TRANSLACE™ Spinal Tethering System 8219500 (UDI 00763000108397)Recalls & reported problems

Medtronic Sofamor Danek, Inc. · TETHER 8219500 TETHER · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00763000108397), received between Oct 23, 2019 and Mar 3, 2025 — classified as 2 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Device Damaged by Another Device” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Damaged by Another Device6
  • Break5

Patient problems

  • No Known Impact Or Consequence To Patient10
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 3, 2025Break; Break

    INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PATIENT UNDERGONE T2-L2 POSTERIOR FUSION FOR SCOLIOSIS. IT WAS REPORTED THAT PRODUCT SHEARED AND SNAPPED WHEN THE TENCHENER WAS APPLIED.NO PATIENT SYMPTOMS WERE REPORTED. NO FURTHER COMPLICATIONS WERE REPORTED/ANTICIPATED. ADDITIONAL INFORMATION WAS RECEIVED FROM THE REP THAT CONNECTOR PERFORMED AS EXPECTED, NO ISSUES WITH THE CONNECTOR. NO REVISION SURGERY PLANNED DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21502660 ↗
  • MalfunctionMay 13, 2020Break; Break

    IT WAS REPORTED THAT THE PATIENT PRESENTED WITH SCOLIOSIS; AND UNDERWENT A CORRECTION SURGERY AT T6-L3. INTRA-OP, AFTER PLACEMENT WAS PERFORMED, THE TETHERS BROKE WHILE APPLYING TENSION WITH THE TENSIONER. TENSION WAS APPLIED WHEN CORRECTION WAS ONCE PERFORMED. AFTER PROVISIONAL FIXATION, LOOSENING WAS PERFORMED. AND WHEN TENSION WAS APPLIED AGAIN DURING THE SECOND CORRECTION, THE TETHERS BROKE. THEREFORE, PRODUCT FROM ANOTHER MANUFACTURER (NESPLON) WAS USED. THE SURGEON WONDERED WHETHER CORRECTION WAS PERFORMED EXCESSIVELY. THE BROKEN PRODUCT WAS REMOVED WITH NO FRAGMENT OF THE PRODUCT REMAINING INSIDE THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10052258 ↗
  • MalfunctionMay 13, 2020Break; Break

    IT WAS REPORTED THAT THE PATIENT PRESENTED WITH SCOLIOSIS; AND UNDERWENT A CORRECTION SURGERY AT T6-L3. INTRA-OP, AFTER PLACEMENT WAS PERFORMED, THE TETHERS BROKE WHILE APPLYING TENSION WITH THE TENSIONER. TENSION WAS APPLIED WHEN CORRECTION WAS ONCE PERFORMED. AFTER PROVISIONAL FIXATION, LOOSENING WAS PERFORMED. AND WHEN TENSION WAS APPLIED AGAIN DURING THE SECOND CORRECTION, THE TETHERS BROKE. THEREFORE, PRODUCT FROM ANOTHER MANUFACTURER (NESPLON) WAS USED. THE SURGEON WONDERED WHETHER CORRECTION WAS PERFORMED EXCESSIVELY. THE BROKEN PRODUCT WAS REMOVED WITH NO FRAGMENT OF THE PRODUCT REMAINING INSIDE THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10052243 ↗
  • MalfunctionDec 6, 2019Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT THE PATIENT PRESENTED WITH SCOLIOSIS; AND UNDERWENT SPINAL FUSION, SAGITTAL RE-ALIGNMENT AND DEROTATION. INTRA-OP, DURING CURVE CORRECTION, WHILE TENSIONING THE TETHER, THE INSTRUMENT STARTED DAMAGING THE TISSUE OF THE TETHER; HENCE, THE PHYSICIAN HAD TO STOP APPLYING TENSION. THEN, THE DAMAGED DISTAL PART OF THE TETHER WAS CUT AND REMOVED. THE BROKEN PART DID NOT REMAIN INSIDE THE PATIENT. REPORTEDLY, A VERY GOOD CORRECTION OF THE SCOLIOTIC CURVE COULD NOT BE PERFORMED DUE TO THIS EVENT; HOWEVER, NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9429403 ↗
  • MalfunctionDec 6, 2019Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT THE PATIENT PRESENTED WITH SCOLIOSIS; AND UNDERWENT SPINAL FUSION, SAGITTAL RE-ALIGNMENT AND DEROTATION. INTRA-OP, DURING CURVE CORRECTION, WHILE TENSIONING THE TETHER, THE INSTRUMENT STARTED DAMAGING THE TISSUE OF THE TETHER; HENCE, THE PHYSICIAN HAD TO STOP APPLYING TENSION. THEN, THE DAMAGED DISTAL PART OF THE TETHER WAS CUT AND REMOVED. THE BROKEN PART DID NOT REMAIN INSIDE THE PATIENT. REPORTEDLY, A VERY GOOD CORRECTION OF THE SCOLIOTIC CURVE COULD NOT BE PERFORMED DUE TO THIS EVENT; HOWEVER, NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9429319 ↗
  • MalfunctionDec 6, 2019Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT THE PATIENT PRESENTED WITH SCOLIOSIS; AND UNDERWENT SPINAL FUSION, SAGITTAL RE-ALIGNMENT AND DEROTATION. INTRA-OP, DURING CURVE CORRECTION, WHILE TENSIONING THE TETHER, THE INSTRUMENT STARTED DAMAGING THE TISSUE OF THE TETHER; HENCE, THE PHYSICIAN HAD TO STOP APPLYING TENSION. THEN, THE DAMAGED DISTAL PART OF THE TETHER WAS CUT AND REMOVED. THE BROKEN PART DID NOT REMAIN INSIDE THE PATIENT. REPORTEDLY, A VERY GOOD CORRECTION OF THE SCOLIOTIC CURVE COULD NOT BE PERFORMED DUE TO THIS EVENT; HOWEVER, NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9428825 ↗
  • MalfunctionNov 8, 2019Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT THE PATIENT UNDERWENT SCOLIOTIC CURVE CORRECTION AFTER SHE WAS DIAGNOSED WITH SCOLIOSIS. INTRA-OP, WHEN THE SURGEON PERFORMED A HUGE TRACTION WITH CLAMPS WITH THE AIM OF RE-ALIGNING THE ANTERIO-POSTERIOR CURVE, THE TETHERS GOT DAMAGED IN CORRESPONDENCE TO THE TENSIONER. THE DISTAL PART OF THE TETHER WAS CUT AS INDICATED IN THE SURGICAL TECHNIQUE. DEROTATION WAS APPLIED USING GRIPPER AND APPLYING TENSION TO THE CLAMP. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9294284 ↗
  • MalfunctionNov 8, 2019Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT THE PATIENT UNDERWENT SCOLIOTIC CURVE CORRECTION AFTER SHE WAS DIAGNOSED WITH SCOLIOSIS. INTRA-OP, WHEN THE SURGEON PERFORMED A HUGE TRACTION WITH CLAMPS WITH THE AIM OF RE-ALIGNING THE ANTERIO-POSTERIOR CURVE, THE TETHERS GOT DAMAGED IN CORRESPONDENCE TO THE TENSIONER. THE DISTAL PART OF THE TETHER WAS CUT AS INDICATED IN THE SURGICAL TECHNIQUE. DEROTATION WAS APPLIED USING GRIPPER AND APPLYING TENSION TO THE CLAMP. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9294256 ↗
  • MalfunctionNov 8, 2019Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT THE PATIENT UNDERWENT SCOLIOTIC CURVE CORRECTION AFTER SHE WAS DIAGNOSED WITH SCOLIOSIS. INTRA-OP, WHEN THE SURGEON PERFORMED A HUGE TRACTION WITH CLAMPS WITH THE AIM OF RE-ALIGNING THE ANTERIO-POSTERIOR CURVE, THE TETHERS GOT DAMAGED IN CORRESPONDENCE TO THE TENSIONER. THE DISTAL PART OF THE TETHER WAS CUT AS INDICATED IN THE SURGICAL TECHNIQUE. DEROTATION WAS APPLIED USING GRIPPER AND APPLYING TENSION TO THE CLAMP. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9294245 ↗
  • InjuryOct 23, 2019Break; Break

