDE report 22409635 ↗
  • MalfunctionJul 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA. THE BLOOD GLUCOSE VALUE AT THE TIME OF THE EVENT WAS 360 MG/DL. THE HYPERGLYCEMIA WAS TREATED WITH AN INSULIN PUMP. THE EVENT INVOLVED PRODUCT(S) MMT-1780KL. TROUBLESHOOTING WAS PERFORMED, AND THE CUSTOMER WAS USING THE INSULIN PUMP WITHIN 48 HOURS OF THE EVENT. ALSO, THE CUSTOMER WAS USING THE AUTO MODE FEATURE WAS ACTIVE OR NOT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. NO PRODUCT RETURN IS REQUIRED FOR MMT-1780KL.

    Patient problems: Hyperglycemia

    MAUDE report 22409592 ↗
  • MalfunctionJun 14, 2023Break; Break

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A CRACK ON THE BACK OF THE PUMP. CUSTOMER STATED THAT THE INSULIN PUMP FALLS OFF ACCIDENTALLY. TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER CONFIRMED THAT THERE WAS NO DAMAGE TO THE PUMP'S RESERVOIR LIP RING AND THERE WAS NO OUT-OF-BOX DAMAGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO BACKUP PLAN PER HEALTHCARE PROFESSIONAL (HCP'S) INSTRUCTIONS. THE DEVICE WAS EXPECTED TO BE RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17123715 ↗
  • MalfunctionMar 14, 2023Obstruction of Flow; Obstruction of Flow

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT CUSTOMER REPORTED INSULIN FLOW BLOCK ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER WILL CONTINUE TO USE THE DEVICE. THE INSULIN PUMP WILL NOT BE RETURNED FOR THE PRODUCT ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16535435 ↗
  • MalfunctionAug 2, 2022Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15146185 ↗
  • MalfunctionAug 2, 2022Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RETAINER RING WAS CRACKED. THE RESERVOIR WAS ABLE TO LOCK IN THE PLACE. CUSTOMER STATED THAT THE INSULIN PUMP HAD NEITHER BEEN BUMPED NOR DROPPED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE TO USE THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15144575 ↗
  • MalfunctionAug 2, 2022Break; Break

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP BELT CLIP WAS BROKEN AND THE INSULIN PUMP WAS DAMAGED. THE TYPE OF DAMAGE WAS PART COMING OFF. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED, THE CUSTOMER WILL DISCONTINUE THE USE OF DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15143640 ↗
  • MalfunctionJul 28, 2022Break; Patient Device Interaction Problem; Break; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Hyperglycemia; Hypoglycemia

    MAUDE report 15121198 ↗
  • MalfunctionJul 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT CUSTOMER STATED BLOOD ON SITE ISSUE. CUSTOMER STATED THAT THE SITE WAS ACTIVELY BLEEDING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Pain

    MAUDE report 15121083 ↗
  • MalfunctionJul 26, 2022Break; Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15102036 ↗
  • MalfunctionJul 26, 2022Break; No Apparent Adverse Event; Patient Device Interaction Problem; Break; No Apparent Adverse Event; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RETAINER RING WAS CRACK AND DAMAGED. THE LOCATION OF DAMAGE WAS THE RETAINER RING. THE RESERVOIR WAS ABLE TO LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED, THE CUSTOMER WILL DISCONTINUE THE USE OF DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15101882 ↗
  • MalfunctionJul 25, 2022Break; Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15092357 ↗
  • MalfunctionJul 25, 2022Break; Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15092247 ↗
  • MalfunctionJul 20, 2022Obstruction of Flow; Patient Device Interaction Problem; Obstruction of Flow; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD INSULIN FLOW BLOCKED ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15063551 ↗
  • MalfunctionJul 20, 2022Break; Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15063468 ↗
  • MalfunctionJul 19, 2022Break; Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15052279 ↗
  • MalfunctionJul 19, 2022Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THERE WAS CRACK ON THE BATTERY COMPARTMENT FROM TOP TO BOTTOM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED, THE CUSTOMER WILL DISCONTINUE THE USE OF DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15050522 ↗
  • MalfunctionJul 18, 2022No Display/Image; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED, BLANK DISPLAY OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15044653 ↗
  • MalfunctionJul 18, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15044489 ↗
  • MalfunctionJul 18, 2022Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15044437 ↗
  • MalfunctionJul 18, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT. UPDATED SUMMARY FOR SVN (B)(4). INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP CENTER BUTTON HAS A CRACKS. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15043327 ↗
  • MalfunctionJul 18, 2022Break; Break

