MiniMed®670G MMT-1760KPKRecalls & reported problems

Medtronic Minimed, Inc. · KIT MMT-1760KPK 670G V4.10 EN MG · Class 3

All models of MiniMed 670G (23) · MiniMed670G (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Medtronic MiniMed 670G insulin pump delivers insulin automatically. Its 82 adverse event reports, with no deaths and 9 injuries, mostly mention cracked battery caps, blank screens, alarms, and power loss, often with no reported effect on the patient. Three open recalls cover insulin over- or under-delivery from height changes, shorter battery life after impacts, and battery cap deterioration cutting power.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
82
Deaths reported
0
Injuries reported
9
Recalls naming it
3

Summary

The FDA's MAUDE database contains 82 adverse event reports that list this exact device (DI 00763000096984), received between Oct 15, 2018 and Jul 8, 2022 — classified as 9 injuries, 73 malfunctions.

The most frequently reported specific device problem is “Device Alarm System” (15 reports).

3 FDA recall records name this device's identifier, 3 still open.

12 other recalls cover devices cleared under the same FDA submission (P160017); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Feb 13, 2026Class IIOpen, Classified

    All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

    Z-1739-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Jul 31, 2024Class IOpen, Classified

    Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.

    Z-0003-2025 · Root cause: Device Design · Recall details · FDA record ↗

  3. May 23, 2022Class IIOpen, Classified

    Due to battery cap deterioration, this may potentially result in an incomplete battery circuit (similar to when battery cap is not attached) and a loss of power/functionality of the insulin infusion pumps.

    Z-1355-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Alarm System15
  • Appropriate Device Problem Term/Code Not Available14
  • No Display/Image12
  • Patient Device Interaction Problem12
  • No Apparent Adverse Event10
  • Adverse Event Without Identified Device or Use Problem7
  • Power Problem7
  • Circuit Failure6

Patient problems

  • No Consequences Or Impact To Patient56
  • No Clinical Signs, Symptoms or Conditions13
  • Hyperglycemia9
  • Hypoglycemia7
  • Insufficient Information2

Recent adverse event reports

The 25 most recent of 82 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 8, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14953644 ↗
  • MalfunctionJul 4, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14906754 ↗
  • MalfunctionJul 4, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14906208 ↗
  • MalfunctionJul 3, 2022Break

    MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14899565 ↗
  • MalfunctionJul 2, 2022Computer Software Problem; Communication or Transmission Problem

    MEDTRONIC RECEIVED INFORMATION THAT PUMP ERROR 3, PUMP ERROR 54 OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14898364 ↗
  • MalfunctionJun 27, 2022Break; No Display/Image; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED, DAMAGE (PHYSICAL OR COSMETIC), BLANK DISPLAY OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14838042 ↗
  • MalfunctionJun 27, 2022Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14831365 ↗
  • MalfunctionJun 26, 2022No Display/Image; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION THAT BLANK DISPLAY, BATTERY CAP CRACKED/DAMAGED OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14823826 ↗
  • MalfunctionJun 25, 2022No Display/Image; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION THAT BATTERY CAP CRACKED/DAMAGED, BLANK DISPLAY OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14820240 ↗
  • MalfunctionFeb 10, 2021Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A CRACK ON THE BACK OF THE PUMP TOWARDS BATTERY COMPARTMENT SIDE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11306001 ↗
  • MalfunctionFeb 2, 2021Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RETAINER RING WAS CRACK. THE CUSTOMER STATED THAT THE RESERVOIR DID LOCK INTO PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 11264582 ↗
  • MalfunctionJan 29, 2021Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A DETACHED RETAINER RING. THE CUSTOMER STATED THAT THE RESERVOIR WAS ABLE TO LOCK INTO PLACE WHEN INSERTED INTO INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 11250243 ↗
  • MalfunctionOct 13, 2020Electrical /Electronic Property Problem; Loss of Data; Electrical /Electronic Property Problem; Loss of Data