    IT WAS REPORTED THAT POST-OP, TRANSLACE BANDS SNAPPED AT SOME POINT. HENCE, ON (B)(6) 2019 A REVISION SURGERY WAS PREFORMED DUE TO TRANSLACE BREAKAGE. NO PATIENT COMPLICATIONS WERE REPORTED DUE TO THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9225060 ↗
  • InjuryOct 23, 2019Break; Break

    IT WAS REPORTED THAT POST-OP, TRANSLACE BANDS SNAPPED AT SOME POINT. HENCE, ON (B)(6) 2019 A REVISION SURGERY WAS PREFORMED DUE TO TRANSLACE BREAKAGE. NO PATIENT COMPLICATIONS WERE REPORTED DUE TO THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9225048 ↗

FDA clearance and classification

Cleared through 510(k) K163181 on May 16, 2017 (decision: Substantially Equivalent). Submitted by Medtronic Sofamor Danek USA, Inc. as “TRANSLACE™ Spinal Tethering System”. The FDA review took 183 days. FDA record ↗

The FDA regulates product code OWI (bone fixation cerclage, sublaminar) under 21 CFR 888.3010 as a Class II device. 21 CFR 888.3010 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: Acts as a bone anchor for temporary stabilization, or used in conjunction with other medical implants of similar metals when wiring is needed, during development of a spinal fusion.

Common questions about the TRANSLACE™ Spinal Tethering System 8219500

Has the TRANSLACE™ Spinal Tethering System 8219500 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00763000108397), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the TRANSLACE™ Spinal Tethering System 8219500?

The FDA's MAUDE database lists 11 adverse event reports for this device, received between Oct 23, 2019 and Mar 3, 2025. In those reports, the most common specific device problems are “Device Damaged by Another Device” (6) and “Break” (5). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the TRANSLACE™ Spinal Tethering System 8219500?

None of the reports are classified as deaths, 2 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the TRANSLACE™ Spinal Tethering System 8219500 FDA approved?

It was cleared through 510(k) K163181 on May 16, 2017. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the TRANSLACE™ Spinal Tethering System 8219500 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the TRANSLACE™ Spinal Tethering System 8219500, and is it still sold?

The FDA's device database lists Medtronic Sofamor Danek, Inc. as the company responsible for it (brand TRANSLACE™ Spinal Tethering System, model 8219500). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Bone fixation cerclage, sublaminar devices

All Bone fixation cerclage, sublaminar devices →