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RETAINER RING WAS BROKEN. CUSTOMER STATED THAT THE RESERVOIR WAS ABLE TO LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15039897 ↗
  • MalfunctionJul 18, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    INFORMATION RECEIVED BY MEDTRONIC INDICATES THAT THE NEW PUMP WAS NOT SHIPPED TO CUSTOMER. CUSTOMER STATED HER PUMP FAILED ON THURSDAY AND WAS RETURN BACK TO THE SENDER. NO HARM REQUIRING MEDICAL INTERVENTION WAS REQUIRED. THE DEVICE WAS RETURN BACK FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15039737 ↗
  • DeathJul 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    (B)(4) CREATED RMA TO RECEIVE IN INFUSION SET AND RESERVOIRS RECEIVED IN WITH PUMP. SHIP: NOTHING / RETURN: MMT-864 / MMT-332A (B)(4) REFERENCED REPORT PROVIDED TO MEDTRONIC OTHER (B)(6) 2019 FORWARDED TO MEDTRONIC TECHNICAL SUPPORT DEPARTMENT / INCIDENT DATE FROM (B)(6) 2019. START REPORT. CUSTOMER'S WIFE WOULD LIKE TO RETURN THE CUSTOMERS PUMP AND DOES NOT WANT IT BACK. SEND BOX TO: (B)(6) END REPORT. OUTCOME: RETURN IS ALREADY BEING PROCESSED. (B)(4) CAMPUSSHIP RETURN REQUEST FOR THE RETURN OF THE DEVICE/PUMP TO MEDTRONIC. (B)(4) CORRECTED TO REFLECT THE PROPER INFUSION SET PRODUCT ID AND BATCH NUMBER AS WELL AS SVN (B)(4) CORRECTED TO REFLECT THE PROPER BATCH NUMBER. (B)(4) I UPDATED THE SVN TO REFLECT THE PROPER REPORTED BY TYPE AND REPORTED BY SINCE THE SPOUSE/(B)(6) OF THE CUSTOMER CONTACTED THE MEDTRONIC 24-HOUR TECHNICAL SUPPORT DEPARTMENT TO PROVIDE THE DEATH REPORT. (B)(4) INITIAL NOTES: CUSTOMER'S WIFE CALLED AND SAID SHE RECEIVED A CALL ABOUT HER DECEASED HUSBANDS PUMP AND SHE DOES NOT KNOW WHAT TO DO. INQUIRED WHAT LED UP TO THE COMPLAINT. CUSTOMER RESPONSE: POS NOTIFIED OF CUSTOMER PASSING. WIFE WAS CONTACTED TO INFORMATION REGARDING THE EVENTS OF CUSTOMER'S DEATH. D

    Patient problems: Cancer

    MAUDE report 15038508 ↗
  • MalfunctionJul 18, 2022Circuit Failure; Intermittent Communication Failure; Circuit Failure; Intermittent Communication Failure

    MEDTRONIC RECEIVED INFORMATION THAT PUMP ERROR 81, PUMP ERROR 19, PUMP ERROR 13 OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15038126 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160017. One submission often covers many models, so check the product and lot details in each record.

  • Z-2981-2026 · Jul 13, 2026 · Open, Classified
    MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
  • Z-1742-2026 · Feb 13, 2026 · Open, Classified
    MiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
  • Z-1261-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 670G (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
  • Z-1266-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 780G (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
  • Z-1265-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 770G (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
  • Z-0193-2023 · Sep 20, 2022 · Open, Classified
    Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742
  • Z-0855-2022 · Jan 27, 2022 · Open, Classified
    MiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782; MiniMed 770G: MMT-1880, MMT-1881
  • Z-1771-2021 · Mar 15, 2021 · Open, Classified
    MiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL; PUMP MMT-1886 MM780G 6.5W BLE MG; KIT MMT-1895WW MM780G 6.5V BLE SF MMOL; KIT MMT-1896WW MM780G 6.5W BLE SF M
  • Z-0516-2021 · Apr 30, 2020 · Open, Classified
    MiniMed 670G, insulin pump, Ref/Model # MMT-1760KPK, MMT-1762KCN, MMT-1762WWK - Product Usage: indicated for continuous or periodic monitoring of glucose levels in the fluid under
  • Z-0956-2020 · Nov 21, 2019 · Open, Classified
    Medtronic MiniMed 670G System with SmartGuard, Ref #s/Model Number(s) - MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, and MM
  • Z-2605-2017 · May 8, 2017 · Terminated
    MiniMed 670G Insulin Infusion Pump, Catalog No. MMT-1580, MMT-1780, MMT-1782, MMT-1760.
  • Z-2604-2017 · May 8, 2017 · Terminated
    MiniMed 640G Insulin Infusion Pump, Catalog No. MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1551, MMT-1751, MMT-1552, MMT-1752.