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP RECEIVED PUMP ERROR. CUSTOMER WAS ABLE TO CLEAR ALARM AND REWIND PROCESS. CUSTOMER STATED THAT INSULIN PUMP EXPERIENCED UNEXPECTED RESTART. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10673117 ↗
  • MalfunctionSep 17, 2020No Apparent Adverse Event; Patient Device Interaction Problem; No Apparent Adverse Event; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RETAINER RING WAS MISSING. THE RESERVOIR WAS ABLE TO LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10550955 ↗
  • MalfunctionJul 6, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD ISSUE WITH KEYPAD AND NOT RESOLVED BY INSERTING NEW BATTERY. CUSTOMERS STATED THAT NUMBERS WERE NOT RAMPING ON THEIR OWN. CUSTOMER STATED THAT THE SCROLL BAR WAS NOT MOVING WITH NO INPUT. CUSTOMER STATED THAT THE BUTTON WAS NOT UNRESPONSIVE DURING AN EASY BOLUS ATTEMPT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10237790 ↗
  • MalfunctionJul 6, 2020Power Problem; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THEIR INSULIN PUMP HAD CRITICAL PUMP ERROR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10232550 ↗
  • MalfunctionJul 1, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD STUCK BUTTON ALARM. THE CUSTOMER STATED THAT THE HOME SCREEN WAS NOT RESPONSIVE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10217076 ↗
  • MalfunctionJun 29, 2020Power Problem; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD CRITICAL PUMP ERROR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10205442 ↗
  • MalfunctionJun 24, 2020No Display/Image; Mechanical Problem; No Display/Image; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MULTIPLE INSULIN PUMP ERRORS AND THE SCREEN WAS BLANK. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10188101 ↗
  • MalfunctionJun 22, 2020Moisture or Humidity Problem; Power Problem; Moisture or Humidity Problem; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD AN OPEN BOOK IMAGE. THE CUSTOMER WAS DOING ACTIVITY IN THE POOL AND THERE WAS NO OTHER ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10177888 ↗
  • MalfunctionMay 22, 2020Power Problem; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER HAD TO CHANGE BATTERY THREE TIMES. THE CUSTOMER STATED THAT THE BATTERY CAP OR THE BATTERY COMPARTMENT IS NOT DAMAGED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 10081762 ↗
  • MalfunctionMay 12, 2020Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CLEAR PLASTIC RESERVOIR RING WAS LOOSE AND FALLEN OFF HOWEVER THE RESERVOIR WAS ABLE TO LOCK IN PLACE WHEN INSERTED INTO THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10045961 ↗
  • MalfunctionMay 6, 2020Power Problem; Power Problem

    CUSTOMER REPORTED VIA PHONE CALL THAT THEY GOT CRITICAL PUMP ERROR. CUSTOMER BLOOD GLUCOSE READING WAS UNKNOWN. TROUBLESHOOT WAS PERFORMED. CUSTOMER REPORTS THEY RECEIVED THE OPEN BOOK IMAGE ON THE PUMP SCREEN. INSULIN PUMP WILL BE RETURNING FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10031500 ↗
  • MalfunctionApr 20, 2020Connection Problem; Patient Device Interaction Problem; Connection Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A LOOSE RETAINER RING. THE CUSTOMER STATED THAT THE RESERVOIR DOES NOT LOCK INTO PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9976110 ↗
  • MalfunctionApr 19, 2020Connection Problem; Patient Device Interaction Problem; Connection Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A LOOSE RETAINER RING. THE CUSTOMER STATED THAT THE RESERVOIR DOES NOT LOCK INTO PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9975841 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160017. One submission often covers many models, so check the product and lot details in each record.

  • Z-2981-2026 · Jul 13, 2026 · Open, Classified
    MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
  • Z-1742-2026 · Feb 13, 2026 · Open, Classified
    MiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
  • Z-1266-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 780G (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
  • Z-1261-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 670G (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
  • Z-1265-2025 · Jan 31, 2025 · Open, Classified
    MiniMed insulin pump, REF: 770G (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
  • Z-0193-2023 · Sep 20, 2022 · Open, Classified
    Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742
  • Z-0855-2022 · Jan 27, 2022 · Open, Classified
    MiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782; MiniMed 770G: MMT-1880, MMT-1881
  • Z-1771-2021 · Mar 15, 2021 · Open, Classified
    MiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL; PUMP MMT-1886 MM780G 6.5W BLE MG; KIT MMT-1895WW MM780G 6.5V BLE SF MMOL; KIT MMT-1896WW MM780G 6.5W BLE SF M
  • Z-0516-2021 · Apr 30, 2020 · Open, Classified
    MiniMed 670G, insulin pump, Ref/Model # MMT-1760KPK, MMT-1762KCN, MMT-1762WWK - Product Usage: indicated for continuous or periodic monitoring of glucose levels in the fluid under
  • Z-0956-2020 · Nov 21, 2019 · Open, Classified
    Medtronic MiniMed 670G System with SmartGuard, Ref #s/Model Number(s) - MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, and MM
  • Z-2604-2017 · May 8, 2017 · Terminated
    MiniMed 640G Insulin Infusion Pump, Catalog No. MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1551, MMT-1751, MMT-1552, MMT-1752.
  • Z-2605-2017 · May 8, 2017 · Terminated
    MiniMed 670G Insulin Infusion Pump, Catalog No. MMT-1580, MMT-1780, MMT-1782, MMT-1760.

Other Automated insulin dosing , threshold suspend devices

All Automated insulin dosing , threshold suspend devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page